BAYLAN AUTOMATED TEST BENCH
|
|
- Thomas Curtis
- 5 years ago
- Views:
Transcription
1 BAYLAN AUTOMATED TEST BENCH At this test bench, the electromagnetic flow meters are used with the measurement method. Calibrations of flowmeters are made with the help of the water collection tank and a calibrated scale. Amount of the water, which passes inside of watermeters, is measured with help of electromagnetic flowmeters. This data and data of optical sensors on the watermeters are used to calculate errors of watermeters. System can be worked by software, which is installed to computer by BAYLAN, except movement of the pistons. Its movement is controlled by buttons on the panel. Tests can be performed for all watermeters between DN15-DN40 at any flowrate. The table, which is below, shows how many watermeter can be connected to the test bench for different lengths, diameters, Qn and measuring flow rates. Length (mm) Piece Qn (m³/h) DN , , , , , TEST BENCH PARTS The main structure of test bench is stainless steel. Steel frame is coated with epoxy paint. On the test bench, an air conditioner is used to regulate pressure. Connection pipes are made of POM, impact and shatter resistant. Picture 1
2 Information about Picture 1 1- Pneumatic water supply valve 2- Water inlet valves 3- Panel buttons 4- Computer monitor 5- Water collection tank 6- Electromagnetic flowmeters 7- Water flow control valves Picture 2 Information about Picture 2 1- Pressure piston for the static pressure test 2- Optical sensor connector 2. Scale Maximum capacity of the scale is 300 kg. It has Five-digit LED displays and 2g sensitivity. It works at 220 VAC and 50 Hz (it is not affected by voltage dropping until 170 VAC).
3 Picture 3 3. Manometer Working pressure is between 0 and 25 bars. Moving part is made of brass. Pressure element is made of CuSn8. Storages material is steel and front rings material is abs plastic. Manometer connection is R 1/8"brass. Accuracy is 2,5%. Operating temperature range is from -25 C to +60 C. 4. Flowmeters Picture 4 There are 3 electromagnetic flowmeters or different flowrates. First flowmeter is for l/s. Second floweter is for l/s. Third flowmeter is for l/s.
4 Picture 5 Information about Picture-5 1- Electromagnetic flowmeter (Qmax-Qn) 2- Electromagnetic flowmeter (Qt) 3- Electromagnetic flowmeter (Qmin) 4- Emergency stop button 5- Manometer 6- Valve for regulating flowrate (Qmax-Qn) 7- Valve for regulating flowrate (Qt) 8- Valve for regulating flowrate (Qmin) Picture 6
5 Information about Picture-6 1- Electromagnetic flowmeter measuring part. 5. Buttons Information about Picture-7 Picture 7 1- Compression buttons (piston): Press two yellow buttons at the same time to compress watermeters to test bench. 2- Test bench is active: Lights is on while the test board is active. 3- Release button (piston): Make sure that the water pressure is zero before releasing piston in the system. 4- Watermeters are compressed. 5- Emergency stop button. 6. Sensor Sensors are 1. class accuracy and they shows excellent performance in measuring range. They have the highest detection performance between 20 and 50mm. There are 8 pieces of graphic display led at the sensor to program easily and monitor its working. There is 1. class detection visible red light, wavelength 650 nm on the sensor. It works from 10 to 30 V DC voltages, and its maximum current is 35 ma (electric charge). It is protected from reverse polarity, high voltage and fluctuations. Display screen (8-segment) shows signal strength, reference point of laser. Green LED: Power ON Yellow LED: Output transmission Connect watermeter to test bench for programming of the sensor. Provide low flowrate (like 90l/h). Press (+) button on the sensor until 7. and 8. LEDs are on. If one LED is on, programming is accepted. If 1, 3, 6 and 8 light is on, programming is not accepted. In this case, procedure is repeated. After completed programming, all the light will be on by one by from left to right and right to left.
6 Programming process is done once, there is no need to program for each test. For correct placement of sensor, flowrate must be suitable. If sensors are placed perfectly, light moves from right to left and from left to right until last light. However, since light passes two light from middle line, reading is sufficient. 7. Test procedure Picture 8 Connect apparatus. its nominal diameter must be equal to watermeters nominal diameter. Check o-rings which are at the connection regions. Connect watermeters to test bench by paying attention to direction of flow and watermeters. Space can be completed between additional pipes. Outlet and inlet of pipes' lengths must be certain length. Press two yellow buttons at the same time (Picture-7) and tighten the watermeters (the green light will be on). Check all connections and make sure that there is no space connection. There are two different water inlet valve (Picture-1) for different types of pressure for water supplies. Do not open both valve at the same time. Otherwise, the water with high pressure will flow to low pressure line. Open only one of them. First of all click "AIR RELEASE" button by mouse or touch screen (Picture-10) to release water. Check pipes, connections and watermeters for leak. If there is leakage, click "Close Valve" (Picture-10) or emergency stop (Picture-5 and 7). If there is no leakage, place sensor. To begin test, press "TEST" button on the screen (Picture-10). After the pressing "TEST" button, the screen will display the watermeter types and flowrates to select (Picture-11). First of all, select watermeter types then flowrate and at the end click "LOAD" button for loading datas. To change datas, press "EDIT" button. Press "START TEST" to start test (Picture-13). At the end of the check manometer. Pressure must be zero to release piston. Adjust flowrate for reading manually. Select type of watermeter and flowrate for reading from sensor. Amount of water can be checked from flowmeter, computer or weighing
7 machine. Close valves after desired amount of water. The following table shows Amount of water for Qmax, Qt, Qmin according to their classes. Qn (l/h) Qmax (l/h) Qmin=0,02Qn (l/h) Amount of Water B Class Qt=0,08Qn (l/h) Qmin=0,01Qn (l/h) C Class Qt=0,0,15Qn (l/h) 1500 (60 kg) 3000 (80 kg) 30 (6 kg) 120 (15 kg) 15 (6 kg) 22,5 (15 kg) 2500 (70 kg) 5000 (100 kg) 50 (6 kg) 200 (15 kg) 25 (6 kg) 37,5 (15 kg) 3500 (100 kg) 7000 (100 kg) 70 (10 kg) 280 (25 kg) 35 (6 kg) 52,5 (15 kg) Read last index of watermeter for manuel test and read actual amount of water from electromagnetic flowmeter. Calibrate watermeters by adjustment screw according to error values of watermeters. After calibrations of the watermeters are done, error values of Qn, Qmax, Qmin, Qt must be determined. After calibration and test of the watermeters are done, click "CLOSE VALVE". Drop pressure to zero. Release piston to remove watermeters. Note: For Accuracy final test results, test conditions must be compatible with the following conditions. Relative humidity: 45% - 75% from Ambient atmospheric pressure: 86 kpa kpa (0.86 bar bar) Ambient temperature: 15 C - 25 C The test panel and the surrounding area should not be dusty and dirty. During each test, the temperature and relative humidity shouldn't change more than 5 C or 10%. Information about Picture Meter connection pipes 2 - Optical Sensors Picture 9
8 Picture 10 Picture 11
9 Picture Static Pressure Test Picture 13 There is a pressure piston for static pressure test. The pneumatic piston, is controlled by controller. Before the test, water supply is opened and water flows in the system until no air remains. After the
10 evacuation of the air, valves are closed. Water supply valve is closed (Picture-1) and pneumatic pressure of system is increased by controller. At the same time increasing at the manometer can be observed. After reaching the desired pressure in the system, operator can check watermeters. When time is completed, release pressure piston. Before releasing watermeter piston, check manometer. Picture -14 Picture Software of BAYLAN Test Bench Baylan Test Benches Software are developed by R&D department of Baylan. Watermeters are automatically tested by this software.
11 Saves the test results Charts and draws the error curve Checks the test panel automatically 10. Menus Setup Menu The serial port settings and user options can be set from "Settings" menu Serial Port The serial port parameters should be as follows: Baud Rate: 9600 Data Bits: 8 Stop Bits: 1 Parity: None Flow Control: None In this menu you can read records data of communication of the test bench.
12 Users Each user can enter the program with its own password. User can process according to authority of him. Only "admin" user has the authority to save new user. "Normal" user is only authorized to change their own password Report Menu The test reports can be read after recorded test results. All records are listed if any box is not checked.
13 Error curve graph can be drawn after selection of test and sensor number. After selecting test and sensor, preview and the printing process can be done by "Print" button Commands 3. Open Tank Command The drain valve opens. 4. Close Tank Command The drain valve closes. 5. Release Air Command All valves are open. 6. Close Valve Command The valve closes. 7.Refresh Command To check latest status of the test bench. Incoming information is displayed on the screen. 11. Test Menu At the entering test menu, "Tank Open?" prompt will be seen. For calibration test, tank must be closed.
14 There are 3 submenus at the type menus Type of watermeter Test of watermeter Test point of watermeter Type of watermeter: Information about characteristic of watermeter Type Code Class Water Pulse/Liter Qn : Shows name and code of watermeter. : Shows class of watermeter. : Shows type of wateremter (Cold-Warm). : Shows puls number of watermeter.(tour of pointer per liter). : Shows flowrate at the Qn. Watermeter Testing: Test Code : Specifies the test name or code. Water : Shows type of watermeter (hot or cold type). Test Type : Specifies the type of test. There is only one type of test "Start Stop" test. Water Meter Test Points: Shows the parameters of the selected test point. Point Type : Type of flow indicates that the test be done.
15 Point Flow : Shows flowrate of the test. Test Duration : Test duration in seconds. Volume : The volume of water will pass from watermeter during test. It is calculated automatically multiplying the test time and flowrate. Flowmeter : Shows the flowrate to be selected according to the test point. Constant Flowmeters :Flow meter pulse would indicate how many of each per liter. Max Flow : the maximum allowable flowrate indicates at the test point. Flow Min : the minimum allowable flow rate indicates at the test point. Max Error Rate : The maximum allowable error rate indicates at the test point. Minimum Error Rate : the minimum allowable error rate indicates at the test point. New Button Type Code, Test Code and Test Point menus, whichever is selected, the selected section is shown in blue. Select one of them and click "NEW" to enter data which menu you choose. After adding parameters, click OK.
16 Edit Button To modify data, click data whichever you want to change then click EDIT. Delete Button To delete data, click data whichever you want to delete then click DELETE. Load Button To start test, click LOAD button to load data of your test. 12. Test Menu Click "Test Start" to start test. Datas about weight and temperature will be seen on the screen. To finish test, click "Test Finish" button and test result so far will be seen on the screen. To cancel the test click "Test Cancel" After test finishes, pals and errors will be seen on the screen.
17 12.1. Ok Button After test finishes, to save error values, click Ok button New Test Button
18 Click "AIR RELEASE" button to adjust flowrate. Click "New Test" button then the selected meter type and test type test points appears. Select the test which you want then click "LOAD" button to start test again Save Button To save data, click "SAVE" button. By that way, test number is increased. If operator don't click "SAVE" button, test number will be same Serial No Button To add serial number to test results, click SERIAL NO button Calibration Test Test menu and calibration test are same. At the calibration test, values of flowmeters and weigthingmachine are compared. Flowmeter calibration can be done according to calibration value. 13. Parameters These parameters are very important for the testing process. Pulse liter : The number of pulses produced when 1 liter of water passes Test Duration : Test duration in seconds. Qn : Flowrate The first weight : The weight of tank at the begining of the test First heat : The water temperature at the begining of the test First Pals : 0 Last weight : The weight of tank at the end of the test
19 Last temperature : The water temperature at the end of the test Last pulse : The number of pulses produced by the watermeter during the test process Net Weight : The weight difference between the first and the last weight. Avg. Temperature : The average temperature of the first and last temperature. Density : The density of water according to the average temperature. Constant Flowmeter : The number of flowmeters pulses. Error : The error between the value of the amount of volume as measured by the watermeter and value of the real volume. OK : The error is not within the accepted limits, it is zero otherwise one. Weight and density information is only used in the calibration process. Volume information is obtained by flowmeters. Test Procedure Step 1: Adjust (COM) port settings from the "Settings Menu". Please skip this step after making the settings once. 2. Step Two: Click "Test Menu" button and select the parameters. Skip this step if you are already entered these parameters. Step 3: Enter "Test Menu". Select the type of watermeter test type and test points. Step 4: Click "Upload" button Step 5: Click "Test Start" button. Step 6: After finishing the test click "OK" button. If you want to do another test, click Air discharge then adjust flowrate and click "New Test" button. Return to step 4. Step 7: After all the tests, click "Save" button to save the test results.
20 Calibration of flowmeter by weightingmachine Calibration test result is used for calibrate GKL number of flowmeter. For calibrating GKL number: Press Button 1 for 2.5 sec. then release button. Select Quick Setup by Button 1 Move to Process input by Button 3 or Button 4 Select Process input by Button 1 Move to GKL Number by Button 3 or Button 4 Select GKL by Button 1 Change GKL by Button 1, Button 3 and Button 4 Ex: Calibration value is from calibration test and GKL number is GKL number + [(GKL number x calibration number)/100]=new GKL number [(1000 x -2) / 100] = 980 Due to continuous development of our products, we reserve the right to modify our product design or construction without prior notice.
Liquid Turbine Flow Meter OwnerÊs Manual
Liquid Turbine Flow Meter OwnerÊs Manual Manual LWGY-G Rev.1.0 Content 1.0 GENERAL INFORMATION... 1 2.0 SPECIFICATIONS... 2 3.0 OPERATION CONDITIONS... 3 4.0 CAUTIONS FOR INSTALLATION... 4 5.0 DIMENSION...
More informationOptions guide. Electromagnetic flowmeters. M9xx
Options guide Electromagnetic flowmeters M9xx Ordering codes Model Description M 910 -V0 0 0 0 DN50 PN16 Example Sensor M Basic Ex Ex certified (remote version only) Transmitter 910E Economic version 910
More informationDomestic Water Meter. Elin Wasserwerkstechnik. Contact Water Meter MTM for cold water. Features
Domestic Water Meter Elin Wasserwerkstechnik Contact Water Meter MTM for cold water Features Nominal diameter DN15 to DN50 Cold water up to 30 C (safe up to 50 C) Operating pressure up to 16 bar Control
More informationModel TUR-200D Turbine Flowmeter
General Specifications Model TUR-200D Turbine Flowmeter Liquid flows through the turbine housing causing an internal rotor to spin. As the rotor spins, an electrical signal is generated in the pickup coil.
More informationECONOMISER SERIES E2T USER MANUAL
TURBO S.R.L. Electronic Control Systems for Dust Collectors e-mail: info@turbocontrols.it web: www.turbocontrols.it TEL. ++39 (0)362 574024 FAX ++39 (0)362 574092 ECONOMISER SERIES E2T USER MANUAL 24/06/2014
More informationMagFlux Q 8200 Series
MagFlux Q 8200 Series 3.06 ELECTROMAGNETIC FLOW METER General MagFlux Q Electromagnetic Flow Meter, created in composite materials, designed with an optimized construction which secures optimal performance.
More informationCLA-VAL e-drive-34. User Manual. Motorised Pilots. CLA-VAL Europe LIN072UE - 04/16
User Manual CLA-VAL Europe www.cla-val.ch cla-val@cla-val.ch 1 - LIN072UE - 04/16 Table of Contents 1 Introduction... 3 1.1 Precautions Before Starting... 3 1.2 Troubleshooting... 3 1.3 General Disclaimer...
More informationHYDRUS ULTRASONIC METER
APPLICATION Static ultrasonic water meter for accurate measuring and recording for all applications of water supply. FEATURES 4 Real data communication, open metering telegram 4 Long-term stability under
More informationVantage 6000 Portable Ultrasonic Flow Meter
Vantage 6000 Portable Ultrasonic Flow Meter Instruction Manual Eastech Badger Flow Technology Group 4250 S. 76 th East Ave. Tulsa OK 74145 800-226-3569 Description The Vantage 6000 portable transit time
More informationAccurate measurement of compressed air consumption and detection of leaks. Measuring the individual consumption per customer / cost centre
Inline Flowmeter for compressed air and gases DN15 - DN50 / 16 bar The flow meter of the series, based on the measurement principle of thermal mass flow, is ideally suited for the measurement of flow of
More informationSonoMeter 30 Energy Meters
SonoMeter 30 Energy Meters Description MID examination certificate no.: LT-1621-MI004-020 The Danfoss SonoMeter 30 is a range of ultrasonic, compact energy meters intended for measuring energy consumption
More informationPaddle-wheel flow controller with optical principle for On/Off control
Paddle-wheel flow controller with optical principle for On/Off control Type 8039 can be combined with... Indication, monitoring, transmitting and On/Off control in one device. Programmable s (transistor
More informationIBM UPS Extend Run Battery Pack Operation and Setup Guide
IBM UPS Extend Run Battery Pack Operation and Setup Guide 02R2717 1: INSTALLATION Introduction The IBM Uninterruptible Power Supply (UPS) Extend Run Battery Pack provides extended load protection and
More informationDominion PX. Quick Setup Guide. Before You Begin. Mounting Zero U Models Using L-Bracket. Zero U Size. 1U and 2U Size
Dominion PX Quick Setup Guide Thank you for purchasing the Dominion PX intelligent power distribution unit (PDU). The intended use of the Raritan Dominion PX is distribution of power to information technology
More informationStatic Balancing Valve
Rev.A TJL...Series DN15~DN50 TJF...Series DN50~DN500 Static Balancing Valve Features Introduction Digital Handwheel Accurate scale on digital handwheel which could accurate to 0.1 cycle. The humanized
More informationArndt & Voß GmbH Elektronik - Meßtechnik
torque measuring unit Contents: Page 1. Functions 1 2. Components 2 3. Connection and control elements 3 4. Operating Mode 4 Torque measuring with bar graph 5 Single value chart 6 5. Selection of Setup
More informationBERMAD Waterworks. Insertion Flow Meter Device (IFM)
IFM-MUT1222PRV BERMAD's IFM-MUT1222-PRV SENSOR is designed for measuring the flow rate of valves sized DN80 to DN600 (3-24 ). This document describes the installation process of the sensor. For larger
More informationSHARKY 775 COMPACT ENERGY METER ULTRASONIC
APPLICATION The ultrasonic compact energy meter can be used for measuring the energy consumption in heating / cooling application for billing purposes. FEATURES 4 Approval for ultrasonic meter with dynamic
More informationElectronic Limit Switch Type
Electronic Limit Switch Type 3738-20 with optional solenoid valve for on/off rotary actuators Application Electronic limit switch for on/off applications to indicate the end position of rotary actuators.
More informationPRODUCT / TEST MANUAL 2L9K2 BATTERY EARTH FAULT RELAY
Sheet 1 of 5 Order Number Serial Number PRODUCT / TEST MANUAL 2L9K2 BATTERY EARTH FAULT RELAY Issue Date Level B 27/05/1999 Initial issue. Summary of changes Due to RMS continuous product improvement policy
More informationDSL-240 MkIII Single Pass Dust Monitor Measures mg/m 3 using DDP
Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions
More informationFluidic Oscillation Compact Heat Meter
Fluidic Oscillation Compact Heat Meter Application The Superstatic 749 is an autonomous compact thermal energy meter consisting of a flow meter a detachable integrator with a wide range of communications
More informationISTEC MAGNETIC METERS
ISTEC MAGNETIC METERS MEASURE ANY CONDUCTIVE LIQUIDS OR SLURRIES (EVEN ABRASIVE OR CORROSIVE) WITH THE NEW MICROPROCESSOR-BASED MAGNETIC METER -- A QUALITY PRODUCT - SIMPLE - RELIABLE - AT AN AFFORDABLE
More informationSolid State Electronic Pressure Switch with Integral LED Display Model PSD-10
Solid State Electronic Pressure Switch with Integral LED Display Model PSD-10 Electronic Pressure Measurement WIKA Datasheet PSD-10 Applications Hydraulics and pneumatics Filter monitoring Pump control
More informationEG DYNAMIC user manual
Timing Advance Processor EG DYNAMIC user manual ver. 1.1.0 dated 2012-10-01 This instruction can be also downloaded from: http://www.europegas.pl/en/technical-support/service-manuals Latest software version
More informationGuangzhou Junliye Import&Export Co., Ltd. CW-6H Injector Diagnostic & Cleaning System. Operation manual
CW-6H Injector Diagnostic & Cleaning System Operation manual OPERATING MANUAL INDEX 1. Technical Characteristics 2 2. Installation.3 3. Equipment Description..4 4. Control Board.6 5. Operation.9 I. MANUAL
More informationParker Hannifin GmbH & Co. KG Tube Fittings Division Europe Am Metallwerk 9, Bielefeld Phone Fax
Parker Hannifin GmbH & Co. KG Tube Fittings Division Europe Am Metallwerk 9, 33659 Bielefeld Phone ++49-521-4048-0 Fax ++49-521-4048-4280 Operating Instruction ServiceJunior SensoControl Please read carefully
More informationStainless steel(compensation) DN25, 40, Flange Standard GB/T Others DIN,ANSI,JIS Thread Standard BSPP(male) Others BSPP(Female),NPT.,etc. A
SGW GAS TURBINE FLOW METER Overview SGW gas turbine flow meter which assimilates the advanced technology of flow meters from home and abroad with optimum design, is a new generation of gas flow meter with
More informationTROUBLESHOOTING TP. Index
TROUBLESHOOTING TP Index E1 POWER CUT DETECTED... 2 E2 LOCK ERROR... 3 E3 DRAINAGE FAILURE / WATER IN TUB... 6 E4 MAXIMUM WATER LEVEL REACHED... 6 E5 FAULT WATER INLET... 6 E6 HEATING FAULT... 6 E7 MAXIMUM
More informationCompact Static Heat Meter of High-Tech Composite
Compact Static Heat Meter of High-Tech Composite Application The Superstatic 789 is a lightweight and robust compact heat meter consisting of a high-tech composite flow meter, a detachable integrator with
More informationEW130 Series Multijet Water Meters DN15 40 FOR COLD AND WARM POTABLE WATER APPLICATIONS
EW10 Series Multijet Water Meters DN15 0 FOR COLD AND WARM POTABLE WATER APPLICATIONS CONTENTS PRODUCT DATA Fig. 1. EW100BM / EW101BM Contents... 1 General... 2 Application... 2 Features... 2 Design...
More informationDigital Scale. Revision 1.0 August 17, Contents subject to change without notice.
Digital Scale Revision 1.0 August 17, 2000 Contents subject to change without notice. Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont, MN 56031 Tel (800) 637-0529 Tel (507) 238-8702 Fax
More informationSonoMeter 30 Energy Meters
SonoMeter 30 Energy Meters Description The Danfoss SonoMeter 30 is a range of ultrasonic, compact energy meters intended for measuring energy consumption in heating and cooling applications for billing
More informationBRAKE TESTER DECELEROMETER
OC3010_GBM_21009 BRAKE TESTER DECELEROMETER OWNER S MANUAL Version 8.++ ORBIT CONTROLS AG Zürcherstrasse 137 CH-8952 Schlieren/ZH Tel: + 41 44 730 2753 Fax: + 41 44 730 2783 info@orbitcontrols.ch www.orbitcontrols.ch
More informationVersaFlow Vortex 100 Vortex Flow Meter Specifications
VersaFlow Vortex 100 Vortex Flow Meter Specifications 34-VF-03-05 May 2011 The All-In-One Solution The VERSAFLOW is the only vortex flowmeter with integrated and temperature compensation in 2- wire technology.
More informationAgXcel GX5 Install Guide for Integration into Trimble. where precision meets the soil... PO Box 1611 Kearney, NE
where precision meets the soil... AgXcel GX5 Install Guide for Integration into Trimble PO Box 1611 Kearney, NE 68848 877.218.1981 www.agxcel.com where precision meets the soil... AgXcel Integration into
More informationPaddle-wheel flow controller for On/Off control
Paddle-wheel flow controller for On/Off control Type 803 can be combined with... Indication, monitoring, transmitting and On/Off control in one device Programmable s (transistor or relay) Automatic-calibration:
More informationInstruction of connection and programming of the VECTOR controller
Instruction of connection and programming of the VECTOR controller 1. Connection of wiring 1.1.VECTOR Connection diagram Fig. 1 VECTOR Diagram of connection to the vehicle wiring. 1.2.Connection of wiring
More informationHYDRUS ULTRASONIC METER
HYDRUS APPLICATION Static ultrasonic water meter for accurate measuring and recording for all applications of water supply. FEATURES 4 Ultrasonic water meter with long-term stability under difficult conditions
More informationUltrasonic heat meter
Ultrasonic heat meter Installation and user guide Flow sensor-vmc-p Energy integrator Compact heat meter Temperature senor CONTENTS Kapitel Please read this manual prior to installation Page 3 2! T! Important
More informationDynojet Research, Inc. All Rights Reserved. Optical RPM Sensor Installation Guide.
1993-2001 Dynojet Research, Inc. All Rights Reserved.. This manual is copyrighted by Dynojet Research, Inc., hereafter referred to as Dynojet, and all rights are reserved. This manual, as well as the software
More informationEx PROFICIENCY TESTING PROGRAM
Procedure Instruction i (1) Contents 1. General performance... 1 2. Procedure... 2 2.1 Functional test of Test Box "i"... 5 2.2 Calibration... 6 2.3 Contact resistance according to IEC 60079-11:2007...
More informationIssue 2.0 December EPAS Midi User Manual EPAS35
Issue 2.0 December 2017 EPAS Midi EPAS35 CONTENTS 1 Introduction 4 1.1 What is EPAS Desktop Pro? 4 1.2 About This Manual 4 1.3 Typographical Conventions 5 1.4 Getting Technical Support 5 2 Getting Started
More informationIntegrated Flow Control Module for Slurries and Chemicals LFC-7000
Integrated Flow Control Module for Slurries and Chemicals LFC-7000 Operating Instructions and Quick-Start Guide TABLE OF CONTENTS Introduction 3 Storage and Handling 3 Installation Instructions 4 Operating
More informationFlash Point Process Analyzer
Norms and standards ASTM D 56 ASTM D 93; DIN EN ISO 2719; IP 34 DIN EN ISO 13736; IP170 DIN 51755 Application The BARTEC BENKE Flash Point Process Analyzer (FPA-4) is a continuously measuring analyzer,
More informationEW447-EW452 Series Mechanical Heat Meters FOR HEATING AND COOLING APPLICATIONS
EW447-EW452 Series Mechanical Heat Meters FOR HEATING AND COOLING APPLICATIONS Application PRODUCT DATA Static compact heat meter with electronic measurement, consisting of electronic energy integrator
More informationModel Pressure range ON/OFF output Linear output
Digital Sensor Sensor with Easy-to-Read LED Display status can be checked at a glance from the red digital pressure value and analog bar displays. Measurement pressure is averaged by the chattering prevention
More informationThe Multivariable Air Flow Meter
The Multivariable Air Flow Meter Model MVC10A/MVC10F OVERVIEW The Multivariable Mass Flow Meter model MVC10A/ MVC10F, a compact has all function necessary to measure nitrogen (N 2 ) gas, carbon dioxcide
More informationSPG-DIGI. Operating Instruction. Please read carefully before use!
WALTER STAUFFENBERG GMBH & CO.KG Im Ehrenfeld 4 D-58791 Werdohl Postfach 1745 D-58777 Werdohl Germany Tel.: +49 (0) 2392 / 916-0 Fax: +49 (0) 2392 / 2505 E-Mail: sales@stauff.com Internet: www.stauff.com
More informationDSL-220 MkIII Single Pass Opacity Monitor Measures 0-100% Opacity
Ideal for monitoring opacity and smoke levels in the exhaust gas of industrial combustion or air filtration processes. In situ measurement directly in exhaust gas flow using the widely accepted light transmission
More informationProduct Data Sheet 2012 Rev 5 FLOWMETRIX SA SAFMAG Electromagnetic Flowmeter Doc Date
FLOWMETRIX SA Doc Date 17.07.2012 PERFORMANCE * 0,5% of rate (Standard) * 0,25% of rate (Optional) * 1000 : 1 operable flow range * Excellent repeatability * Optional Bi-directional measurement * Advanced
More informationmass flowmeter for compressed air
mass flowmeter for compressed air VARIOMASS ECO pressure and temperature compensated easy and cost effective installation of the sensor no pressure loss caused by the sensor no moving parts, i.e. maintenance
More informationHeat Engines Lab 12 SAFETY
HB 1-05-09 Heat Engines 1 Lab 12 1 i Heat Engines Lab 12 Equipment SWS, 600 ml pyrex beaker with handle for ice water, 350 ml pyrex beaker with handle for boiling water, 11x14x3 in tray, pressure sensor,
More informationProduct Sheet Pre-Fabricated Flushing Bypass Assembly for Plant Applications Hero Plant DN15-150
Product Sheet Pre-Fabricated Flushing Bypass Assembly for Plant Applications Hero Plant DN15-150 Description Hero Plant is a pre-fabricated, flushing bypass arrangement for balance, control, isolation,
More informationMarsh McBirney Flo-Dar
Marsh McBirney Flo-Dar Accurate Flow Measurement Flo-Dar TM provides the user with highly accurate flow measurements under a wide range of flows and site conditions. By measuring the velocity of the fluid
More informationNote: If anything is damaged or missing, contact your customer representative immediately.
Package Contents The package includes: 1 Chassis 1-18 leafs according to the amount ordered 18 - X leaf blanks X = the amount ordered 1 leaf fan module 1 spine fan module 9 spines 1-2 management modules
More informationULTRASONIC FLOW SENSOR IFX-M4-01
ULTRASONIC FLOW SENSOR IFX-M4-01 TECHNICAL DESCRIPTION, OPERATING INSTRUCTION IFX-M4-01 2012-09-10 TABLE OF CONTENTS 1. APPLICATION... 3 2. TECHNICAL DATA... 4 3. PACKAGE CONTENT... 6 4. OPERATING PRINCIPLE...
More informationFLEC AIR PUMP Operators Manual JULY QUI-1030 VERSION 1.2
FLEC AIR PUMP 1001 Operators Manual JULY 2009 QUI-1030 VERSION 1.2 This page left intentionally blank FLEC AIR PUMP 1001 Operators Manual Contents 1. Introduction 4 2. Operating instructions 5 2.1. Operating
More informationElectromagnetic flowmeter. Descriptions. Specifications
Descriptions The EFM Series is the most economical flanged electromagnetic flowmeter on the market. It is used in 3 to 12 pipe in HVAC, municipal or industrial water, wastewater, pump stations and packaged
More informationUTM10 Series Ultrasonic Transit-time Flowmeters
Page 1 of 5 TI-P197-01 MI Issue 1 UTM10 Series Ultrasonic Transit-time Flowmeters Description UTM10 ultrasonic flow and energy meters clamp onto the outside of pipes and do not make contact with the internal
More informationThe Multivariable Air Flow Meter
No. SS2-MVC200-0100 OVERVIEW The Multivariable Mass Flow Meter AIRcube, a compact has all function necessary to measure nitrogen (N 2 ) gas, carbon dioxcide (CO 2 ) gas and air. The model MVC10A/MVC10F
More informationDatasheet PDCSY-MW-U. Technical Overview. Features. Product warranty and total quality commitment.
Datasheet Ultrasonic Flow Sensors & Integrator Technical Overview Ultrasonic flow sensors have no moving parts in the volume flow, this makes them almost wear free and noiseless. They measure the flow
More informationTOPAS PMW-basic. Customer benefits. Product features
TOPAS PMW-basic The TOPAS PMW-basic product line covers a wide range of applications in hot water measurement. This innovative system offers all types of water measurement right up to data integration
More informationHeat / Chill Meter PolluCom M Installation and Operating Instructions. PolluCom M. Installation and Operating Instructions. M H 1210 INT, Page 1
PolluCom M Installation and Operating Instructions M H 1210 INT, Page 1 The compact meter PolluCom M is suitable for the measurement of energy consumption in heating or chilling plants filled with the
More informationAppendix 9: New Features in v3.5 B
Appendix 9: New Features in v3.5 B Port Flow Analyzer has had many updates since this user manual was written for the original v3.0 for Windows. These include 3.0 A through v3.0 E, v3.5 and now v3.5 B.
More informationECT Display Driver Installation for AP2 Module
ECT Display Driver Installation for AP2 Module Overview The ECT Display Driver is a small module with a removable wire harness that mounts behind the driver's foot well cover. All wiring connections are
More informationSWII Users Manual. Intercomp Co County Road 116 Minneapolis, MN (763) Fax
Manual Intercomp Co. 3839 County Road 116 Minneapolis, MN 55340 (763)-476-2531 1-800-328-3336 Fax 763-476-2613 www.intercompracing.com Manual #: 700036-I Table of Contents INTRODUCTION...3 SPECIFICATIONS...3
More informationDL 01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05
DL 0 Battery Powered for Leak Testing Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger graphic display stainless
More informationHeating / Cooling Meter PolluCom E Installation and Operation Instructions
Our compact meter PolluCom E is used for measuring energy consumption in plants carrying water as heating or cooling liquid. These installation and operation instructions specify how to install and operate
More informationaquaconcept Product features Customer benefits
aquaconcept The unique aquaconcept modular design covers a wide range of water applications. This innovative system offers all types of water measurement right up to data integration into your specific
More informationSITRANS F flowmeters SITRANS F US. SITRANS FUT1010 Ultrasonic flowmeter. Overview. Application
Overview SITRANS F flowmeters Application Liquid applications Gas applications Pipelines Custody transfer, allocation, Upstream Production wells, line balance, gathering, separation interface/densitometer
More informationSuperstatic 749. Fluidic Oscillation Compact Heat Meter
Superstatic 749 Fluidic Oscillation Compact Heat Meter Application The Superstatic 749 is an autonomous compact thermal energy meter consisting of a flow meter an integrator and a pair of temperature sensors.
More informationArmature Reaction and Saturation Effect
Exercise 3-1 Armature Reaction and Saturation Effect EXERCISE OBJECTIVE When you have completed this exercise, you will be able to demonstrate some of the effects of armature reaction and saturation in
More informationAVL PLU 131U FLOW METER
CONSUMPTION MEASUREMENT TECHNOLOGIES AVL PLU 131U FLOW METER AVL PLU131 UREA FLOW MEASUREMENT SENSORS Function Summary The AVL PLU 131U-020 Flow Sensor has been designed for transient flow rate measurement
More information2/2-way Diaphragm Valve with stainless steel design, weld end or clamp connection, DN 8-25
2/2-way Diaphragm Valve with stainless steel design, weld end or clamp connection, DN 8-25 Type 2103 can be combined with Hermetical separation of fl uids from the operating mechanism by diaphragm Zero
More informationTesting Instrumentation Gasboard 5020 for exhaust gases from petrol or LPG engines
Testing Instrumentation Gasboard 5020 for exhaust gases from petrol or LPG engines GASBOARD 5020 is an analyser for the control of exhaust gases from gasoline or LPG engines; it integrates non dispersive
More informationEPAS Desktop Pro Software User Manual
Software User Manual Issue 1.10 Contents 1 Introduction 4 1.1 What is EPAS Desktop Pro? 4 1.2 About This Manual 4 1.3 Typographical Conventions 5 1.4 Getting Technical Support 5 2 Getting Started 6 2.1
More informationInline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 )
EE771/EE772 Inline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 ) The inline flow meter EE771/EE772, based on the measurement principle of thermal mass flow, is ideally suited for the
More informationT O PA S Applications Features Your benefits
TOPAS Hot water Applications The TOPAS hot water meter is a velocity flowmeter based on the well-established multijet measuring principle. It operates over a wide measuring range and is used in all areas
More informationCOMMON RAIL INJECTOR TESTER STARDEX 0601
COMMON RAIL INJECTOR TESTER STARDEX 0601 1. Security information Read this operating manual carefully before using Stardex 0601 ( Device or Test bench ). The device must be connected only to the three-phase
More informationUFM 800 W, C and UFM 800 W Hot Tapping. Ultrasonic flowmeters. for water and wastewater
KROHNE 12/2006 7.02343.22.00 MAR UFM 800 W, C and UFM 800 W Hot Tapping Ultrasonic flowmeter for water and wastewater...with weld-in sensors for metal pipelines...with built-on or built-in sensors for
More informationSPEEDRIVE INSTRUCTIONS MANUAL
EN SPEEDRIVE INSTRUCTIONS MANUAL Safety warning. The following symbols shown beside a paragraph represent danger warnings associated to the failure to comply with the corresponding instructions. DANGER!
More information2SGT. Electronic Pressure Instrument. Three Separate Functions in a Single Loop-Powered Device
2SGT Electronic Pressure Instrument Three Separate Functions in a Single Loop-Powered Device Switch Two Independent, Programmable SPST Solid State Relays Program switch open and close values in engineering
More informationCONTENTS 1. INTRODUCTION SAFTY INSTRUCTION CABLE CONNECTION SYSTEM DESCRIPTION OPERATION... 9
USER MANUAL 1 CONTENTS 1. INTRODUCTION... 1 2. SAFTY INSTRUCTION... 3 3. CABLE CONNECTION... 4 4. SYSTEM DESCRIPTION... 5 5. OPERATION... 9 6. TROUBLE SHOOTING GUIDE... 25 7. OPERATION MODES..... 27 8.
More informationWATER METERS.
WTER METERS Hydrodynamic design High protection Multi-jet technology ensures uniform distribution of load on the turbine located thanks to the water inlet diffuser. The movement activates the magnetic
More informationMotor Tuning Instructions
6/20/12 Motor Tuning Instructions Before you begin tuning: 1. Make sure Pro-Motion is installed. 2. Hook up motor drive, motor, and computer. - Connect motor drive to computer using a USB to Serial Com
More informationDMT PERFORMING UNDER PRESSURE DMT. Pressure Scanner Features. Applications. Description 1/7
Features User Accessible Memory for Test Configuration Management ±0.05% FS System Accuracy EU Throughput Rates of 500 Hz Auto-Negotiating 10/100 BaseT Ethernet with TCP & UDP Protocol Pressure Ranges
More informationVZFA/VZOA MISURATORI DI PORTATA FUEL OIL METERS. CONTOIL VZFA/VZOA , versions for special requirements / applications
VZFA/VZOA FUEL OIL METERS MISURATORI DI PORTATA CONTOIL VZFA/VZOA 5...50, versions for special requirements / applications For applications requiring an increased accuracy of ± % or better, such as: Versions
More informationADU 5. Automatic Distillation Unit
ADU 5 Automatic Distillation Unit Atmospheric Distillation Distillation tests are used to characterize the volatility of petrochemical products. These volatility characteristics are a critical measurement
More informationCORONA E ELECTRONIC METER MULTI-JET
APPLICATION Fully electronic compact water meter with impeller scanning for recording volume. Highly accurate recording of all billing data at medium temperatures up to 90 C. FEATURES 4 Electronic sensor
More informationAutomated Seat Belt Switch Defect Detector
pp. 10-16 Krishi Sanskriti Publications http://www.krishisanskriti.org/publication.html Automated Seat Belt Switch Defect Detector Department of Electrical and Computer Engineering, Sri Lanka Institute
More informationTest Report Project Nr. 2013/020
PISTON Ltd. 1033 Budapest Szőlőkert u. 4./B HUNGARY Tel: Fax: http: Email: (361)-275-0033 (361)-275-0034 www.pistonmedical.com info@pistonmedical.com Test Report Project Nr. 2013/020 Name and address of
More informationFocus On Training. Notes:
Model Type: Mechanical Pumps Page 1 1. Title Page Tier 2 Series Pump T-2 Components 2. Specification Sheet Understand the reason for using latest edition specification Under the use of SSI and Internet
More informationSAILMON SYSTEM LOAD CELL BOX MANUAL. Sailmon LoadCellBox Manual V1.0 Page 1
SAILMON SYSTEM LOAD CELL BOX MANUAL Sailmon LoadCellBox Manual V1.0 Page 1 Sailmon LoadCellBox Manual V1.0 Page 2 CONTENTS 1. Product Description...4 2. Technical Specifications...5 3. Installation...6
More informationDomestic water meters
EnergyMetering Domestic water meters Multi-jet wet dial meters Multi-jet dry dial meters Multi-jet semi-dry dial meters/roller Protect Cartridge meters All that counts. Cold water meters Hot water meters
More informationUser manual ma Loop Calibrator Jofra macal
User manual ma Loop Calibrator Jofra macal User Manual macal ma loop Calibrator Copyright 2005 AMETEK DENMARK A/S LIST OF CONTENTS 1. INTRODUCTION 1.1 PACKING LIST... 2 2. SAFETY INSTRUCTIONS... 2 3.
More informationMATRIX LLC LEVEL CONTROL PANEL
MATRIX LLC LEVEL CONTROL PANEL Installation Operation Maintenance Manual 1-26-09 REV-0.3 1 INDEX Introduction. 3 Product Configurations..... 4 I/O Descriptions. 8 Operation o Intercooler, Accumulator,
More informationProduct Information Stepper Motor-driven Diaphragm Dosing Pump MEMDOS SMART LP
Product Information Stepper Motor-driven Diaphragm Dosing Pump Reliable dosing of chemicals Stepper motor driven diaphragm pumps are particularly suitable for highly accurate, reproducible industrial dosing
More informationAccutrol, LLC Product Sheet. VTS Series. US Patents 6,991,177 & 7,543,759. Digital Airflow Measurement Station
Accutrol, LLC Product Sheet VTS Series US Patents 6,991,177 & 7,543,759 Digital Airflow Measurement Station Innovative features of the AccuStation VTS Series! Digital Airflow Sensing Modular Design Intuitive
More informationSmall hydrocarbon flow calibration facilities at NMIJ
APMP/TCFF Workshop Small hydrocarbon flow calibration facilities at NMIJ Yoshiya Terao Original presentations Expansion of Calibration Flow Range of Small Liquid Hydrocarbon Flow Facility at NMIJ by Kar-Hooi
More information