2.0 TEST PROCEDURES AND RESULTS Vibration Test TEST EQUIPMENT AND INSTRUMENTATION QUALITY ASSURANCE PROGRAM 7 ATTACHMENTS

Size: px
Start display at page:

Download "2.0 TEST PROCEDURES AND RESULTS Vibration Test TEST EQUIPMENT AND INSTRUMENTATION QUALITY ASSURANCE PROGRAM 7 ATTACHMENTS"

Transcription

1

2 Page No. 2 TABLE OF CONTENTS Page No. 1.0 INTRODUCTION Scope References Test Specimen Description Summary TEST PROCEDURES AND RESULTS Vibration Test TEST EQUIPMENT AND INSTRUMENTATION QUALITY ASSURANCE PROGRAM 7 Table I Amplitude Displacement 4 Table II Response Accelerometer Mounting 4 Figure 1 Rack No. 1 5 Figure 2 Rack No. 2 6 ATTACHMENTS A PHOTOGRAPHS A-1 B VIBRATION TEST DATA B-1 C INSTRUMENTATION EQUIPMENT SHEET C-1

3 Page No INTRODUCTION 1.1 Scope This report documents the test procedures followed and the results obtained during Vibration Testing performed on one RC/R63 (600)(700) Two-Bay Rack for Optima EPS. The CT Rack Assembly was received at Wyle Laboratories on April 13, The receiving inspection revealed the Rack to be in good condition. Testing was performed at Wyle Laboratories Huntsville, Alabama, Test Facility from April 18 through April 20, References Optima EPS Purchase Order No Wyle Laboratories Quotation No. 542/041059/DB Wyle Laboratories Quality Assurance Program Manual, Revision 2 MIL-STD-810F, Environmental Test Methods and Engineering Guidelines MIL-STD-45662A, Calibration System Requirements ANSI/NCSL Z540-1, Calibration Laboratories and Measuring and Test Equipment, General Requirements ISO , Quality Assurance Requirements for Measuring Equipment 1.3 Test Specimen Description The Rack tested was identified as a Part Number /1 and RC/R63 (600) (700) Two-Bay Rack Assembly. The Rack was attached to the Wyle interface fixture via isolators mounted to the bottom and the back/top of the Rack. At the direction of the Optima EPS Representative, the isolators and shelves supplied with the Rack were not considered part of the Rack Assembly that was qualified by this test. At the direction of the Optima EPS Representative, the individual shelves were loaded with steel weights in order to simulate actual assembly loading. The location and individual shelf weight distributions are defined in Figures 1 and 2 of this report. 1.4 Summary The CT Rack Assembly was successfully subjected to a Vibration Test Program in accordance with MIL-STD-810F and references presented in Paragraph 1.2, above. The test results contained herein apply only to the specimen identified in this report.

4 Page No TEST PROCEDURES AND RESULTS 2.1 Vibration Test The CT Rack Assembly was subjected to a Random Vibration Test for 60 minutes in each of the three mutually perpendicular axes. The attachment between the Rack Assembly and the Wyle interface fixture was through the Optima EPS-supplied isolation mounts. These mounts were located on the bottom of the Rack Assembly at five locations and on the back top of the Rack Assembly at two locations. Each bay of the Rack Assembly was supplied to Wyle with seven individual shelves located in each bay. Figures 1 and 2 of this report identify the individual shelf weight as loaded on each shelf. The Vibration Testing was performed on an Electrodynamic Exciter in accordance with MIL-STD-810F and references presented in Paragraph 1.2, above. Table I identifies spectrum modifications requested by Optima EPS prior to test. These modifications reduced the overall amplitude displacement in order for all Vibration Testing to be performed on one exciter. Table I. Amplitude Displacement Axis Frequency (Hz) Original Amplitude Modified Amplitude Vertical Transverse Longitudinal Two vibration control accelerometers were mounted on the opposite diagonal corners at the interface of the Rack Assembly and the Wyle-supplied interface test fixture. A total of three response accelerometers were mounted to the Rack Assembly as shown in Table II, below. Table II. Response Accelerometer Mounting Accelerometer No. Function Location A1 Control Vibration Table A2 Control Vibration Table A3 Response Center of Racks, Top A4 Response Center of Racks, Middle A5 Response Center of Racks, Bottom

5 Page No TEST PROCEDURES AND RESULTS (Continued) 2.1 Vibration Test (Continued) 6 kg kg 34 6 kg 33 6 kg kg kg kg kg kg kg 15 kg kg 15 kg kg kg kg kg 15 kg kg 9 15 kg 8 15 kg kg 15 kg 6 15 kg kg 15 kg 35 kg kg Seven shelves provided. Left side is proposed weight distribution, right side is actual weight during test. Figure 1. Rack No. 1

6 Page No TEST PROCEDURES AND RESULTS (Continued) 2.1 Vibration Test (Continued) 6 kg kg 34 2 kg 33 2 kg 32 2 kg kg 5 kg kg kg kg 10 kg kg kg 10 kg kg kg 10 kg kg kg kg kg Seven shelves provided. The left side is proposed weight distribution, the right side is actual weight during test. Figure 2. Rack No. 2

7 Page No TEST PROCEDURES AND RESULTS (Continued) 2.1 Vibration Test (Continued) The Vibration Data Sheets and Data Plots included in Attachment B present the type and sequence of the Vibration Testing. After completion of Vibration Testing in each axis, Wyle Laboratories personnel performed visual inspections. During the inspection after completion of the first (Vertical) axis the lower shelf inside Rack 2 had broken loose from its mounting at all four corners. Also noted was fatigue to the bottom left back isolation mount. After completion of the third (Longitudinal) axis, the same bottom-left back isolation mount exhibited fraying and breakage. According to the Optima EPS Representative, neither of these two findings constitutes an anomaly or failure as a result of test since they are not considered by Optima EPS as part of the item to be qualified by this series of tests. When the test was completed the Rack Assembly was returned to Optima EPS, Tucker, Georgia, for evaluation. Photographs of the vibration test setup are presented in Attachment A. The Instrumentation Equipment Sheet for this test setup is presented in Attachment C. 3.0 TEST EQUIPMENT AND INSTRUMENTATION All instrumentation, measuring, and test equipment used in the performance of this test program were calibrated in accordance with Wyle Laboratories' Quality Assurance Program, which complies with the requirements of ANSI/NCSL Z540-1, ISO , and Military Specification MIL-STD-45662A. Standards used in performing all calibrations are traceable to the National Institute of Standards and Technology (NIST) by report number and date. When no national standards exist, the standards are traceable to international standards or the basis for calibration is otherwise documented. 4.0 QUALITY ASSURANCE PROGRAM All work performed on this test program was completed in accordance with Wyle Laboratories Quality Assurance Program. The Wyle Laboratories,, Quality Management System is registered in compliance with the ISO-9001:2000 International Quality Standard. Registration has been completed by Quality Management Institute (QMI), a Division of Canadian Standards Association (CSA). Wyle Laboratories is accredited (Certificate No.: ) by the American Association for Laboratory Accreditation (A2LA) and the results shown in this test report have been determined in accordance with Wyle's scope of accreditation unless otherwise stated in this report.

8 Page No. A-1 ATTACHMENT A PHOTOGRAPHS

9 Page No. A-2 Photograph No. 1 Typical Vibration Test Setup Photograph No. 2 Typical Vibration Test Setup

10 Page No. A-3 Photograph No. 3 Typical Vibration Test Setup Photograph No. 4 Post-Longitudinal Axis Isolation Mount Damage

11 Page No. A-4 Photograph No. 5 Post-Longitudinal Axis Isolation Mount Damage Photograph No. 6 Post Vertical Axis Damage to Mounting Rail

12 Page No. A-5 Photograph No. 7 Typical Weight Distribution

13 Page No. B-1 ATTACHMENT B VIBRATION TEST DATA

14 Page No. B-2

15 Page No. B-3

16 Page No. B-4

17 Page No. B-5

18 Page No. B-6

19 Page No. B-7

20 Page No. B-8

21 Page No. B-9

22 Page No. B-10

23 Page No. B-11

24 Page No. B-12

25 Page No. B-13

26 Page No. B-14

27 Page No. B-15

28 Page No. B-16

29 Page No. B-17

30 Page No. B-18

31 Page No. B-19

32 Page No. B-1 ATTACHMENT C INSTRUMENTATION EQUIPMENT SHEET

33 Page No. B-2

NATIONAL TECHNICAL SYSTEMS

NATIONAL TECHNICAL SYSTEMS Test Report Number 41418-09.PPC Testing of Pancake Alternator & PMU PREPARED FOR: The Perfect Parts Company Sullivan Products 1 North Haven Street Baltimore, MD 21224 PREPARED BY: National Technical Systems

More information

PRODUCT SPECIFICATION

PRODUCT SPECIFICATION of 3 C Section : For the 200222, 2002323, 2002444, 20025, 200252, 200252, 20026, 200262, 200262, 20028, 200283, 200284, 2002822, 2002824, 200283, 2002832, 2002833, 2002834 and 2002835 series parts..0 GENERAL

More information

Titre / Title HIGH RELIABILITY RF COAXIAL LOADS AND ATTENUATORS GENERIC SPECIFICATION

Titre / Title HIGH RELIABILITY RF COAXIAL LOADS AND ATTENUATORS GENERIC SPECIFICATION Date: December 19 th, 16 ISSUE: 1/B PAGE: 1 / 21 Titre / Title HIGH RELIABILITY RF COAXIAL LOADS AND ATTENUATORS GENERIC SPECIFICATION Rédigé par / Written by Responsabilité / Responsibility Date Signature

More information

Shock and Vibration Test report. Shock and Vibration Summary Test Report

Shock and Vibration Test report. Shock and Vibration Summary Test Report Shock and Vibration Summary Test Report Project number: Date creation: October 26 th, 23 Date revision: Number of pages: 22 D318 Summary Test Report Lt 1 TABLE OF CONTENTS 1 SCOPE...4 2 APPLICABLE DOCUMENTS...4

More information

Lightweight Shock and Type I Vibration Test Report on UPS Model No. PS3200RM for Powerstar Inc. Gaithersburg, MD

Lightweight Shock and Type I Vibration Test Report on UPS Model No. PS3200RM for Powerstar Inc. Gaithersburg, MD Test Report No. 12069.1 No. of Pages 21 Lightweight Shock and Type I Vibration Test Report on UPS Model No. PS3200RM for Powerstar Inc. Gaithersburg, MD NU LABORATORIES 312 Old Allerton Road, Annandale,

More information

Environmental Test Report

Environmental Test Report Environmental Test Report Date: March 14, 2018 Report Number: R5597 Orbital ATK Alliant Techsystems Operations LLC 4700 Nathan Lane Plymouth, MN 55442 United States 3/14/2018 Prepared By: Jessica Tonz

More information

600 Amp Surge Arrester

600 Amp Surge Arrester R&D Center 497 Schooley s Mountain Road Hackettstown, NJ 07840-3906 Tele: 908-852-1122 Fax: 908-813-2091 TEST REPORT NUMBER: 379-17-12164 DATE: 4-3-2012 REPORT OF TEST ON 600 Amp Surge Arrester Statements

More information

Environmental Qualification

Environmental Qualification Environmental Qualification Actuation Solutions for Nuclear Powerplants Environmental Qualification of Rotork nuclear electric actuators and gearboxes Redefining Flow Control Contents Section 1 Environmental

More information

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 *

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * PRODUCT: Heavy Duty Stock Car Steel Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 METTLER TOLEDO (INDUSTRIAL) 1900 Polaris Parkway Columbus, OH 43240 Frank Gardone Phone: 412 213 2263 CALIBRATION Valid To: April 30,

More information

Voluntary Specification for Determining Forced Entry Resistance of Side-Hinged Door Systems

Voluntary Specification for Determining Forced Entry Resistance of Side-Hinged Door Systems Task Group Approval Council Approval Product Group Approval Successful Re-ballot 16-1052 Successful Ballot (#17-1011) Out for Ballot #17-1062 AAMA 1304-XX DRAFT #6 DATED 10/4/2017 Voluntary Specification

More information

Product Specification QSL RF Connectors

Product Specification QSL RF Connectors Product Specification QSL RF Connectors 108-2324 04Feb09 Rev A 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the Tyco Electronics QSL RF Connectors which

More information

Voluntary Specification for Determining Forced Entry Resistance of Side-Hinged Door Systems

Voluntary Specification for Determining Forced Entry Resistance of Side-Hinged Door Systems Task Group Approval Council Approval Product Group Approval Successful Re-ballot 16-1052 Successful Ballot (#17-1011) Out for Ballot #18-1016 AAMA 1304-XX DRAFT #7 DATED 2/21/2018 Voluntary Specification

More information

CLASSIFICATION NOTES. Type Testing Procedure for. Crankcase Explosion Relief Valves

CLASSIFICATION NOTES. Type Testing Procedure for. Crankcase Explosion Relief Valves CLASSIFICATION NOTES Type Testing Procedure for Crankcase Explosion Relief Valves Contents 1. Scope, Application 2. Recognized Standards 3. Purpose 4. Test Facilities 5. Explosion Test Process 6. Testing

More information

NATIONAL TECHNICAL SYSTEMS

NATIONAL TECHNICAL SYSTEMS Test Report Number 41508 09.DIA Testing of Optic Assemblies PREPARED FOR: PREPARED BY: Dialight 1501 Route 34 South Farmingdale, NJ 07727 National Technical Systems 36 Gilbert Street South Tinton Falls,

More information

Jun06 Rev D

Jun06 Rev D Product Specification High Speed Serial Data 2 Connector 108-1965 27Jun06 Rev D 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the Tyco Electronics High

More information

CREEP PERFORMANCE TEST REPORT. Rendered to: TREX COMPANY LLC. Seclusions Composite Privacy Fence System TYPE: Composite Fence System

CREEP PERFORMANCE TEST REPORT. Rendered to: TREX COMPANY LLC. Seclusions Composite Privacy Fence System TYPE: Composite Fence System Architectural Testing CREEP PERFORMANCE TEST REPORT Rendered to: TREX COMPANY LLC Seclusions Composite Privacy Fence System TYPE: Composite Fence System Report No.: Report Date: Expiration Date: 130 Derry

More information

Environmental Qualification Test Report for Spider S3/S5

Environmental Qualification Test Report for Spider S3/S5 32-0048 Environmental Qualification Test Report for Spider S3/S5 Tested for: Spider Tracks Limited FDS Job Nº: 21324 Flight Data Systems Pty. Ltd. ABN: 25 060 690 753 31 McGregors Drive, Keilor Park, Victoria

More information

Lightweight Shock Test Report on Power Supply PS1503-H-MB and Adapter Plate for Powerstar Inc. Gaithersburg, MD

Lightweight Shock Test Report on Power Supply PS1503-H-MB and Adapter Plate for Powerstar Inc. Gaithersburg, MD Test Report No. 11543.1 No. of Pages 14 Lightweight Shock Test Report on Power Supply PS1503-H-MB and Adapter Plate for Powerstar Inc. Gaithersburg, MD NU LABORATORIES 312 Old Allerton Road, Annandale,

More information

Apr11 Rev A

Apr11 Rev A Product Specification SMA In-Series Adapters 04Apr11 Rev A 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the TE Connectivity (TE) SMA In-Series Adapters.

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 METTLER TOLEDO (INDUSTRIAL) 1900 Polaris Parkway Columbus, OH 43240 Frank Gardone Phone: 412 213 2263 CALIBRATION Valid To: April 30,

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 DEKRA IST RELIABILITY SEVICES INC. No. 22 Puding Rd., East Dist. Hsinchu City 300, Taiwan, R.O.C. Phone: 886 3 579 5766 ext 7100 Mr. General Lee general.lee@dekra-ist.com

More information

test with confidence HV Series TM Test Systems Hydraulic Vibration

test with confidence HV Series TM Test Systems Hydraulic Vibration test with confidence HV Series TM Test Systems Hydraulic Vibration Experience. Technology. Value. The Difference. HV Series TM. The Difference. Our philosophy is simple. Provide a system designed for optimum

More information

Aug07 Rev A All Paragraphs Revised

Aug07 Rev A All Paragraphs Revised Qualification Test Report Universal MATE-N-LOK* Connector 110-213 28Aug07 Rev A All Paragraphs Revised 1. INTRODUCTION 1.1. Purpose 1.2. Scope Testing was performed on Universal MATE-N-LOK* connectors

More information

MEZALOK* Stacking Connector System

MEZALOK* Stacking Connector System Product Specification 108-2411 25 FEB 16 Rev D MEZALOK* Stacking Connector System 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the MEZALOK* Stacking Connector

More information

Reliability & Qualification Report for MEMS Based Variable Optical Attenuators

Reliability & Qualification Report for MEMS Based Variable Optical Attenuators Reliability & Qualification Report for MEMS Based Variable Optical Attenuators P/N: MMVOA-1-1550-S-9/125-XX-0.25-1 Page 1 of 21 Revision History Rev Approval Date ECM/ECR Description of Change 00 10/18/2008

More information

DER Commissioning Guidelines Community Scale PV Generation Interconnected Using Xcel Energy s Minnesota Section 10 Tariff Version 1.

DER Commissioning Guidelines Community Scale PV Generation Interconnected Using Xcel Energy s Minnesota Section 10 Tariff Version 1. Community Scale PV Generation Interconnected Using Xcel Energy s Minnesota Section 10 Tariff Version 1.3, 5/16/18 1.0 Scope This document is currently limited in scope to inverter interfaced PV installations

More information

Systems. Dynamic Test. Electrodynamic Vibration Test Systems

Systems. Dynamic Test. Electrodynamic Vibration Test Systems Systems 100 - water 120 kn cooled Electrodynamic Vibration Test Systems SW3-10480APP (100KN) SW3-10LS3-550APP (100KN) SW3-10650APP (100KN) SW3-12480APP (120KN) SW3-12LS3-550APP (120KN) Dynamic Test Technical

More information

Test Group Description. Total Number. Figure 1. Unless otherwise stated, the following environmental conditions prevailed during testing:

Test Group Description. Total Number. Figure 1. Unless otherwise stated, the following environmental conditions prevailed during testing: Qualification Test Report 26 November 12 Rev A1 SL Series Jacks, Category 6 1. INTRODUCTION 1.1. Purpose 1.2. Scope Testing was performed on AMP NETCONNECT* SL Series Category 6 Jacks to determine their

More information

FEBRRUARY 22, 2008 TEST REPORT # QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC.

FEBRRUARY 22, 2008 TEST REPORT # QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC. FEBRRUARY 22, 2008 TEST REPORT #207908 QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH, INC. Contech Research REVISION HISTORY DATE

More information

ASTM D4169 Truck Profile Update Rationale Revision Date: September 22, 2016

ASTM D4169 Truck Profile Update Rationale Revision Date: September 22, 2016 Over the past 10 to 15 years, many truck measurement studies have been performed characterizing various over the road environment(s) and much of the truck measurement data is available in the public domain.

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994 EUSTIS CO., INC./PYROCOM CALIBRATION LAB 12407-B Mukilteo Speedway Lynnwood, WA 98087 Bill LeMesurier Phone: 425 423 9996 CALIBRATION

More information

PERFORMANCE SPECIFICATION SHEET BATTERY, RECHARGEABLE, NICKEL CADMIUM, VENTED, 1.2-VOLT, 13 AMPERE- HOUR, BB-475/A CELL

PERFORMANCE SPECIFICATION SHEET BATTERY, RECHARGEABLE, NICKEL CADMIUM, VENTED, 1.2-VOLT, 13 AMPERE- HOUR, BB-475/A CELL INCH-POUND 25 FEB 2011 PERFORMANCE SPECIFICATION SHEET BATTERY, RECHARGEABLE, NICKEL CADMIUM, VENTED, 1.2-VOLT, 13 AMPERE- HOUR, BB-475/A CELL This specification is approved for use within Army CECOM Life

More information

CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH Anti-Ram Bollards

CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH Anti-Ram Bollards CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH 2003 Anti-Ram Bollards Prepared for: RSA Protective Technologies, LLC 1573 Mimosa Court Upland, CA 91784 Test

More information

CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH Anti-Ram Bollards

CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH Anti-Ram Bollards CRASH TEST REPORT FOR PERIMETER BARRIERS AND GATES TESTED TO SD-STD-02.01, REVISION A, MARCH 2003 Anti-Ram Bollards Prepared for: RSA Protective Technologies, LLC 1573 Mimosa Court Upland, CA 91784 Test

More information

SFI SPECIFICATION 49.2 EFFECTIVE: MARCH 22, 2011 *

SFI SPECIFICATION 49.2 EFFECTIVE: MARCH 22, 2011 * SFI SPECIFICATION 49.2 EFFECTIVE: MARCH 22, 2011 * PRODUCT: Top Fuel Front Wing Assemblies 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for

More information

SFI SPECIFICATION 28.1 EFFECTIVE: AUGUST 25, 2017 *

SFI SPECIFICATION 28.1 EFFECTIVE: AUGUST 25, 2017 * SFI SPECIFICATION 28.1 EFFECTIVE: AUGUST 25, 2017 * PRODUCT: Polymer (Foam-Filled) Fuel Cells 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards

More information

SEP 2016 Rev D2

SEP 2016 Rev D2 Product Specification 108-2112 28 SEP 2016 Rev D2 VAL-U-LOK* Connectors 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for VAL-U-LOK* Connectors. 1.2. Qualification

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EMPIRE SCALE CORPORATION d.b.a. PRECISION SCALE & BALANCE d.b.a. EMPIRE SCALE & BALANCE 140 Rotech Drive Lancaster, NY 14086 Margery

More information

SFI SPECIFICATION 1.2 EFFECTIVE: FEBRUARY 9, 2006* PRODUCT: Multiple Disc Clutch Assemblies for Vehicles with Naturally Aspirated Engines

SFI SPECIFICATION 1.2 EFFECTIVE: FEBRUARY 9, 2006* PRODUCT: Multiple Disc Clutch Assemblies for Vehicles with Naturally Aspirated Engines SFI SPECIFICATION 1.2 EFFECTIVE: FEBRUARY 9, 2006* PRODUCT: Multiple Disc Clutch Assemblies for Vehicles with Naturally Aspirated Engines 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes

More information

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP S-D-A FTSH S-DV-A SAMTEC, INC.

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP S-D-A FTSH S-DV-A SAMTEC, INC. AUGUST 28, 2008 TEST REPORT #208383-3 REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP-130-02-S-D-A FTSH-130-02-S-DV-A SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH, INC.

More information

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM H-D-WT TEM H-D-WT SAMTEC, INC.

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM H-D-WT TEM H-D-WT SAMTEC, INC. AUGUST 28, 2008 TEST REPORT #208383-5 REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM-125-02-03.0-H-D-WT TEM-125-02-03.0-H-D-WT SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH,

More information

MIL-STD-883G METHOD LEAD INTEGRITY

MIL-STD-883G METHOD LEAD INTEGRITY LEAD INTEGRITY 1. PURPOSE. This method provides various tests for determining the integrity of microelectronic device leads (terminals), welds, and seals. Test condition A provides for straight tensile

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 TUV SUD AMERICA - AUBURN HILLS 1755 Atlantic Blvd. Auburn Hills, MI 48326 William (Mac) Elliott Phone: 813 284 2736 MECHANICAL Valid To: August 31, 2018 Certificate

More information

SFI SPECIFICATION 1.3 EFFECTIVE: MAY 23, 2000 * PRODUCT: Nitro-Methane Drag Race Multiple Disc Clutch Assemblies

SFI SPECIFICATION 1.3 EFFECTIVE: MAY 23, 2000 * PRODUCT: Nitro-Methane Drag Race Multiple Disc Clutch Assemblies SFI SPECIFICATION 1.3 EFFECTIVE: MAY 23, 2000 * PRODUCT: Nitro-Methane Drag Race Multiple Disc Clutch Assemblies 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and

More information

Huntsville- Aerospace Testing & Engineering Services Highway 20 West Huntsville, AL 35806

Huntsville- Aerospace Testing & Engineering Services Highway 20 West Huntsville, AL 35806 Huntsville- Aerospace Testing & Engineering Services 7800 Highway 20 West Huntsville, AL 35806 Facility Overview Huntsville Facility Overview Site: Total Square Footage: 92 Acres 160,000 sq. ft. of Laboratory

More information

SmartSensor 105 & Advance NEMA Salt Fog

SmartSensor 105 & Advance NEMA Salt Fog SmartSensor 105 & Advance NEMA Salt Fog Test Report Wavetronix products mentioned: SmartSensor 105 SmartSensor Advance As contracted by Wavetronix, Environ Laboratories LLC performed a NEMA salt fog test

More information

MODEL 205 ce VIBRATION METER

MODEL 205 ce VIBRATION METER OPERATION MANUAL VIBRATION METER (INCLUDES MODEL 205M ce) Page 2 TABLE OF CONTENTS SECTION TITLE PAGE 1. INTRODUCTION 3 Description 3 2. FEATURES 4 Batteries 4 Controls 4 Vibration Pickup 4 3. OPERATION

More information

PERFORMANCE SPECIFICATION SHEET BATTERY, STORAGE, AIRCRAFT, MAINTENANCE-FREE, 24-VOLT, 1.5-AMPERE-HOUR

PERFORMANCE SPECIFICATION SHEET BATTERY, STORAGE, AIRCRAFT, MAINTENANCE-FREE, 24-VOLT, 1.5-AMPERE-HOUR METRIC MIL-PRF-8565/6A (AS) 25 June 2001 SUPERSEDING DOD-B-8565/6(AS) 26 June 1987 PERFORMANCE SPECIFICATION SHEET BATTERY, STORAGE, AIRCRAFT, MAINTENANCE-FREE, 24-VOLT, 1.5-AMPERE-HOUR This specification

More information

Jun 20,2014 Rev A

Jun 20,2014 Rev A Product Specification 108-115067 Jun 20,2014 Rev A DDR4 DIMM Through-hole Memory Socket 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the TE Connectivity

More information

Note to users: This electronic document is not subject to copyright. You may copy it, but you are requested not to disseminate altered copies.

Note to users: This electronic document is not subject to copyright. You may copy it, but you are requested not to disseminate altered copies. Note to users: This electronic document is not subject to copyright. You may copy it, but you are requested not to disseminate altered copies. Technology Assessment Program NIJ Standard for Metallic Handcuffs

More information

SFI SPECIFICATION 1.1 EFFECTIVE: NOVEMBER 9, 2001 *

SFI SPECIFICATION 1.1 EFFECTIVE: NOVEMBER 9, 2001 * SFI SPECIFICATION 1.1 EFFECTIVE: NOVEMBER 9, 2001 * PRODUCT: Replacement Flywheels and Clutch Assemblies 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum

More information

FEDERAL STANDARD PERFORMANCE TEST METHOD FOR INTENSIVE USE CHAIRS

FEDERAL STANDARD PERFORMANCE TEST METHOD FOR INTENSIVE USE CHAIRS METRIC November 20, 2007 SUPERSEDING FNEW-83-269E October 31, 1989 FEDERAL STANDARD PERFORMANCE TEST METHOD FOR INTENSIVE USE CHAIRS The General Services Administration has authorized the use of this federal

More information

DETAIL SPECIFICATION SHEET

DETAIL SPECIFICATION SHEET METRIC MIL-DTL-38999/36 30 November 2006 DETAIL SPECIFICATION SHEET CONNECTORS, ELECTRICAL, CIRCULAR, THREADED, PLUG, LANYARD RELEASE, FAIL-SAFE, REMOVABLE CRIMP CONTACTS, PINS, SHELL SIZE 25, SERIES III,

More information

SFI SPECIFICATION 7.2 EFFECTIVE: DECEMBER 15, 2006 *

SFI SPECIFICATION 7.2 EFFECTIVE: DECEMBER 15, 2006 * SFI SPECIFICATION 7.2 EFFECTIVE: DECEMBER 15, 2006 * PRODUCT: Sportsman Lower Engine Containment Device 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum

More information

ULTRASONIC EXAMINATION

ULTRASONIC EXAMINATION Page 1 of 9 ULTRASONIC EXAMINATION PROCEDURE Doc. No. KNS/UT/01.REV 0 Page: 9 of 9 Date: 15.09.2012 ULTRASONIC EXAMINATION PROCEDURE PREPARED & APPROVED BY NDE- Level III CERTIFIED BY REVIEWED BY Page

More information

Shielding Effectiveness of Bal Spring Canted Coil Springs for EMI

Shielding Effectiveness of Bal Spring Canted Coil Springs for EMI Shielding Effectiveness of Bal Spring Canted Coil Springs for EMI Technical Report TR-92 () (100-85) Bal Seal Engineering, Inc. 19650 Pauling Foothill Ranch CA 92610 Toll Free (800) 366-1006 Tel (949)

More information

THERMAL SHOCK TEST REPORT

THERMAL SHOCK TEST REPORT Report No. :DH1509300024A Version:A THERMAL SHOCK TEST REPORT Company :FLIP CHIP OPTO. INC. Address :No.185, Kewang Rd., Longtan Dist., Taoyuan City 325, Taiwan (R.O.C.) Model Name :Apollo 600-B Lot Number

More information

CHAPTER 6 MECHANICAL SHOCK TESTS ON DIP-PCB ASSEMBLY

CHAPTER 6 MECHANICAL SHOCK TESTS ON DIP-PCB ASSEMBLY 135 CHAPTER 6 MECHANICAL SHOCK TESTS ON DIP-PCB ASSEMBLY 6.1 INTRODUCTION Shock is often defined as a rapid transfer of energy to a mechanical system, which results in a significant increase in the stress,

More information

Certificate of Calibration

Certificate of Calibration Model Number Manufacturer Description Serial Number DSOX2014A Keysight Technologies Inc Oscilloscope, 4-channel, 100MHz MY52161639 Date of Calibration 5 Dec 2014 Procedure STE-50114571-A.02.00 Temperature

More information

SFI SPECIFICATION 14.3 EFFECTIVE: AUGUST 25, 2017 *

SFI SPECIFICATION 14.3 EFFECTIVE: AUGUST 25, 2017 * SFI SPECIFICATION 14.3 EFFECTIVE: AUGUST 25, 2017 * PRODUCT: Nitro-Methane Fuel Supercharger Restraint Devices 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and

More information

PDR circuit breaker - Ground fault Datasheet

PDR circuit breaker - Ground fault Datasheet - Ground fault Datasheet protection device Features Blue housing 1-3 poles plus neutral 7-100mA Preciese, temperature independent operation Choise of time delays, terminals and actuator colours Application

More information

OFS Brands Test Facility Report No: OFSB Date: April 2, 2015 Page: 1 of 29. Test Report For: ANSI/BIFMA X Storage Units.

OFS Brands Test Facility Report No: OFSB Date: April 2, 2015 Page: 1 of 29. Test Report For: ANSI/BIFMA X Storage Units. Date: April 2, 2015 Page: 1 of 29 Test Report For: ANSI/BIFMA X5.9-2012 Storage Units FastenLink This report shall not be reproduced, except in full, without written permission from OFS Brands Quality

More information

SFI SPECIFICATION 15.2 EFFECTIVE: MAY 1, 2008 *

SFI SPECIFICATION 15.2 EFFECTIVE: MAY 1, 2008 * SFI SPECIFICATION 15.2 EFFECTIVE: MAY 1, 2008 * PRODUCT: Drag Race Front Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for evaluating

More information

SFI SPECIFICATION 15.4 EFFECTIVE: JULY 20, 2010 * PRODUCT: Top Fuel and Funny Car Drag Race Drive Beadlock Wheels

SFI SPECIFICATION 15.4 EFFECTIVE: JULY 20, 2010 * PRODUCT: Top Fuel and Funny Car Drag Race Drive Beadlock Wheels SFI SPECIFICATION 15.4 EFFECTIVE: JULY 20, 2010 * PRODUCT: Top Fuel and Funny Car Drag Race Drive Beadlock Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures

More information

OIML Weights & Weight Sets

OIML Weights & Weight Sets OIML Weights & s PRECISE WEIGHTS FOR YOUR MOST PRECISE WORK Troemner, an ISO 9001 registered company, is the world s leading supplier of precision weights, weight sets, mass standards and associated calibration/certification

More information

UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS

UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS EDUCATION A TECHNICAL PRESENTATION FROM THE LEADING MANUFACTURER OF METROLOGY INSTRUMENTS Bulletin No. 2183 EDU-14003A-H Mitutoyo Institute of Metrology

More information

Heating oven with mechanical control

Heating oven with mechanical control ED series 23 Drying ovens / Heating ovens Heating oven with mechanical control The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C. Thanks to the

More information

MILITARY SPECIFICATION MOUNT, TELESCOPE AND QUADRANT: M172

MILITARY SPECIFICATION MOUNT, TELESCOPE AND QUADRANT: M172 MILITARY SPECIFICATION MOUNT, TELESCOPE AND QUADRANT: M172 30 MARCH 1990 SUPERSEDING MIL-M-48559(MU) 12 September 1975 This specification is approved for use by the U.S. Army Armament, Munitions and-chemical

More information

U.S. DEPARTMENT OF TRANSPORTATION

U.S. DEPARTMENT OF TRANSPORTATION TP-218-04 June 27, 2003 U.S. DEPARTMENT OF TRANSPORTATION NATIONAL HIGHWAY TRAFFIC SAFETY ADMINISTRATION LABORATORY TEST PROCEDURE FOR FMVSS 218 Motorcycle Helmets SAFETY ASSURANCE Office of Vehicle Safety

More information

REPORT NO INSPECTION, TEST AND EVALUATION OF. Interlock Kits SUBMITTED TO MASTER ELECTRICAL SERVICES 3107 LANVALE AVENUE RICHMOND, VA 23230

REPORT NO INSPECTION, TEST AND EVALUATION OF. Interlock Kits SUBMITTED TO MASTER ELECTRICAL SERVICES 3107 LANVALE AVENUE RICHMOND, VA 23230 : November 16, 2009 REPORT NO. 52431-01 INSPECTION, TEST AND EVALUATION OF Interlock Kits SUBMITTED TO MASTER ELECTRICAL SERVICES 3107 LANVALE AVENUE RICHMOND, VA 23230 A WYLE LABORATORIES, INC. WYLE is

More information

DETAIL SPECIFICATION SHEET

DETAIL SPECIFICATION SHEET METRIC MIL-DTL-38999/31E 12 March 2014 SUPERSEDING MIL-DTL-38999/31D 19 April 2002 DETAIL SPECIFICATION SHEET CONNECTORS, ELECTRICAL, CIRCULAR, THREADED, PLUG, LANYARD RELEASE, FAIL-SAFE, REMOVABLE CRIMP

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ELEMENT MATERIALS TECHNOLOGY MINNEAPOLIS LLC (A subsidiary of Element Materials Technology Minneapolis Inc.) 9725 Girard Avenue South Minneapolis, MN 55431

More information

COMFORT PRODUCT INC. Technical Report: (1215) Initial Report Date: August 03, 2015 Date Received: July 31, 2015 Page 1 of 22

COMFORT PRODUCT INC. Technical Report: (1215) Initial Report Date: August 03, 2015 Date Received: July 31, 2015 Page 1 of 22 CONSUER PRODUCTS SERVICES DIVISION COFORT PRODUCT INC Date Received: July 31, 2015 Page 1 of 22 S.TAN LOT 222, LORONG 4D, SUBANG NEW VILLAGE, 40150 SHAH ALA, SELANGOR, ALAYSIA. Sample Description: 2 DRAWER

More information

Automotive Vehicles Bumper fitment on M1 Category of Vehicles Test Methods

Automotive Vehicles Bumper fitment on M1 Category of Vehicles Test Methods AUTOMOTIVE INDUSTRY STANDARD Automotive Vehicles Bumper fitment on M1 Category of Vehicles Test Methods PRINTED BY THE AUTOMOTIVE RESEARCH ASSOCIATION OF INDIA P. B. NO. 832. PUNE 411 004 ON BEHALF OF

More information

May08 Rev C

May08 Rev C Product Specification 0.64 mm Generation Y Terminal 108-2296 20May08 Rev C 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the Tyco Electronics 0.64 mm Generation

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ELITE ELECTRONIC ENGINEERING, INC. 1 1516 Centre Circle Downers Grove, IL 60515 Larry Brooks (QA Manager) Phone: 630 495 9770 ext. 166 Email: lbrooks@elitetest.com

More information

May2015 Rev A

May2015 Rev A Product Specification 108-60104 15May2015 Rev A ANGLED ENTRY JACK FOR CATEGORY 6 APPLICATIONS 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the TE Connectivity

More information

SFI SPECIFICATION 34.1 EFFECTIVE: January 13, 2017*

SFI SPECIFICATION 34.1 EFFECTIVE: January 13, 2017* SFI SPECIFICATION 34.1 EFFECTIVE: January 13, 2017* PRODUCT: Screw-Type Superchargers 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for evaluating

More information

TEST REPORT. Rendered to: HOMELAND VINYL PRODUCTS, INC. For: PVC Guardrail System Utilizing New Nylon/PVC Rail Mounting Bracket

TEST REPORT. Rendered to: HOMELAND VINYL PRODUCTS, INC. For: PVC Guardrail System Utilizing New Nylon/PVC Rail Mounting Bracket TEST REPORT Rendered to: HOMELAND VINYL PRODUCTS, INC. For: PVC Guardrail System Utilizing New Nylon/PVC Rail Mounting Bracket Report No: Report Date: 10/31/12 130 Derry Court York, PA 17406-8405 phone:

More information

SFI SPECIFICATION 4.2 EFFECTIVE: OCTOBER 26, 1994*

SFI SPECIFICATION 4.2 EFFECTIVE: OCTOBER 26, 1994* SFI SPECIFICATION 4.2 EFFECTIVE: OCTOBER 26, 1994* PRODUCT: Tractor Blankets 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for evaluating

More information

Radiant High Voltage. Displacement Measurement Fixture. Construction. Introduction. Figure 1

Radiant High Voltage. Displacement Measurement Fixture. Construction. Introduction. Figure 1 Radiant High Voltage Displacement Measurement Fixture Introduction Radiant Technologies, Inc. offers four types of high voltage test fixtures. One, the High Voltage Test Fixture (HVTF), has been very popular

More information

EU Mutual Recognition Technical Requirements

EU Mutual Recognition Technical Requirements 1. PRODUCT DESCRIPTION... 1 1.a General description of the product... 1 1.b Application limitations... 1 1.c Intended use... 2 1.d System context... 2 2. DESIGN EVALUATION... 2 2.a Engineering evaluation

More information

Heating oven with natural convection

Heating oven with natural convection ED series 23 Drying ovens / Heating ovens Heating oven with natural convection The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C. Thanks to the

More information

SFI SPECIFICATION 28.3 EFFECTIVE: APRIL 21, 2014 *

SFI SPECIFICATION 28.3 EFFECTIVE: APRIL 21, 2014 * SFI SPECIFICATION 28.3 EFFECTIVE: APRIL 21, 2014 * PRODUCT: Competition Fuel Cell Bladder 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for

More information

Door Locks and Door Retention Components

Door Locks and Door Retention Components TECHNICAL STANDARDS DOCUMENT No. 206, Revision 2R Door Locks and Door Retention Components The text of this document is based on Federal Motor Vehicle Safety Standard No. 206, Door Locks and Door Retention

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 INDEPENDENT ENGINEERING LABORATORIES, INC. 1 2400 E. South Street Jackson, MI 49201 Ms. Jessica Stremich Phone: 517 788 6590 E-mail: IELaccounting@ielinc.com

More information

Mitraset 19 Housings. Housing for electronic instruments

Mitraset 19 Housings. Housing for electronic instruments Housing for electronic instruments Mitraset 19 housings are in accordance with MIL-T 28800, MIL-STD 810 F, EIA and IEEE Standards and are designed to accommodate electronic devices in a 19 rack. They also

More information

SPECIFICATION SHEET Pulse tool / Impact Wrench Testing System MODEL : IWT-101

SPECIFICATION SHEET Pulse tool / Impact Wrench Testing System MODEL : IWT-101 Applications Calibration of Impact tools Calibration oil pulse tools And also any type of manual wrenches Importance of Torque tool calibration: Controlling torque of a torque tool is important for companies

More information

TEST METHODS CONCERNING TRANSPORT EQUIPMENT

TEST METHODS CONCERNING TRANSPORT EQUIPMENT PART IV TEST METHODS CONCERNING TRANSPORT EQUIPMENT - 403 - CONTENTS OF PART IV Section Page 40. INTRODUCTION TO PART IV... 407 40.1 PURPOSE... 407 40.2 SCOPE... 407 41. DYNAMIC LONGITUDINAL IMPACT TEST

More information

VIBRATION AND MECHANICAL SHOCK TEST REPORT

VIBRATION AND MECHANICAL SHOCK TEST REPORT Report No. :DH1701190024A-1 Version:A VIBRATION AND MECHANICAL SHOCK TEST REPORT Company :OTO Photonics Inc. Address :9F-5,No. 27, Guanxin Rd., East Dist., Hsinchu City 30072, Taiwan(R.O.C.) Sample Name

More information

Downloaded from

Downloaded from [INCH-POUND] 28 April, 1995 SUPERSEDING W-C-596F/GEN 24 June, 1983 FEDERAL SPECIFICATION CONNECTOR, ELECTRICAL, POWER, GENERAL SPECIFICATION FOR The General Services Administration has authorized the use

More information

MGA Research Corporation

MGA Research Corporation MGA Research Corporation Real Time Simulation Testing Gerald Roesser David Nagle Thomas Hutter MGA Research Corporation 1 MGA Research Corporation PRESENTERS Gerald Roesser BSEE MGA Associate since 2001

More information

Compliance Certification Services Inc. Report No: MIL(CS101) Date of Issue: August 12, 2009 MIL STD 461E/ F TEST REPORT.

Compliance Certification Services Inc. Report No: MIL(CS101) Date of Issue: August 12, 2009 MIL STD 461E/ F TEST REPORT. Compliance Certification Services Inc. MIL STD 461E/ F TEST REPORT For 20.1 Multimedia LCD Monitor Model Number: WMRM920-PIP Trade Name: itech Issued to itech Company LLC 41758 Christy Street, Fremont

More information

SAMTEC POWER CHARACTERIZATION

SAMTEC POWER CHARACTERIZATION Project Number: Tracking Code: Power Report Rev 3 Requested by: Eric Mings Date: 9/11/2012 Product Rev: AE Part #: MMSD-25-24-L-12.00-D-K-LUS Lot #: «Lot_Number» Tech: Tony Wagoner Eng: Eric Mings Part

More information

Heating chamber with mechanical convection

Heating chamber with mechanical convection FED series 240 Drying ovens / Heating ovens Heating chamber with mechanical convection The multi-talented BINDER heating chamber of the FED series provides almost unlimited capacity and is particularly

More information

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration) CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration) Approved By: Chief Executive Officer: Ron Josias Executive - Accreditation: Mpho Phaloane Author:

More information

Heating chamber with mechanical convection

Heating chamber with mechanical convection FED series 720 Drying ovens / Heating ovens Heating chamber with mechanical convection The multi-talented BINDER heating chamber of the FED series provides almost unlimited capacity and is particularly

More information

SFI SPECIFICATION 35.1 EFFECTIVE: AUGUST 26, 2008 *

SFI SPECIFICATION 35.1 EFFECTIVE: AUGUST 26, 2008 * SFI SPECIFICATION 35.1 EFFECTIVE: AUGUST 26, 2008 * PRODUCT: Stock Car Steel Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards for evaluating

More information

5. STRUCTURAL INTEGRITY 5.1 STRUCTURAL STRENGTH AND DISTORTION TESTS - STRUCTURAL SHAKEDOWN TEST 5.1-I. DISCUSSION

5. STRUCTURAL INTEGRITY 5.1 STRUCTURAL STRENGTH AND DISTORTION TESTS - STRUCTURAL SHAKEDOWN TEST 5.1-I. DISCUSSION 5. STRUCTURAL INTEGRITY 5.1 STRUCTURAL STRENGTH AND DISTORTION TESTS - STRUCTURAL SHAKEDOWN TEST 5.1-I. DISCUSSION The objective of this test is to determine certain static characteristics (e.g., bus floor

More information