DIAGNOSTIC TESTING USING ASHRAE 110. Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc

Size: px
Start display at page:

Download "DIAGNOSTIC TESTING USING ASHRAE 110. Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc"

Transcription

1 DIAGNOSTIC TESTING USING ASHRAE 110 Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc

2 ASHARE 110 A Method of Testing Three components Face velocity measurements Smoke visualization Tracer gas testing Owner sets acceptance levels The problem What do you do when the results are unacceptable?

3 General Approach Operational Problems Tracer gas leakage Reentry Hood problems Aerodynamics Operations Room Conditions Other Considerations

4 Gas Delivery System Leakage Symptoms Steady increase No reduction when gas is turned off Analysis Use meter as a leak meter Solution Stop leak Place cylinder inside hood

5 Cylinder Inside the Hood

6 Reentry Tracer gas released inside the hood Characteristics Delayed onset Tracer gas in the room Often slow decay

7 Potential Reentry

8 Room Concentration

9 Repeat with Different Wind

10 Smoke Visualization ASHRAE 110 Failure Smoke spills out Diagnostic testing Lazy air Reverse air flow Interior roll

11 Reverse Flow Reverse flow

12 Lazy Flow Lazy Flow

13 Good sweeping flow Good Sweeping Action

14 Roll in top of hood Smoke Showing the Hood Roll

15 Summary Poor flow characteristic can pass but can contribute to poor performance Repeat somke tests with mannequin Vortex formed by mannequin interacts with the weak airflow in the hood Can show spillage where not seen without the mannequin

16 Equipment in the Hood

17 Other Issues Equipment in hood Sash Position Mandates other than left, right, center Difference between use and design Heat sources or disturbances in hood Work position Operator position

18 Room Air Supply The supply ventilation design is as important (or more important than) the exhaust ventilation design Rule of thumb: Room air currents in the laboratory near the hood should be less than half, preferably less than one third, of the face velocity

19 Room Air Currents Cross Draft Grid & Probe Orientation Top View Horizontal Draft Vertical Perpendicular Horizontal Left Center Right Vertical Draft Stand Perpendicular Draft High Midpoint Low Front View Side View

20 Minor Room Air Currents Hood 628, Position A, 52.5 inches 100 Velocity, fpm Vertical Perpendicular Horizontal Time, Seconds

21 Significant Room Air Currents Hood 628, Position AB, 52.5 inches 100 Velocity, fpm Vertical Perpendicular Horizontal Time, seconds

22 Case History New 12-foot bench top hoods Acceptance level AI 0.05 Preliminary tests not acceptable Diagnostic testing Reverse flow at work surface High Turbulence in laboratory High Cross drafts near hood Temperature variable in laboratory

23 Original Design

24 Approach Change diffusers to reduce room air turbulence Modify airfoil on laboratory hoods Modify room temperature control Recalibrate CV supply and exhaust boxes Rewire part of Building Management System Readout matched boxes

25 Diffuser Replacement

26 Original Diffuser and Hood Hood Rating 0.45

27 Change Diffuser Hood Rating 0.03

28 Changed Diffuser Modified Airfoil Hood Rating <0.01

29 Conclusions In many instances ASHRAE 110 can be used diagnostically to determine the probable cause of poor hood performance

30 Other Test Dynamic tests Human as a mannequin Artificial clutter Walk by tests Large apparatus (walk in hoods) Floor mounted fans

31 Philosophy Provides additional information relative to robustness Often no correlation to safety Interpretation is difficult

32 Caution Hoods are designed for specific strength Some hood manufacturers modifies the hood to provide strength in the top of the hood Reduce roll Airfoil that tends to lift the air up Some manufacturers provide a strong base Exceptional turning action by airfoil Piston flow in lower portion of the hood Many manufactures design for general flow

33 Manufacture Recommended Tests Higher release rate is better controlled by a hood strong at the top A lower diffuser is better controlled by a hood strong along the work surface Boxes and induced flow may work bit the geometry may be critical Some tests can be self serving

34 Conclusion Special tests are useful and provide additional information concerning the robustness of the hood Often suggested by user but have no comparative basis Hood manufactures recommend tests that favor their hoods

Lab Safety & Energy Optimization. Introduction & Agenda

Lab Safety & Energy Optimization. Introduction & Agenda Lab Safety & Energy Optimization Determining Appropriate Airflow Specifications for VAV Fume Hoods Thomas C. Smith tcsmith@labhoodpro.com Introduction & Agenda Thomas C. Smith President, BSME (NCSU), MSEE

More information

DEPAUL UNIVERSITY. Fume Hoods Manual. Environmental Health & Safety. April 2017

DEPAUL UNIVERSITY. Fume Hoods Manual. Environmental Health & Safety. April 2017 DEPAUL UNIVERSITY Fume Hoods Manual Environmental Health & Safety April 2017 1 TABLE OF CONTENTS SECTION PAGE NO. 1. INTRODUCTION 2 2. TYPES OF FUME HOODS 2 2.1 CONVENTIONAL FUME HOODS 2 2.2 HIGH PERFORMANCE

More information

Supreme AirLV- FUME HOODS. KEWAUNEE encouraging new discovery LOW CONSTANT VOLUME

Supreme AirLV- FUME HOODS. KEWAUNEE encouraging new discovery LOW CONSTANT VOLUME Supreme AirLV- LOW CONSTANT VOLUME FUME HOODS KEWAUNEE encouraging new discovery Supreme Air LV Fume Hood Low Velocity Low Volume Full Access Hold-Open Catch for set-up sash position Flush Mounted Convergent

More information

VOLU-probe/SS. Stainless Steel Pitot Airflow Traverse Probes

VOLU-probe/SS. Stainless Steel Pitot Airflow Traverse Probes Stainless Steel Pitot Airflow Traverse Probes The VOLU-probe/SS Stainless Steel Pitot Airflow Traverse Probe is ideally suited for new installations or retrofit applications requiring accurate airflow

More information

Fume Hood Standard. The appropriate maintenance/engineering division on each campus or other business unit operating on UofT property shall:

Fume Hood Standard. The appropriate maintenance/engineering division on each campus or other business unit operating on UofT property shall: Fume Hood Standard A fume hood is an enclosed working chamber fitted with an exhaust ventilation system. It is one component of an exposure control system which is designed to contain, dilute and disperse

More information

Design Standard - Fume Hoods & Fume Hood Exhausts

Design Standard - Fume Hoods & Fume Hood Exhausts Design Standard - Fume Hoods & Fume Hood Exhausts 1.0 Introduction The fume hood is the primary control device in most laboratories for protecting employees and students from exposure to hazardous chemicals.

More information

FUME HOOD SELECTION GUIDE

FUME HOOD SELECTION GUIDE FUME HOOD SELECTION GUIDE FUME HOOD SELECTION GUIDE Laboratory fume hoods are one of the most vital pieces of safety equipment in a working laboratory. Fume hoods are designed to safely exhaust dangerous

More information

USER'S GUIDE TO FUME HOODS. Procedures for Increasing the Effectiveness of Laboratory Hoods. Revised December, Hood Basics

USER'S GUIDE TO FUME HOODS. Procedures for Increasing the Effectiveness of Laboratory Hoods. Revised December, Hood Basics USER'S GUIDE TO FUME HOODS Procedures for Increasing the Effectiveness of Laboratory Hoods Revised December, 1998 A laboratory fume hood system is designed to protect the operator from undesirable substances

More information

Protector XStream Laboratory Hood is shown with SpillStopper Work Surface , Protector Acid Storage Cabinet and Protector

Protector XStream Laboratory Hood is shown with SpillStopper Work Surface , Protector Acid Storage Cabinet and Protector Protector XStream Laboratory Hood 9840600 is shown with SpillStopper Work Surface 9849800, Protector Acid Storage Cabinet 9901100 and Protector Standard Storage Cabinet 9900100. Protector XStream Laboratory

More information

Requires Ductwork and Blower

Requires Ductwork and Blower 6' Protector XStream Laboratory Hood 9840600 is shown with SpillStopper Work Surface 9849800, Protector Acid Storage Cabinet 9901100 and Protector Standard Storage Cabinet 9900100. Requires Ductwork and

More information

3.5 Exposure Control Measures

3.5 Exposure Control Measures 3.5 Exposure Control Measures 3.5.1 Chemical Fume Hoods Chemical fume hoods are the most reliable and primary engineering control used in the laboratory to protect against the inhalation of hazardous vapors

More information

How to Use a Chemical Fume Hood Safely

How to Use a Chemical Fume Hood Safely Published on UC Davis Safety Services (https://safetyservices.ucdavis.edu) How to Use a Chemical Fume Hood Safely SafetyNet #: 35 A chemical fume hood is designed to protect the worker from volatile and

More information

ASHRAE 110 Test Report Cover Sheet

ASHRAE 110 Test Report Cover Sheet SHRE 110 Test Report over Sheet lient: mgen orporation One mgen center rive Thousand Oaks 91320-3424 Ray Russell Test Report # : 1 Test ate # : 1/6/2005 Lead Technician: Joe Tech Test Report Scope and

More information

---:...-- Environmental Health & Safety RADIATION PROCEDURES MANUAL. Procedure Cover Sheet. Procedure Number: EHS REV 2

---:...-- Environmental Health & Safety RADIATION PROCEDURES MANUAL. Procedure Cover Sheet. Procedure Number: EHS REV 2 Environmental Health & Safety RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Fume Hoods Procedure Number: EHS-08-03-REV 2 Effective Date: 18 September, 2018 ---:...-- Approved By_-=--------::.--...l~---

More information

Introducing NJK Precision Airflow Measuring Stations

Introducing NJK Precision Airflow Measuring Stations Introducing Precision Airflow Measuring Stations A new and unique airflow measuring station that accurately measures airflows at very low flow rates and can operate in physically restrictive applications

More information

Spring Air Systems FN-B-MB-10/4.5-VE Wall-Mounted Canopy Exhaust Hood Performance Report

Spring Air Systems FN-B-MB-10/4.5-VE Wall-Mounted Canopy Exhaust Hood Performance Report Spring Air Systems FN-B-MB-10/4.5-VE Wall-Mounted Canopy Exhaust Hood Performance Report Application of ASTM Standard Test Method F1704-09 FSTC Report 501311275 Food Service Technology Center Prepared

More information

Table of Contents. Page 2 Page 2 Page 2 Page 2 Page 3 Page 3 Page 3 Page 3 Page 3 Page 4 Page 4 Page 4 Page 5 Page 5 Page 5

Table of Contents. Page 2 Page 2 Page 2 Page 2 Page 3 Page 3 Page 3 Page 3 Page 3 Page 4 Page 4 Page 4 Page 5 Page 5 Page 5 -02 - Location and Mounting: AirflowMonitoringStation Table of Contents #1 - Right Angle wo/turning Vanes Mounting #2 - Right Angle w/turning Vanes Mounting #3 - Right Angle w/rounded Elbow Mounting #4

More information

TABLE OF CONTENTS STANDARDS FOR THE DESIGN, CONSTRUCTION, MAINTENANCE AND USE OF LABORATORY FUME HOODS. Effective August 2010

TABLE OF CONTENTS STANDARDS FOR THE DESIGN, CONSTRUCTION, MAINTENANCE AND USE OF LABORATORY FUME HOODS. Effective August 2010 TABLE OF CONTENTS TABLE OF CONTENTS... 1 INTRODUCTION... 2 BACKGROUND... 2 FUME HOOD DESIGN AND CONSTRUCTION... 3 VARIABLE AIR VOLUME FUME HOODS... 3 RECIRCULATING HOODS... 3 SPECIAL LABORATORY FUME HOODS...

More information

SECTION LABORATORY FUME HOODS UNIFLOW LE SERIES

SECTION LABORATORY FUME HOODS UNIFLOW LE SERIES SECTION 11610.1 LABORATORY LE FUME HOODS UNIFLOW LE SERIES REVISED AUGUST 2, 2013 LABORATORY FUME HOODS UNIFLOW LE SERIES PART 1 GENERAL 1.00 SUMMARY A. Section Includes: 1. Laboratory Fume Hoods B. Related

More information

Daikin Steam Coils. Installation and Maintenance Manual IM 901. Types HI-F5, HI-F8, & E-F5. Group: Applied Air. Part Number: IM 901

Daikin Steam Coils. Installation and Maintenance Manual IM 901. Types HI-F5, HI-F8, & E-F5. Group: Applied Air. Part Number: IM 901 Installation and Maintenance Manual IM 901 Daikin Steam Coils Group: Applied Air Part Number: IM 901 Date: February 2008 Types HI-F5, HI-F8, & E-F5 2008 Daikin Applied Contents Introduction... 3 General

More information

SECTION LABORATORY FUME HOODS

SECTION LABORATORY FUME HOODS SECTION 115313 LABORATORY FUME HOODS PART 1 - GENERAL 1.1 SUMMARY A. This Section includes Laboratory Fume Hoods: B. Related Sections include the following: 1. [Division 12 Section Painted Steel Laboratory

More information

General Purpose Walk-In Fume Hood

General Purpose Walk-In Fume Hood General Purpose Walk-In Fume Hood with Vertical Rising Sash Available Models: Open By-Pass Restricted By-Pass Auxiliary Air Available Option: Adjustable Baffles Air Flow and Static Pressure Alarms Service

More information

2.6. Air Flow Control Valve Type PRD

2.6. Air Flow Control Valve Type PRD 2.6. Air Flow Control Valve Type PRD Page 1/10 Air Flow Control Valve, Type PRD Page 2/10 Air Flow Control Valve Type PRD Description and Design The PRD remains the air valve of choice for critical heating,

More information

Scope and Application of ASTM 1704, Standard Test Method for Capture and Containment Performance of Commercial Kitchen Exhaust Ventilation Systems

Scope and Application of ASTM 1704, Standard Test Method for Capture and Containment Performance of Commercial Kitchen Exhaust Ventilation Systems Scope and Application of ASTM 1704, Standard Test Method for Capture and Containment Performance of Commercial Kitchen Exhaust Ventilation Systems The capture and containment exhaust air flow rates for

More information

AUDIO-VISUAL EQUIPMENT

AUDIO-VISUAL EQUIPMENT 11 52 00 AUDIO-VISUAL EQUIPMENT 1. GENERAL A. Related sections: i. 00 00 13 Designing Learning Environments ii. 11 52 13 Projection Screens iii. 12 56 52 Audio-Visual Equipment iv. 72 41 00 General Audio-Visual

More information

Bench fume hood with horizontal sash - 14" working sash height. Bench fume hood with vertical sliding doors 14" working gap between doors

Bench fume hood with horizontal sash - 14 working sash height. Bench fume hood with vertical sliding doors 14 working gap between doors UNIVERSITY OF HOUSTON ENVIRONMENTAL HEALTH AND SAFETY POLICIES AND PROCEDURES SECTION: SUBJECT: ENVIRONMENTAL HEALTH AND SAFETY LABORATORY FUME HOODS I. PURPOSE These laboratory fume hood survey procedures

More information

General Purpose Walk-In Fume Hood

General Purpose Walk-In Fume Hood General Purpose Walk-In Fume Hood with Vertical Rising Sash Available Models: Open By-Pass Restricted By-Pass Auxiliary Air Available Option: Adjustable Baffles Air Flow and Static Pressure Alarms Service

More information

PennBarry Fan Guide Specification. Laboratory Exhaust System: Model vplume

PennBarry Fan Guide Specification. Laboratory Exhaust System: Model vplume PennBarry Fan Guide Specification Laboratory Exhaust System: Model vplume This document 2018 PennBarry vplume is a commercial lab exhaust fan that ensures a greater margin of safety by effectively dispersing

More information

Study on Flow Fields in Variable Area Nozzles for Radial Turbines

Study on Flow Fields in Variable Area Nozzles for Radial Turbines Vol. 4 No. 2 August 27 Study on Fields in Variable Area Nozzles for Radial Turbines TAMAKI Hideaki : Doctor of Engineering, P. E. Jp, Manager, Turbo Machinery Department, Product Development Center, Corporate

More information

Design Requirements The system shall be designed, manufactured, tested and installed in compliance with the current versions or issuance of:

Design Requirements The system shall be designed, manufactured, tested and installed in compliance with the current versions or issuance of: SECTION 15877 FUME HOODS 1. General 1.1. Summary These specifications provide a performance specification for a chemical laboratory Fume Hood (herein referred to as FH). The intent of the specification

More information

LFBC - LATERAL FLOW BIO CONTAINMENT FOR HPAPI PROCESSING FLOW SCIENCES, INC.

LFBC - LATERAL FLOW BIO CONTAINMENT FOR HPAPI PROCESSING FLOW SCIENCES, INC. LFBC - LATERAL FLOW BIO CONTAINMENT FOR HPAPI PROCESSING FLOW SCIENCES, INC. LFBC THE FLOW SCIENCES GLOVEBOX WORKSTATION SERIES safety. The Glovebox Workstation series provides provides containment containment

More information

STATE UNIVERSITY CONSTRUCTION FUND

STATE UNIVERSITY CONSTRUCTION FUND DIRECTIVE 15H-8 Issue date: June 2016 1. General Design Criteria LABORATORY DESIGN a. Design a Pressure Independent Variable Volume system that satisfies the airflow demand of the fume hoods, minimum air

More information

Gaylord 30-inch Tall ELX-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report

Gaylord 30-inch Tall ELX-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report Gaylord 30-inch Tall ELX-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report Application of ASTM Standard Method F1704-09 FSTC Report 5011.09.06-R1 Food Service Technology Center Prepared by:

More information

Fume cupboards and extraction devices

Fume cupboards and extraction devices Fume cupboards and extraction devices Energy efficiency, maximum ergonomics and a larger internal workspace make working with our new fume cupboards even safer and more convenient. A new design together

More information

Laboratory Fume Hood. Operation & Maintenance Manual APR 2005

Laboratory Fume Hood. Operation & Maintenance Manual APR 2005 Laboratory Fume Hood Operation & Maintenance Manual APR 2005 TABLE OF CONTENTS I. Introduction p. 3 II. Fume Hood Services p. 4 III. Operation Recommendations p. 4 IV. General Fume Hood Maintenance p.

More information

Laboratory Demonstration Space Located at Price Research Center North Winnipeg. Fume Hoods. ehpricesales.com

Laboratory Demonstration Space Located at Price Research Center North Winnipeg. Fume Hoods. ehpricesales.com Laboratory Demonstration Space Located at Price Research Center North Winnipeg Fume Hoods ehpricesales.com 2 www.ehpricesales.com FUME HOODS Contents Dynamic Bench Fume Hood Features High Performance Fume

More information

.1 Special Building Areas 8. Laboratories..3 Laboratory Casework (section ).4 Chemical storage cabinets (section )

.1 Special Building Areas 8. Laboratories..3 Laboratory Casework (section ).4 Chemical storage cabinets (section ) Part 1 General 1.1 Summary.1 Unless otherwise indicated, follow the standards below when specifying Laboratory Fume Hoods. These standards are not intended to restrict or replace professional judgment..2

More information

FVC2100/2200 TEK-AIR TECHNICAL PRODUCT DATA SHEET FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER. Application. General Description

FVC2100/2200 TEK-AIR TECHNICAL PRODUCT DATA SHEET FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER. Application. General Description TEK-AIR TECHNICAL PRODUCT DATA SHEET FVC2100/2200 FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER MODEL 2100: Constant face velocity, variable volume MODEL 2200: Constant face velocity, variable volume

More information

Ventilated Balance Enclosures

Ventilated Balance Enclosures Ventilated Balance Enclosures by PLAS n LABS Meets and exceeds ANSI/ASHRAE 110:2016 standard for containment & performance PLAS n LABS, INC. www.plas-labs.com PLI@PLAS-LABS.com 517-372-7177 Ventilated

More information

Air Sentry High Performance Fume Hood Operation & Maintenance Manual

Air Sentry High Performance Fume Hood Operation & Maintenance Manual Air Sentry High Performance Fume Hood Operation & Maintenance Manual Revision 8 TABLE OF CONTENTS I. Introduction p. 3 II. High Performance Fume Hoods p. 4 III. Vortex Control System p. 5 IV. Uncrating

More information

FOR USE IN SOLUTION AIR HANDLING UNITS

FOR USE IN SOLUTION AIR HANDLING UNITS AMS-60 AIRFLOW MONITORING DEVICE APPLICATION GUIDE Supersedes: NOTHING Form 102.20-AG1 (303) FOR USE IN SOLUTION AIR HANDLING UNITS GENERAL The YORK AMS-60 is an airflow monitoring device which combines

More information

Alarge portion of kitchen ventilation planning is dedicated to

Alarge portion of kitchen ventilation planning is dedicated to By Richard T. Swierczyna, Associate Member ASHRAE, & Paul A. Sobiski, Associate Member ASHRAE Alarge portion of kitchen ventilation planning is dedicated to properly exhausting cooking effluent. Appliance

More information

Element D Services Heating Ventilating and Air Conditioning ODG010107

Element D Services Heating Ventilating and Air Conditioning ODG010107 Element D Services Heating Ventilating and Air Conditioning ODG010107 D304202 Laboratory Exhaust and Ventilation Part 1. General 1. This section supplements Design Guideline Element D3042 on air handling

More information

Facts, Fun and Fallacies about Fin-less Model Rocket Design

Facts, Fun and Fallacies about Fin-less Model Rocket Design Facts, Fun and Fallacies about Fin-less Model Rocket Design Introduction Fin-less model rocket design has long been a subject of debate among rocketeers wishing to build and fly true scale models of space

More information

Pre-lab Questions: Please review chapters 19 and 20 of your textbook

Pre-lab Questions: Please review chapters 19 and 20 of your textbook Introduction Magnetism and electricity are closely related. Moving charges make magnetic fields. Wires carrying electrical current in a part of space where there is a magnetic field experience a force.

More information

Air Flow Factors. Alnor Model 6070, 6000, 2220, or 2220-A Probe Orientation

Air Flow Factors. Alnor Model 6070, 6000, 2220, or 2220-A Probe Orientation ir Flow Factors Model: TDC irflow Measurements Procedure 1. Record diffuser velocities with lnor probe in the locations and orientation shown the velocity Measurement Location Chart. Note: Each diagram

More information

January, 2014 Page 1 of 6

January, 2014 Page 1 of 6 Part 1 General 1.1 Summary.1 Unless otherwise indicated, follow the standards below when specifying chemical fume hood modifications or installations. These standards are not intended to restrict or replace

More information

Demand Based Static Pressure Reset Control for Laboratories

Demand Based Static Pressure Reset Control for Laboratories Accutrol, LLC Product Sheet Demand Based Static Pressure Reset Control for Laboratories Accutrol, LLC 21 Commerce Drive, Danbury, CT 06810 203-445-9991 www.accutrolllc.com Contents and specifications are

More information

Fume Hood Operation & Maintenance Manual

Fume Hood Operation & Maintenance Manual Fume Hood Operation & Maintenance Manual Fume Hood Operation & Maintenance Manual TABLE OF CONTENTS Fume Hood Applications 2 Safety 3 Aerodynamic Principles & Safety 4 Fume Hood Cleaning 7 Basic Maintenance

More information

Ground Effect and Turbulence Simulation at the Pininfarina Wind Tunnel. Giuseppe Carlino Aerodynamic and Aeroacoustic Research Center

Ground Effect and Turbulence Simulation at the Pininfarina Wind Tunnel. Giuseppe Carlino Aerodynamic and Aeroacoustic Research Center Ground Effect and Turbulence Simulation at the Pininfarina Wind Tunnel Giuseppe Carlino Aerodynamic and Aeroacoustic Research Center The Aerodynamic and Aeroacoustic Research Center The Full Scale Automotive

More information

Compressor Noise Control

Compressor Noise Control Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 1972 Compressor Noise Control G. M. Diehl Ingersoll-Rand Research Follow this and additional

More information

AUTOMOTIVE TESTING AND OPTIMIZATION. Tools for designing tomorrow's vehicles

AUTOMOTIVE TESTING AND OPTIMIZATION. Tools for designing tomorrow's vehicles AUTOMOTIVE TESTING AND OPTIMIZATION Tools for designing tomorrow's vehicles 2 Measurement of flow around the side mirror by Particle Image Velocimetry (PIV). Courtesy of Visteon Deutschland GmbH Our advanced

More information

Live Wire Owner s Manual

Live Wire Owner s Manual Live Wire Owner s Manual Battery Replacement Introduction Features Precautions Tool Description Purpose Using the Tool Model MB0414L, MB0414H www.origocorp.com Operating the Tool How it Works Quick Summary

More information

Energy Recovery Laboratory Exhaust

Energy Recovery Laboratory Exhaust Energy Recovery Laboratory Exhaust Model Vektor -ERS For Vektor-M, MH and MS Systems January 2018 Model Vektor -ERS Greenheck s Vektor -ERS is a pre-engineered laboratory exhaust energy recovery system.

More information

new fume hoods providing performance and protection

new fume hoods providing performance and protection new fume hoods providing performance and protection five questions to get us started 1. Will the use of your hood require a special application hood? 2. How high of a priority is energy conservation? 3.

More information

Faraday s Electromagnetic Lab

Faraday s Electromagnetic Lab Name Section Date CONCEPTUAL PHYSICS Electromagnetic Induction: Generators and Alternating Current Tech Lab Electromagnetism Sim Faraday s Electromagnetic Lab Purpose To manipulate simulated magnets, compasses,

More information

Fume Hoods Overview OVERVIEW

Fume Hoods Overview OVERVIEW OVERVIEW What Is A Fume Hood - A laboratory chemical fume hood is first and foremost a safety device. Users need to be able to rely on their fume hood as a primary containment device to protect them from

More information

The A4 Paper Test Page 1 of (C)

The A4 Paper Test Page 1 of (C) The A4 Paper Test Page 1 of 5 2010-10-25 (C) A4 Paper Test_R1.doc Version: R1 The A4 Paper Test Without access to suitable measuring equipment, the most effective method to correctly calibrate the oven

More information

Chapter 11: Flow over bodies. Lift and drag

Chapter 11: Flow over bodies. Lift and drag Chapter 11: Flow over bodies. Lift and drag Objectives Have an intuitive understanding of the various physical phenomena such as drag, friction and pressure drag, drag reduction, and lift. Calculate the

More information

Computer-Assisted Induction Aluminum

Computer-Assisted Induction Aluminum Home Computer-Assisted Induction Aluminum Brazing November 11, 2003 Coupled electromagnetic and thermal computer simulation provides a sufficient basis for process optimization and quality improvement

More information

Gaylord Company, Model EL-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report

Gaylord Company, Model EL-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report Gaylord Company, Model EL-ND-BDL-54 Wall-Mounted Canopy Exhaust Hood Performance Report Application of ASTM Standard Test Method F 1704-05 Food Service Technology Center (www.fishnick.com) -R1 Prepared

More information

LABORATORY FUME HOODS

LABORATORY FUME HOODS Fume Hoods Fume Hood Applications Laboratory professionals and students across the country, and all over the world, depend on the safety of our hoods. Your protection is our first concern and the fume

More information

Protector XStream Laboratory Hoods

Protector XStream Laboratory Hoods High Performance Hood offers enhanced containment combined with energy savings. With its unsurpassed containment and substantial energy savings potential, the patented* Protector XStream Fume Hood delivers

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION Wind tunnel investigation of waste air re-entry with wall ventilation P. Broas Technical Research Centre of Finland, Ship Laboratory, Tekniikantie 12, SF-02150, Espoo, Finland ABSTRACT A wind tunnel investigation

More information

A. Perform a vacuum gauge test to determine engine condition and performance.

A. Perform a vacuum gauge test to determine engine condition and performance. ENGINE REPAIR UNIT 2: ENGINE DIAGNOSIS, REMOVAL, AND INSTALLATION LESSON 2: ENGINE DIAGNOSTIC TESTS NOTE: Testing the engine s mechanical condition is required when the cause of a problem is not located

More information

Element D Services Heating, Ventilating, and Air Conditioning

Element D Services Heating, Ventilating, and Air Conditioning PART 1 - GENERAL 1.01 OVERVIEW A. This section supplements Design Guideline Element D3042 on exhaust and ventilation with specific criteria for projects involving design of Laboratory Exhaust and Systems.

More information

The Influence of the Phase Difference between the Crank Angle of the Pilot and that of the Stoker on the Drag Acting on a Tandem Bike

The Influence of the Phase Difference between the Crank Angle of the Pilot and that of the Stoker on the Drag Acting on a Tandem Bike Proceedings The Influence of the Phase Difference between the Crank Angle of the Pilot and that of the Stoker on the Drag Acting on a Tandem Bike Kazuya Seo * and Satoshi Eda Department of Education, Art

More information

UNIVERSITY OF ROCHESTER DESIGN STANDARDS JANUARY 2012 SECTION FUME HOODS AND LOCAL EXHAUST SYSTEMS

UNIVERSITY OF ROCHESTER DESIGN STANDARDS JANUARY 2012 SECTION FUME HOODS AND LOCAL EXHAUST SYSTEMS SECTION 15860 - FUME HOODS AND LOCAL EXHAUST SYSTEMS This standard describes fundamental good practices relating to the design, selection, installation, commissioning, and testing of local exhaust systems

More information

Engine Performance Troubleshooting Tree - Signature and ISX CM870

Engine Performance Troubleshooting Tree - Signature and ISX CM870 File: 70-t02-1001 Page 1 of 64 Engine Performance Troubleshooting Tree - Signature and ISX CM870 This troubleshooting procedure should be followed for the following symptoms: Engine Acceleration or Response

More information

Chapter 1: Introduction to Automobile Aerodynamics...1

Chapter 1: Introduction to Automobile Aerodynamics...1 Preface...xv Acknowledgments...xix Chapter 1: Introduction to Automobile Aerodynamics...1 1.1 Scope... 1 1.1.1 The Role of Aerodynamics in Vehicle Design... 1 1.1.2 The Character of Vehicle Aerodynamics...

More information

AUDIO-VISUAL EQUIPMENT

AUDIO-VISUAL EQUIPMENT 11 52 00 AUDIO-VISUAL EQUIPMENT 1. GENERAL A. Related sections: i. 00 00 13 Designing Learning Environments ii. 11 52 13 Projection Screens iii. 12 56 52 Audio-Visual Equipment iv. 72 41 00 General Audio-Visual

More information

Thermo Scientific Hamilton Horizon Fume Hoods. Operation, Maintenance and Installation Instructions

Thermo Scientific Hamilton Horizon Fume Hoods. Operation, Maintenance and Installation Instructions Thermo Scientific Hamilton Horizon Fume Hoods Operation, Maintenance and Installation Instructions Thermo Scientific Horizon Fume Hood Table of Contents Introduction.............................................................

More information

VOLU-probe. Pitot Airflow Traverse Probes. amca. Accurate airflow measurement for demanding applications. CERTIFIED RATInGS

VOLU-probe. Pitot Airflow Traverse Probes. amca. Accurate airflow measurement for demanding applications. CERTIFIED RATInGS amca CERTIFIED RATInGS VOLU-probe Pitot Airflow Traverse Probes FLOW MEASUREMENT STATION PERFORMANCE MOVEMENT AND CONTROL ASSOCIATION INTERNATIONAL, INC. Accurate airflow measurement for demanding applications

More information

Content - Silencers. Silencers for spiral systems. Silencers for rectangular systems

Content - Silencers. Silencers for spiral systems. Silencers for rectangular systems Silencer Catalog Content - Silencers General information on noise... 3 General information on sound attenuation... 4 Silencer specification... 7 2 Silencers for spiral systems Silencers for rectangular

More information

A. Each cabinet must be constructed and installed in accordance with EN12469:2000.

A. Each cabinet must be constructed and installed in accordance with EN12469:2000. BIOLOGICAL SAFETY CABINETS Class II, Biological Safety Cabinet PART 1 GENERAL 1.1 REFERENCES The publications listed below form a part of this section to the extent referenced. The publications are referred

More information

AD-1250 Airflow Measuring Station

AD-1250 Airflow Measuring Station AD-1250 Airflow Measuring Station Product Bulletin Code No. LIT-12011472 Issued March 2016 Refer to the QuickLIT website for the most up-to-date version of this document. The AD-1250 Airflow Measuring

More information

Protector XStream Laboratory Fume Hoods

Protector XStream Laboratory Fume Hoods User s Manual Protector XStream Laboratory Fume Hoods Models 9840400, 9840401, 9840402, 9840403, 9840500, 9840501, 9840502, 9840503, 9840600, 9840601, 9840602, 9840603, 9840800, 9840801, 9840802, 9840803

More information

General Purpose Permanent Magnet Motor Drive without Speed and Position Sensor

General Purpose Permanent Magnet Motor Drive without Speed and Position Sensor General Purpose Permanent Magnet Motor Drive without Speed and Position Sensor Jun Kang, PhD Yaskawa Electric America, Inc. 1. Power consumption by electric motors Fig.1 Yaskawa V1000 Drive and a PM motor

More information

Fan Efficiency Grades and System Effect and Their Effects on HVAC Systems

Fan Efficiency Grades and System Effect and Their Effects on HVAC Systems Fan Efficiency Grades and System Effect and Their Effects on HVAC Systems Matt Spink, P.E. Greenheck Fan Corporation Learning Objectives Understanding Fan Efficiency Grades (FEG) Selection of fans within

More information

BRAKE SYSTEM DESIGN AND THEORY

BRAKE SYSTEM DESIGN AND THEORY RAKE SYSTEM DESIGN AND THEORY Aircraft brake systems perform multiple functions. They must be able to hold the aircraft back at full static engine run-up, provide adequate control during ground taxi operations,

More information

Modeling, Structural & CFD Analysis and Optimization of UAV

Modeling, Structural & CFD Analysis and Optimization of UAV Modeling, Structural & CFD Analysis and Optimization of UAV Dr Lazaros Tsioraklidis Department of Unified Engineering InterFEA Engineering, Tantalou 7 Thessaloniki GREECE Next Generation tools for UAV

More information

SECTION DUCTWORK ACCESSORIES

SECTION DUCTWORK ACCESSORIES SECTION 15910 DUCTWORK ACCESSORIES PART 1- GENERAL SCOPE This sections includes accessories used in the installation of duct systems. Included are the following topics: PART 1- GENERAL Related Work Reference

More information

VIBRATION OF AUTOMOBILE SIDE VIEW MIRROR DUE TO AERODYNAMIC INPUTS

VIBRATION OF AUTOMOBILE SIDE VIEW MIRROR DUE TO AERODYNAMIC INPUTS Proceedings of the International Conference on Mechanical Engineering 25 (ICME25) 28-3 December 25, Dhaka, Bangladesh ICME5- VIBRATION OF AUTOMOBILE SIDE VIEW MIRROR DUE TO AERODYNAMIC INPUTS Rajneesh

More information

LABORATORY HOOD MANUAL

LABORATORY HOOD MANUAL LABORATORY HOOD MANUAL DRAKE UNIVERSITY ENVIRONMENTAL HEALTH & SAFETY November 2015 TABLE OF CONTENTS I. INTRODUCTION... 1 II. FUME HOOD SERVICE... 2 Fume Hood Service Provided by EH&S... 2 Fume Hood Service

More information

Aerodynamics and its application for vehicles / Járműáramlástan DIRECTIONAL STABILITY SIDE WIND EFFECT

Aerodynamics and its application for vehicles / Járműáramlástan DIRECTIONAL STABILITY SIDE WIND EFFECT Aerodynamics and its application for vehicles / Járműáramlástan DIRECTIONAL STABILITY SIDE WIND EFFECT Aerodynamics and its application for vehicles SIDE WIND EFFECT FORCES: LIFT force: L = c L (1/2rv

More information

Energy Recovery Laboratory Exhaust

Energy Recovery Laboratory Exhaust Energy Recovery Laboratory Exhaust Model Vektor ERS-M For Vektor-M Fan Systems September 2012 Model Page Header Vektor ERS-M Greenheck s Vektor -ERS is a pre-engineered laboratory exhaust energy recovery

More information

Boilers, Steam Generators, and Process Heaters (Oxides of Nitrogen) - Adopted 10/13/94, Amended 4/6/95, 7/10/97

Boilers, Steam Generators, and Process Heaters (Oxides of Nitrogen) - Adopted 10/13/94, Amended 4/6/95, 7/10/97 RULE 425.2 Boilers, Steam Generators, and Process Heaters (Oxides of Nitrogen) - Adopted 10/13/94, Amended 4/6/95, 7/10/97 I. Purpose The purpose of this Rule is to limit oxides of nitrogen (NOx) emissions

More information

Almost 200 years ago, Faraday looked for evidence that a magnetic field would induce an electric current with this apparatus:

Almost 200 years ago, Faraday looked for evidence that a magnetic field would induce an electric current with this apparatus: Chapter 21 Electromagnetic Induction and Faraday s Law Chapter 21 Induced EMF Faraday s Law of Induction; Lenz s Law EMF Induced in a Moving Conductor Changing Magnetic Flux Produces an E Field Inductance

More information

Part 1 of this series discussed optimizing outdoor

Part 1 of this series discussed optimizing outdoor Many methods await when it comes to reducing exhaust rates. Fortunately, engineers can avail themselves of multiple tactics, from the hood specification to proper commissioning and air balance. By Doug

More information

STUDY OF THE AERODYNAMIC NOISE CHARACTERISTICS OF BLUFF BODIES AS A PANTOGRAPH MEMBER

STUDY OF THE AERODYNAMIC NOISE CHARACTERISTICS OF BLUFF BODIES AS A PANTOGRAPH MEMBER STUDY OF THE AERODYNAMIC NOISE CHARACTERISTICS OF BLUFF BODIES AS A PANTOGRAPH MEMBER PACS REFERENCE: 43.5.Lj IKEDA Mitsuru Railway Technical Research Institute -8-38, Hikari-cho, Kokubunji-shi, Tokyo,

More information

A Study of EGR Stratification in an Engine Cylinder

A Study of EGR Stratification in an Engine Cylinder A Study of EGR Stratification in an Engine Cylinder Bassem Ramadan Kettering University ABSTRACT One strategy to decrease the amount of oxides of nitrogen formed and emitted from certain combustion devices,

More information

Pre-lab Questions: Please review chapters 19 and 20 of your textbook

Pre-lab Questions: Please review chapters 19 and 20 of your textbook Introduction Magnetism and electricity are closely related. Moving charges make magnetic fields. Wires carrying electrical current in a part of space where there is a magnetic field experience a force.

More information

INSTALLATION AND OPERATING INSTRUCTIONS FOR CONCRETE VAULT TYPE UNITS

INSTALLATION AND OPERATING INSTRUCTIONS FOR CONCRETE VAULT TYPE UNITS INSTALLATION AND OPERATING INSTRUCTIONS FOR CONCRETE VAULT TYPE UNITS CONTENTS Introduction Safety Operating Principle System Installation Initial Startup System Operations Troubleshooting Maintenance

More information

NJK Precision Product NJK-02 Sensor Installation Guide

NJK Precision Product NJK-02 Sensor Installation Guide Precision -02 Sensor 2017 Product Line Precision Product Table of Contents Precision Company Overview - Page 2-02 Installation Guidelines - Page 3-02 Recommended Installations - Page 5-02 Non-Recommended

More information

RULE BOILERS, STEAM GENERATORS, AND PROCESS HEATERS (Adopted 5/11/93, Revised 6/13/95, 6/13/00, 9/11/12)

RULE BOILERS, STEAM GENERATORS, AND PROCESS HEATERS (Adopted 5/11/93, Revised 6/13/95, 6/13/00, 9/11/12) VENTURA COUNTY AIR POLLUTION CONTROL DISTRICT RULE 74.15.1 - BOILERS, STEAM GENERATORS, AND PROCESS HEATERS (Adopted 5/11/93, Revised 6/13/95, 6/13/00, 9/11/12) A. Applicability The provisions of this

More information

Special edition paper

Special edition paper Development of a Brake System for Shinkansen Speed Increase Hiroshi Arai* Satoru Kanno* Kenji Fujino* Hiroyuki Kato* Koji Asano* In efforts to increase Shinkansen speeds toward a 360 km/h operating speed,

More information

HYDRAULIC ACTUATOR REPLACEMENT USING ELECTROMECHANICAL TECHNOLOGY

HYDRAULIC ACTUATOR REPLACEMENT USING ELECTROMECHANICAL TECHNOLOGY HYDRAULIC ACTUATOR REPLACEMENT USING ELECTROMECHANICAL TECHNOLOGY SCOPE This white paper discusses several issues encountered by Lee Air with past projects that involved the replacement of Hydraulic Actuators

More information

User s Manual. LabShield. Laboratory Fume Hoods. Air Foil By-Pass Models: 6714, 6716, ADA Models: 8796, 8798, Demonstration Hood 6725G

User s Manual. LabShield. Laboratory Fume Hoods. Air Foil By-Pass Models: 6714, 6716, ADA Models: 8796, 8798, Demonstration Hood 6725G User s Manual LabShield By CampbellRhea Laboratory Fume Hoods Air Foil By-Pass Models: 6714, 6716, 6718 ADA Models: 8796, 8798, 8800 Demonstration Hood 6725G CampbellRhea 1865 North Market Street Paris

More information

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES FLS F6.60 MAGMETER FLOW SENSOR The new F6.60 and F6.63 are flow meters without moving mechanical parts which can be applied for the measurement of dirty liquids so long as they are conductive and homogeneous.

More information

Application Guidelines

Application Guidelines Grilles and Registers Application Guidelines Introduction Overview It is the intent of this section to give some basic application guidelines for the use of supply and return grilles and registers in a

More information