Instruction Manual. Model 307A Analog Insulation & Continuity Meter

Size: px
Start display at page:

Download "Instruction Manual. Model 307A Analog Insulation & Continuity Meter"

Transcription

1 Instruction Manual Model 307A Analog Insulation & Continuity Meter

2 Index Page Safety Precautions... 1 Safety Notes... 2 Features Connections... 3 Specifications Why Test is necessary... 5 Instrument Layout... 6 Lid Instruction Label... 7 Preparation for Measurement... 8 Functions Battery & Fuse Replacement Calibration & Servicing Cleaning & Storage Limited One-year Warranty Service Information YOUR COMMENTS ARE IMPORTANT TO US We have become a market leader in test and measurement in recent years, due to the company's ability to design and bring to the market innovative instruments and devices which offer real customer benefits. The cornerstone of this ability is the company's focus on customer requirements and the emphasis placed on end-user satisfaction. We appreciate customer information and requirement to improve or design new products.

3 1. Safety Precautions Electricity can cause severe injuries even with low voltages or currents. Therefore it is extremely important that you read the following information before using your Analog Insulation Tester. 1.1 This Instrument must only be used and operated by a competent trained person and in strict accordance with the instructions. We will not accept liability for any damage or injury caused by misuse or non compliance with instructions and safety procedures. 1.2 This Instrument must not be used on live circuits. Ensure all circuits are de-energised before testing. See paragraph 1.8 for details of built-in warning features should your Analog Insulation tester be connected to a live system. 1.3 Never open Your Analog Insulation Tester except for battery replacement. (See Battery replacement section). 1.4 Always inspect you Analog Insulation tester and test leads before use for any sign of abnormality or damage. If any abnormal conditions exist (broken test leads, cracked case, display faulty etc...) do not attempt to take any measurement or use the tester. Return your Analog Insulation tester to your nearest Distributor for Service. 1.5 Never replace the protective fuse with any other than the specified or approved equivalent. 1.6 Your Analog Insulation tester has been designed with your safety in mind. However, no design can completely protect against incorrect use. Electrical circuits can be dangerous and/or lethal when a lack of caution or poor safety practice is used. Use caution in the presence of voltage above 24V as these pose a shock hazard. 1.7 Pay attention to cautions and warnings which will inform you of potentially dangerous procedures. 1.8 Your Analog Insulation Tester has a live circuit warning bleeper. If it is connected to an AC live circuit, a beep of twice the frequency of the voltage present will be heard. DO NOT proceed to test and immediately disconnect the instrument from the circuit. In addition, the warning light will lit if the voltage is above 100Vdc or 70Vac. When AC voltage is present, before testing, it's value is displayed on the AC scale. -1-

4 2. Safety Notes Rated environmental conditions (1). Indoor use. (2). Installation Category. (3). Pollution Degree 2. (4). Altitude up to 2000M Meter is protected throughout by double insulation or reinforced insulation. 3. Features Warning Risk of electric shock. Caution Refer to this manual before using the meter. High Quality Taut Band movement. Three Insulation test voltages 1-250Vdc -100MW 2-500Vdc -200MW Vdc -400MW Two Continuity Test on Low Ohms 500W & 3W Designed to meet IEC safety standard.. Small and Lightweight, all in one case (do not need bag). AC voltmeter with linear scale up to 600Vac. 200mA continuity short circuit current. 1mA test current on Insulation Test at nominal voltage. Automatic discharge of capacitance and inductive circuit with charge stored in the circuit under test. Live Warning and display of external voltage presence. Fuse, Air Gap, Crowbar and Overload Protected. On line battery monitoring shows if battery is ok.. -2-

5 Auto null of the test leads on continuity tests. Very Low Battery Consumption. On-Load battery check (+/-205mA load for worst case) operates on 8 dry batteries (AA, R6P type..) Mirror scale. Push and Turn locking switch for long and hand free testing. Designed to meet international standards. Supplied with High Quality test leads. 4. Connections CAT. III LINE MAX 600V EARTH -3-

6 5. Specifications INSULATION Test Voltage 250Vdc 500Vdc 1000Vdc +10%-0% +10%-0% +10%-0% Mirror scale 100MW 200MW 400MW Mid Scale 1MW 2MW 4MW Scale Multiplier x1/2 x1 x2 Accuracy ±3% Output Short-Circuit Current 1.3mA Regulated Output Voltage (up to 1mA current) 263.5V / 525V / 1052V CONTINUITY Low W 0-3W W Test Leads / Fuse Zero W Adjustment by knob Output Short-Circuit Current 205mA Accuracy ±1.5% of scale length AC Voltage Range Accuracy 0-600Vac ±3% of scale length Voltage Warning Warning Light Circuit Live lit from 90Vdc / 70Vac Buzzer Beep from 24Vac / dc Battery Check Battery Check indicate good batteries from 8Vdc to 13Vdc during a Load test of 205mA. Battery OK Battery OK Led lit from 8Vdc and is operative while testing. -4-

7 6.Why test is necessary? INSULATION Every electrical apparatus and installation need to be safe for the user and for the equipment itself. Electrical conductors of electricity need to be insulated from each other, so that they do not create electrical hazard or unnecessary consumption. Badly insulated circuits can create leakage current which can be dangerous and trip your GFCI, RCCB or ELCB.. Each country regulate those levels at which the insulation is acceptable. Generally, Insulation resistance measurements are done between each conductor and the earth, and between each conductors. CONTINUITY Checking the continuity of wires, complete circuits, connections, closure of contacts, circuit breakers, fuses, bounding resistance of connections, etc... Are all very important. -5-

8 7. Instrument Layout N-E FOR Vac: DO NOT PRESS PRESS TO TEST P-E INSULATION MW 1000V/400MW 500V/200MW 250V/100MW 1 LOCK FOR CONTINUOUS TESTING, PRESS AND TURN CATIII 600V ALWAYS ENSURE CIRCUIT TO TEST IS FREE OF VOLTAGE BEFORE PROCEEDING WITH CONTINUITY OR INSULATION TESTING. INSULATION-CONTINUITY METER 1. Test Button Switch. 2. Function Selector 3. Battery OK indicator 4. Mirror Scale 5. Live Circuit Warning Light 6. Test Leads and Fuse Zeroing Knob -6-3W 500W Batt. Check W CONTINUITY 0W ZERO OHM ADJUST 2 6

9 8. Lid Instruction Label INSTRUCTIONS ANALOG INSULATION-CONTINUITY TESTER WARNING: This instrument must only be used by a competent trained individual. Consult the full operating instructions. Never press the test button before connecting test leads to circuit to test. INITIAL CHECKS: 1. Switch to "Batt. Check" and depress the test button. If the pointer does not move to "BATT. GOOD", the battery needs to be replaced before proceeding. 2. Connect test leads to instrument, switch to 3W. Press and turn the test button (continuous mode), short the test leads. The pointer should swing from infinite towards zero. INSULATION TESTS MW Ranges 1. Select the desired insulation test voltage range, 250V, 500V, or 1000V. 2. Connect the leads to the instrument and circuit under test. 3. Check the circuit is not LIVE. 4. Press the test button. Read the red MW scale directly for 500V, multiply by 0.5 (or divide by 2) for 250V and multiply by 2 for 1000V. CONTINUITY TESTS - W Ranges 1. Select the desired ohm range, 3W or 500W. 2. Short the test leads, press test button and adjust the ohms zero ADJ to zero the pointer on the 0W (green scale) 3. Check the circuit is not LIVE. 4. Connect the test leads to the circuit under test. Press the test button. Read the selected range directly. GENERAL: For AC Voltmeter, do not press test button, this is the default mode of the instrument. AC Voltmeter can works without batteries. Insulation or Continuity mode: for continous operation, press and turn the test button. SAFETY PRECAUTION The circuit must not be LIVE, conduct initial checks first. If at any time, the "LIVE"circuit light is lit, or the warning buzzer sounds - DO NOT PROCEED, the circuit is live. Using the instrument in Insulation Mode may leave the circuits charged up if test leads are removed too quickly. Avoid this by releasing the test button while the test leads are still connected to the circuit for a few seconds. -7-

10 9.Preparation for Measurement Before testing Always check the following. At Power "ON", check that Bat. OK led lit. And check that there is no visual damage to the Instrument or test leads. Check the test Leads continuity 1. Connect the leads to the Instrument. 2. Zero the test leads while on the 3 ohm range.. 3. This will indicate your that continuity of the test leads is ok. 4. Verify that the test leads insulation is in good condition. 10. Functions 10.1 Battery Check FOR Vac: DO NOT PRESS PRESS TO TEST N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN 1000V/400MW 500V/200MW 250V/100MW 3W 500W Batt. Check INSULATION MW P-E W CONTINUITY Turn the function selector to Batt. Check. This function has a load which draw about 205mA when test is performed, and therefore it is doing a worst case battery test. Then, press the test button, the pointer should be in the Bat. Good area. During the test, the bat. OK Led (on line battery check) must lit if the pointer is in the bat. Good area. -8-

11 10.2 DC Warning The DC warning buzzer will beep continuously when DC voltage is higher than 30Vdc on the test probes and the test button is Not pressed. The neon light "circuit live" will lit when the voltage on the test probes is higher than 90Vdc and the test button is not pressed AC Warning The AC warning buzzer will beep continuously when AC voltage is higher than 20Vac on the test probes and the test button is Not pressed. The neon light "circuit live" will lit when the voltage on the test probes is higher than 65Vdc and the test button is not pressed AC Measurement The AC measurement is automatic on this instrument. as soon as AC voltage is present on the test leads, the instrument will display the AC voltage from 20 to 600Vac on the linear scale. -9-

12 10.5 Low ohms measurement 0-500W FOR Vac: DO NOT PRESS PRESS TO TEST 1000V/400MW 500V/200MW 250V/100MW INSULATION MW N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN Always check for voltage before testing and measuring on a circuit. This instrument is intended for measuring Low W and Insulation resistance on un-energized circuits only. Use the procedure explained at points 7.2, 7.3, 7.4. The first procedure to follow, is to zero the test leads and the fuse resistance. The instrument is equipped with a Zero W Knob. First, short circuit the test leads by connecting them together, then, press the test button and adjust the zero W knob until the pointer is precisely on the "0" of the 500W scale. Use the mirror scale to be precise With the pointer. Connect the test leads to the circuit to be measured. For short test, press button and keep pressed. For long test or hand free Measurements, press and turn the test button W 500W Batt. Check P-E W CONTINUITY 0W ZERO OHM ADJUST

13 10.6 Low ohms measurement 0-3W FOR Vac: DO NOT PRESS PRESS TO TEST 1000V/400MW 500V/200MW 250V/100MW INSULATION MW N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN Batt. Check Always check for voltage before testing and measuring on a circuit. This instrument is intended for measuring Low W and Insulation resistance on un-energized circuits only. Use the procedure explained at points 7.2, 7.3, 7.4. The first procedure to follow, is to zero the test leads and the fuse resistance. The instrument is equipped with a Zero W Knob. First, short circuit the test leads by connecting them together, then, press the test button and adjust the Zero W knob until the pointer is precisely on the "0" of the 3W scale. Use the mirror scale to be precise with the pointer. Connect the test leads to the circuit to be measured. For short test, press button and keep pressed. P-E For long test or hand free Measurements, press and turn the test button W 500W W CONTINUITY 0W ZERO OHM ADJUST

14 10.7 Insulation Resistance 250Vdc FOR Vac: DO NOT PRESS PRESS TO TEST 1000V/400MW 500V/200MW 250V/100MW INSULATION MW N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN Batt. Check P-E Always check for voltage before testing and measuring on a circuit. This instrument is intended for measuring Low W and Insulation resistance on un-energized circuits only. Use the procedure explained at points 7.2, 7.3, 7.4. Check the test leads and fuse resistance byzeroing the test leads and fuse as per the 7.6 procedure. Connect the test leads to the circuit to be measured and wait for a few seconds. The instrument will automatically discharge any remaining energy which could be present on the circuit, and will check for voltage at the same time. Once you are sure that the circuit to be tested is not Energized, then press the button for a short test duration or press and turn the button for a long test. Once you end the test, allow a few seconds for the Instrument to automatically discharge the circuit. 3W 500W W CONTINUITY -12-

15 10.8 Insulation Resistance 500Vdc FOR Vac: DO NOT PRESS PRESS TO TEST 1000V/400MW 500V/200MW 250V/100MW INSULATION MW N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN Batt. Check P-E Always check for voltage before testing and measuring on a circuit. This instrument is intended for measuring Low W and Insulation resistance on un-energized circuits only. Use the procedure explained at points 7.2, 7.3, 7.4. Check the test leads and fuse resistance byzeroing the test leads and fuse as per the 7.6 procedure. Connect the test leads to the circuit to be measured and wait for a few seconds. The instrument will automatically discharge any remaining energy which could be present on the circuit, and will check for voltage at the same time. Once you are sure that the circuit to be tested is not energized, then press the button for a short test duration or press and turn the button for a long test. Once you end the test, allow a few seconds for the Instrument to automatically discharge the circuit. 3W 500W W CONTINUITY -13-

16 10.9 Insulation Resistance 1000Vdc FOR Vac: DO NOT PRESS PRESS TO TEST 1000V/400MW 500V/200MW 250V/100MW INSULATION MW N-E LOCK FOR CONTINUOUS TESTING, PRESS AND TURN Batt. Check P-E Always check for voltage before testing and measuring on a circuit. This instrument is intended for measuring Low W and Insulation resistance on un-energized circuits only. Use the procedure explained at points 7.2, 7.3, 7.4. Check the test leads and fuse resistance byzeroing the test leads and fuse as per the 7.6 procedure. Connect the test leads to the circuit to be measured and wait for a few seconds. The instrument will automatically discharge any remaining energy which could be present on the circuit, and will check for voltage at the same time. Once you are sure that the circuit to be tested is not energized, then press the button for a short test duration or press and turn the button for a long test. Once you end the test, allow a few seconds for the instrument to automatically discharge the circuit. 3W 500W W CONTINUITY -14-

17 11. Battery & Fuse Replacement 11.1 Battery Replacement Your Analog insulation tester's battery is situated under the tester. The Bat. OK will not lit when battery need to be replaced. Disconnect the test leads from the Instrument, remove the battery cover and the batteries. Replace with eight 1.5V R6 or L6 batteries, taking care to observe the correct polarity. Replace battery and the battery cover Fuse replacement The Fuse is located next to the Batteries. To replace Fuse, proceed as per Battery replacement to open the battery cover, then remove and replace the fuse located under the battery holder. Only replace with same specification fuse. (0.5A) -15-

18 12. Calibration & Servicing Contact your nearest distributor about Calibration Certificate and Servicing. Before returning the Instrument, ensure that : the leads have been checked for continuity and signs of damage. The batteries are in good condition. 13. Cleaning & Storage Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents. If the meter is not to be used for long periods or longer than 60 days, remove the batteries and store them separately. Warning To avoid electrical shock or damage to the meter, do not get water inside the case. -16-

19 14. Limited One-year Warranty B&K Precision warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of one year from date of purchase from an authorized B&K Precision distributor. B&K Precision will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To obtain warranty coverage in the U.S.A., this product must be registered by completing the warranty registration form on within fifteen (15) days of purchase. Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed. B&K Precision shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you. This warranty gives you specific rights and you may have other rights, which vary from state-to-state. B&K Precision Savi Ranch Parkway Yorba Linda, CA

20 15. Service Information Warranty Service: Please return the product in the original packaging with proof of purchase to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Return the product in the original packaging to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on open account must include payment in the form of a money order or credit card. For the most current repair charges please visit and click on "service/repair". Return all merchandise to B&K Precision with pre-paid shipping. The flat-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North American is included for Warranty Service only. For overnight shipments and non-north American shipping fees please contact B&K Precision. B&K Precision Savi Ranch Parkway Yorba Linda, CA Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem. -18-

21 B&K Precision Savi Ranch Parkway Yorba Linda, CA U.S.A. Printed in Taiwan / Ver. 1.0/ B&K Precision Corporation

ANALOGUE INSULATION-CONTINUITY and VOLTAGE METER

ANALOGUE INSULATION-CONTINUITY and VOLTAGE METER 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com BST-IT26 ANALOGUE INSULATION-CONTINUITY and VOLTAGE METER INSTRUCTION MANUAL Index

More information

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter Instruction Manual 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Model 308A Digital Insulation & Continuity Meter INDEX PAGE INSTRUMENT LAYOUT... 1 INTRODUCTION... 2 SAFETY

More information

Instruction Manual. Model 310 Digital Milli-Ohm Meter

Instruction Manual. Model 310 Digital Milli-Ohm Meter Instruction Manual Model 310 Digital Milli-Ohm Meter DESCRIPTION PAGE SAFETY RULES... 01 GENERAL DESCRIPTION... 02-03 FRONT PANEL LAYOUT... 04 PREPARATION FOR USE... 05 PRELIMINARY CHECKS... 05 PRECAUTIONS...

More information

SLA Battery Capacity Analyzer

SLA Battery Capacity Analyzer Model: 600B SLA Battery Capacity Analyzer USER MANUAL Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation,

More information

DC REGULATED POWER SUPPLY

DC REGULATED POWER SUPPLY INSTRUCTION MANUAL Models 1620A/ 1621A/ 1622A/ 1623A/ 1626A/ 1627A MANUAL DE INSTRUCCIONES DC REGULATED POWER SUPPLY TABLE OF CONTENTS 1. INTRODUCTION... 3 2. FEATURES... 3 3. SPECIFICATIONS... 4 4. PRCAUTIONS

More information

HIGH VOLTAGE PROVING UNIT

HIGH VOLTAGE PROVING UNIT HIGH VOLTAGE PROVING UNIT INSTRUCTION MANUAL Index 1. Safety Precautions... 2. Specifications... 3. Features... 4. Connections... 5. Layout of Tester... 6. Instructions Label... 7. Proofing Methods...

More information

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model User's Manual High Voltage Megohmmeter Analog Insulation Tester plus AC Voltage and Continuity Tests Model 380353 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects

More information

Model 1672, 1673 Triple Output Power Supply

Model 1672, 1673 Triple Output Power Supply Model 1672, 1673 Triple Output Power Supply INSTRUCTION MANUAL 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases

More information

Model 1693, 1694 Switching DC Power Supply

Model 1693, 1694 Switching DC Power Supply Model 1693, 1694 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases

More information

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT.

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT. INSULATION TESTER EARTH MAX. AC 600V LINE AC -+ HOLD V k M LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT M H.V. POWER DATA HOLD TEST ON / OFF 250V 500V 1000V ACV OFF 0 CAT.III 600V INSTRUCTION

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION 11/05 Form #271 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 OPERATING INSTRUCTION Visit us at www.testequipmentdepot.com MODEL 3132 ANALOG INSULATION-CONTINUITY TESTER

More information

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 1672 Triple Output Power Supply

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 1672 Triple Output Power Supply 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Instruction Manual Model 1672 Triple Output Power Supply Contents Section Description Page No. CONTENTS 1 1 TEST INSTRUMENT

More information

Model 1900, 1901, 1902 Switching DC Power Supply

Model 1900, 1901, 1902 Switching DC Power Supply Model 1900, 1901, 1902 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all

More information

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions.

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions. 1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions. KYORITSU will not accept liability for any damage or injury caused by misuse

More information

INSTRUCTION MANUAL ANALOGUE INSULATION TESTER KEW 3131M KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD., TOKYO JAPAN

INSTRUCTION MANUAL ANALOGUE INSULATION TESTER KEW 3131M KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD., TOKYO JAPAN INSTRUCTION MANUAL ANALOGUE INSULATION TESTER KEW 3131M KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD., TOKYO JAPAN Contents 1. Safety Precautions... 1 2. Features... 3 3. Specifications... 4 4. Instrument

More information

Model P772 Conductivity Meter. User Manual

Model P772 Conductivity Meter. User Manual Model P772 Conductivity Meter User Manual 1 Table of Contents 1 INTRODUCTION... 1 2 SAFETY SUMMARY... 2 3 COMPLIANCE STATEMENTS... 2 4 PRODUCT CONTENTS AND INSPECTION... 3 5 DEVICE DESCRIPTION... 4 6 OPERATING

More information

DC REGULATED POWER SUPPLY

DC REGULATED POWER SUPPLY INSTRUCTION MANUAL Models 1620A/ 1621A/ 1622A/ 1623A/ 1626A/ 1627A DC REGULATED POWER SUPPLY Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com TABLE OF CONTENTS

More information

Model 1900, 1901, 1902 Switching DC Power Supply

Model 1900, 1901, 1902 Switching DC Power Supply Model 1900, 1901, 1902 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS A.W. Sperry Instruments Inc. AC SNAP-AROUND WITH DC VOLT RANGE MODELS TD-2608 & TD-2608A 1) FEATURES UL Listed Five Functions, 12 Ranges. Tear drop shaped jaws for ease of use in

More information

Digital/Analog Megohmmeter Model 1026 USER MANUAL

Digital/Analog Megohmmeter Model 1026 USER MANUAL Digital/Analog Megohmmeter Model 1026 USER MANUAL Limited Warranty The Megohmmeter Model 1026, is warranted to the owner for a period of 2 years from the date of original purchase against defects in manufacture.

More information

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to RS-3 PRO RS-1007 PRO CAT IV Analog Clamp meter Series Users Manual For detailed specifications and ordering info go to www.testequipmentdepot.com RS-3 PRO RS-1007 PRO CAT IV Analog Clampmeter Series English

More information

User Guide. Model Insulation Tester / Megohmmeter. Introduction

User Guide. Model Insulation Tester / Megohmmeter. Introduction User Guide Model 380363 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380363 provides three test ranges plus continuity

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 8/05 Form #270 OPERATING INSTRUCTION MODEL 3005 DIGITAL INSULATION-CONTINUITY TESTER

More information

NEW 15KV HIGH VOLTAGE INSULATION RESISTANCE TESTER MODEL KM 7015 IN FEATURES : CAT IV 600V. Short Circuit Current 5mA.

NEW 15KV HIGH VOLTAGE INSULATION RESISTANCE TESTER MODEL KM 7015 IN FEATURES : CAT IV 600V. Short Circuit Current 5mA. An ISO 9001:2008 Company 15KV HIGH VOLTAGE INSULATION RESISTANCE TESTER FEATURES : MODEL KM 7015 IN Microprocessor controlled Menu driven. Variable High Voltage : 500 V ~ 15 kv DC. Measuring 30 Insulation

More information

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to PRM-4 Phase Sequence and Motor Rotation Tester Users Manual For detailed specifications and ordering info go to www.testequipmentdepot.com PRM-4 Phase Sequence and Motor Rotation Tester English Users Manual

More information

I I e k ill L ill. INDUSTRIAL DIGITAL EARTH LEAKAGE CIRCUIT BREAKER and 3 PHASE ROTATION TESTER BESANTEK BST-ELC01 INSTRUCTION MANUAL

I I e k ill L ill. INDUSTRIAL DIGITAL EARTH LEAKAGE CIRCUIT BREAKER and 3 PHASE ROTATION TESTER BESANTEK BST-ELC01 INSTRUCTION MANUAL = = = = - - - BESANTEK - - BST-ELC1 INDUSTRIAL DIGITAL EARTH LEAKAGE CIRCUIT BREAKER and 3 PHASE ROTATION TESTER EARTH DIGITAL INDUST RIAL ELCB HAS OT ATl ONTE O C, 1:. : ==== === ;... P E R STE R ==!

More information

IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL

IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN1001/IN1003 ANALOGUE INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL Martindale Electric Company LTD Metrohm House Penfold Trading Estate Imperial Way Watford Herts WD24 4YY T: 01923 441717 F: 01923 446900 Email: sales@martindale-electric.co.uk web: www.martindale-electric.co.uk

More information

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC. INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL INS120P TASCO, INC. THIS TESTER IS DESIGNED FOR USE ONLY BY QUALIFIED ELECTRICIANS. IMPORTANT SAFETY WARNINGS mwarning Read and understand this material

More information

3 PHASE TESTER & MOTOR ROTATION TESTER

3 PHASE TESTER & MOTOR ROTATION TESTER 3 PHASE TESTER & MOTOR ROTATION TESTER R S T TEST CAT. III 500V 3 PHASE TESTER & MOTOR ROTATION TESTER CCW CW BAT. L2 L1 L3 L1 L2 L3 MOTOR ROTATION INSTRUCTION MANUAL INDEX Page 1. INTRODUCTION... 1 2.

More information

Digital Low Resistance Ohm Meter

Digital Low Resistance Ohm Meter BST-MGR01 Digital Low Resistance Ohm Meter INSTRUCTION MANUAL Index 1. Safety Rules... 2. Safety Check... 3. General Description... 4. Operating Instructions... 5. Display... 6. Specifications... 11-12

More information

MARTINDALE MARTINDALE IN2001/IN2003 INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN2001/IN2003 INSULATION & CONTINUITY TESTERS

MARTINDALE MARTINDALE IN2001/IN2003 INSULATION & CONTINUITY TESTERS INSTRUCTION MANUAL IN2001/IN2003 INSULATION & CONTINUITY TESTERS MARTINDALE ELECTRIC Martindale Electric Company LTD Metrohm House Penfold Trading Estate Imperial Way Watford Herts WD24 4YY T: 01923 441717 F: 01923 446900 Email: sales@martindale-electric.co.uk web:

More information

Table of Contents Title Page No.

Table of Contents Title Page No. Table of Contents Title Page No. INTRODUCTION 1 SAFETY PRECAUTIONS / WARNINGS 2 CONTROLS AND INDICATORS 4 PREPARATION AND CAUTION BEFORE USE 5 TESTING PROCEDURES 6 A. AC/DC VOLTAGE MEASUREMENT 6 B. RESISTANCE

More information

INSP-3 Wiring Inspection Tester

INSP-3 Wiring Inspection Tester INSP-3 Wiring Inspection Tester Users Manual For detailed specifications and ordering info go to www.testequipmentdepot.com INSP-3 Wiring Inspection Tester Users Manual English December 2009, Rev.1 2009

More information

A.W. SPERRY INSTRUMENTS INC.

A.W. SPERRY INSTRUMENTS INC. 1/94 Form #245 OPERATING INSTRUCTIONS Model SP-152A POCKET SIZE MULTLTESTER PLEASE READ THESE OPERATING INSTRUCTIONS CAREFULLY Misuse and or abuse of these instruments cannot be prevented by any printed

More information

KEWTECH. KT56 digital multi function tester. Instruction manual

KEWTECH. KT56 digital multi function tester. Instruction manual KEWTECH KT56 digital multi function tester Instruction manual Contents 1 Safety Notice 1 2 Features and Principles of Measurement 3 3 Introduction 6 4 Specifications 7 5 Instrument layout 9 6 Operating

More information

User's Manual. Earth Ground Resistance Tester. Model Introduction

User's Manual. Earth Ground Resistance Tester. Model Introduction User's Manual Earth Ground Resistance Tester Model 382152 Introduction Congratulations on your purchase of the Extech Earth Ground Resistance Tester. This device can measure resistance (in 3 ranges) and

More information

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1 Introduction Descriptions All the information you need in one device. Verify

More information

BATTERY OPERATED DIGITAL MILLI-OHMMETER

BATTERY OPERATED DIGITAL MILLI-OHMMETER BATTERY OPERATED DIGITAL MILLI-OHMMETER C 1 P 1 P 2 C 2 MAX. DC20V MICROPROCESSOR CONTROLLED AC DC HOLD M A X M I N AUTO-HOLD Hz mv km ka ms DIGITAL MILLIOHM METER 2000 200.0 20.00 2000m Test Current Range

More information

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING #61-085 #61-086 Read First: Safety Information This tester complies with UL3111-1, Cat III-600

More information

Digital/Analog Megohmmeter Model 1026

Digital/Analog Megohmmeter Model 1026 Digital/Analog Megohmmeter Model 1026 USER MANUAL 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Table of Contents Warning...3

More information

For technical questions and replacement parts, please call

For technical questions and replacement parts, please call Digital CLAMP MULTIMETER 42396 95652 OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd., Camarillo, CA

More information

KMP Insulation-Continuity Tester. Users Manual

KMP Insulation-Continuity Tester. Users Manual KMP 7036 Insulation-Continuity Tester Users Manual KMP7036 Insulation-Continuity Tester Users Manual English September 2011, Rev.1 2011 Amprobe Test Tools. All rights reserved. Printed in China Limited

More information

HIGH VOLTAGE PROXIMITY DETECTOR

HIGH VOLTAGE PROXIMITY DETECTOR HIGH VOLTAGE PROXIMITY DETECTOR An ISO 9001:2008 Company Model 275 HP (Upto 275KV) / 500 HP (Upto 500KV) The 275HP / 500HP is a HIGH VOLTAGE proximity detector. 275HP has eight voltage detection settings

More information

Safety design conforming to the following provisions

Safety design conforming to the following provisions Safety design conforming to the following provisions of IEC61010. Overvoltage category III 300V AC, pollution degree 2 Overvoltage category II 600V AC/DC, pollution degree 2 Protected throughout by double

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION 11/05 Form #273 OPERATING INSTRUCTION MODEL 4105 Disital Earth Resistance Tester 2150 joshua's Path, Suite 302, Hauppauge, NY 11788 Phone : 1-800-645-5398 or 1-639-231-7050 Fax : 1-639-434-3128 E-mail

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS CM OPERATING INSTRUCTIONS 9 Function, Auto Range Digital Multi-Meter DM6450 INTERTEK Read this owner s manual thoroughly before use and save. C LISTED US I. DISPLAY FUNCTIONS & SYMBOLS 9 6 10 11 14 8 7

More information

Metrohm E3640 FLASH TESTER INSTRUCTION MANUAL. Martindale Electric Co Ltd.

Metrohm E3640 FLASH TESTER INSTRUCTION MANUAL. Martindale Electric Co Ltd. Metrohm Martindale Electric Metrohm House, Imperial Park, Imperial Way, Watford, Hertfordshire, WD24 4PP, UK T: 01923 441717 F: 01923 446900 Email: sales@martindale-electric.co.uk web: www.martindale-electric.co.uk

More information

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

MY /2-DIGIT DIGITAL MULTIMETER Users Manual MY-65 4 1/2-DIGIT DIGITAL MULTIMETER Users Manual Read the Users Manual thoroughly before use. WARRANTY This instrument is warranted to be free from defects in material and workmanship for a period of

More information

PD-20 HIGH VOLTAGE PROBE INSTRUCTION MANUAL

PD-20 HIGH VOLTAGE PROBE INSTRUCTION MANUAL PD-20 HIGH VOLTAGE PROBE INSTRUCTION MANUAL Index 1. Safety precaution... 2. Specifications... 3. Operation... 4. Warning... 5. Cleaning... 6. Rated environmental conditions... 7. Observe the international

More information

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to LH41A Clamp On Ammeter User Manual For detailed specifications and ordering info go to www.testequipmentdepot.com LH41A Clamp On Ammeter Users Manual English April 2007, Rev.2 2007 Amprobe Test Tools.

More information

USER MANUAL. 10KV Digital High Voltage Insulation Tester. Model MG500. Additional User Manual Translations available at

USER MANUAL. 10KV Digital High Voltage Insulation Tester. Model MG500. Additional User Manual Translations available at USER MANUAL 10KV Digital High Voltage Insulation Tester Model MG500 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech Instruments Model MG500.

More information

362 Clamp Meter. Users Manual

362 Clamp Meter. Users Manual 362 Clamp Meter Users Manual PN 3622678 August 2010, Rev. 1, 4/13 2010, 2013 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product names

More information

INSTRUCTION MANUAL M110A Fax: (503)

INSTRUCTION MANUAL M110A Fax: (503) INSTRUCTION MANUAL M110A 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com Introduction The M110A, in addition to high-resolution voltage and ohms scales needed for electrical

More information

7/05 FORM#347 OPERATING INSTRUCTIONS MULTI FUNCTION TESTER MODEL 6030

7/05 FORM#347 OPERATING INSTRUCTIONS MULTI FUNCTION TESTER MODEL 6030 7/05 FORM#347 OPERATING INSTRUCTIONS MULTI FUNCTION TESTER MODEL 6030 Contents 1. Safety Precautions 1 2. Features 3 3. Specifications 4 4. Layout Diagram 7 5. Preparation for Measurement 9 6. Operation

More information

373 Clamp Meter. Users Manual

373 Clamp Meter. Users Manual 373 Clamp Meter Users Manual PN 3612805 July 2010 2010 Fluke Corporation. All rights reserved. Printed in Taiwan. Specifications are subject to change without notice. All product names are trademarks of

More information

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale: Check out what else you can get from Martindale: 17th Edition Testers Accessories Calibration Equipment Continuity Testers Electricians Kits Environmental Products Full Calibration & Repair Service Fuse

More information

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER KEW6201A CONTENTS 1. Safe testing 1 2. Procedure of removing cover. 3 2.1 Method of removing the cover. 3 2.2 Method of storing the cover. 3 3. Product summary

More information

1.Safe testing IMPORTANT:

1.Safe testing IMPORTANT: Electricity is dangerous and can cause injury and death. Always treat it with the greatest of respect and care. If you are not quite sure how to proceed, then stop, take advice from a qualified person.

More information

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc 707 Loop Calibrator Instruction Sheet Introduction The Fluke 707 Loop Calibrator (hereafter, the Calibrator) is a compact sourcing and measuring tool. The Calibrator tests current loops of 0-20 ma or 4-20

More information

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer TIC 300 PRO High Energy Tic Tracer Non Contact AC Voltage Detector Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso TIC 300 PRO High Energy Tic Tracer Non Contact AC Voltage Detector

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Digital Clamp Meter DSA500A IMPORTANT: RECEIVING INSTRUCTIONS Visually inspect all components for shipping damage. If you find damage, notify the carrier at once. Shipping damage

More information

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings 1662/1663/1664 FC Electrical Installation Tester Safety Information Go to www.fluke.com to register your Product and find more information. A Warning identifies conditions and procedures that are dangerous

More information

5.2 PRE-OPERATION PROCEDURE WARRANTY

5.2 PRE-OPERATION PROCEDURE WARRANTY 4/02 Form #314 OPERATING INSTRUCTIONS MODELS DM-210A, 220A, 230A DIGITAL MULTIMETER PLEASE READ THESE OPERATING INSTRUCTIONS CAREFULLY Misuse and or abuse of these instruments cannot be prevented by any

More information

Instruction manual. High voltage insulation resistance tester KEW3121B/3122B

Instruction manual. High voltage insulation resistance tester KEW3121B/3122B Instruction manual High voltage insulation resistance tester KEW3121B/3122B Contents 1. Safety warnings... 1 2. Features... 5 3. Specification... 6 4. Instrument layout... 9 5. Getting started... 10 5-1

More information

Safety design conforming to the following provisions

Safety design conforming to the following provisions Safety design conforming to the following provisions of IEC61010. Overvoltage category III 300V, pollution degree 2, Overvoltage category II 600V, pollution degree 2, Data hold switch for easy reading

More information

TES-1601 INSTRUCTION MANUAL AUTORANGING INSULATION TESTER SAFETY INFORMATION TES ELECTRICAL ELECTRONIC CORP. Warnings and Safety symbols:

TES-1601 INSTRUCTION MANUAL AUTORANGING INSULATION TESTER SAFETY INFORMATION TES ELECTRICAL ELECTRONIC CORP. Warnings and Safety symbols: TES-1601 AUTORANGING INSULATION TESTER INSTRUCTION MANUAL SAFETY INFORMATION The circuit under test must be de-energized and isolated before connections are made except for voltage measurement. Circuit

More information

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY BACK LIGHT D I G I TA L M U LT I M E T E R DT33B OPERATOR LIMITED WARRANTY AND LIMITATION OF LIABILITY This product will be free from defects in material and workmanship for one year from the date of purchase.

More information

Model Insulation Tester / Megohmmeter. User Manual

Model Insulation Tester / Megohmmeter. User Manual Model 380363 Insulation Tester / Megohmmeter User Manual Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380363 provides three test ranges plus continuity

More information

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale:

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale: Check out what else you can get from Martindale: 17th Edition Testers Accessories Calibration Equipment Continuity Testers Electricians Kits Environmental Products Full Calibration & Repair Service Fuse

More information

User s Manual Rev 1.3 GME

User s Manual Rev 1.3 GME User s Manual Rev 1.3 GME TEST INSTRUMENT SAFETY GUIDELINES WARNING An electrical shock of over 10 milliamps of current to pass through the heart will stop most human heartbeats. Voltage as low as 35 volts

More information

MARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions

MARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions 4.4 Storage Conditions The FD650/R and FD500/T or FD600/T should be kept in warm dry conditions away from direct sources of heat or sunlight, with the battery removed and in such a manner as to preserve

More information

User Guide. Model Insulation Tester / Megohmmeter

User Guide. Model Insulation Tester / Megohmmeter User Guide Model 380260 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380260 provides three test ranges plus continuity

More information

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols Table of Contents Title Page Introduction Unpacking the Meter Safety Information International Electrical Symbols The Meter Structure Key Functions Rotary Switch Measurement Operation A. Measuring Voltages

More information

365 Detachable Jaw True-rms Clamp Meter

365 Detachable Jaw True-rms Clamp Meter 365 Detachable Jaw True-rms Clamp Meter Users Manual PN 3622684 August 2010 2010 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product

More information

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings 1587 FC/1587/1577 Insulation Multimeter Safety Information Go to www.fluke.com to register your Product and find more information. A Warning identifies conditions and procedures that are dangerous to the

More information

Users Manual. Phase Rotation Indicator

Users Manual. Phase Rotation Indicator 9040 Phase Rotation Indicator Users Manual PN 2438546 April 2005, Rev.2, 5/11 2005-2011 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are

More information

Megohmmeter Test Probe

Megohmmeter Test Probe Megohmmeter Test Probe User Manual ENGLISH www.aemc.com CHAUVIN ARNOUX GROUP Copyright Chauvin Arnoux, Inc. d.b.a. AEMC Instruments. All rights reserved. No part of this documentation may be reproduced

More information

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER INSTRUCTION MANUAL 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT These instructions contain both information and warnings

More information

BK BK J93 MA BK MA BK MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

BK BK J93 MA BK MA BK MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK n 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com COMPATIBLE CURRENT PROBES & SENSORS FOR POWER QUALITY METERS 193-24-BK 193-36-BK

More information

PY-11 Voltage & Continuity Tester

PY-11 Voltage & Continuity Tester PY-11 Voltage & Continuity Tester Users Manual PY-11 Voltage & Continuity Tester English Users Manual May 2010, Rev.1 2010 Amprobe Test Tools. All rights reserved. Printed in China Limited Warranty and

More information

DIGITAL RCD(ELCB) TESTER

DIGITAL RCD(ELCB) TESTER INSTRUCTION MANUAL DIGITAL RCD(ELCB) TESTER KEW 5410 R KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD. Contents 1. Safety Warnings.... 1 2. Procedure of removing Cover. 3 2-1 Method of removing the Cover.

More information

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT CM69 TRMS AC EARTH LEAKAGE CLAMP METER INSTRUCTION MANUAL 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT These instructions contain both information and warnings

More information

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK n COMPATIBLE CURRENT PROBES & SENSORS FOR POWER QUALITY METERS 193-24-BK 193-36-BK A196-24-BK MA193-10-BK MA193-14-BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK E N G L I S H User Manual Copyright

More information

323/324/325 Clamp Meter

323/324/325 Clamp Meter 323/324/325 Clamp Meter Users Manual PN 4045153 May 2012 2012 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product names are trademarks

More information

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus:

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus: 1.0 Introduction: This probe is designed to provide an additional help to automotive technicians in trouble shooting of electrical circuits problems in the car. Apart from using the normal multi tester,

More information

i400s Calibration Information Introduction

i400s Calibration Information Introduction i400s AC Current Clamp Introduction XWWarning To avoid electric shock or injury, do not perform the performance tests or calibration procedures unless you are qualified to do so. The information provided

More information

Calibration Manual. Detachable Jaw True-rms Clamp Meter

Calibration Manual. Detachable Jaw True-rms Clamp Meter 365 Detachable Jaw True-rms Clamp Meter Calibration Manual December 2010 2010 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks

More information

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free TX900 Microwave Leakage Detector 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Users Manual TX900 Microwave Leakage Detector

More information

Rental Industry Safety Tester Safe Check 5s

Rental Industry Safety Tester Safe Check 5s Rental Industry Safety Tester Safe Check 5s Feb 2006 2006 Clare Instruments Inc. Issue 2.0 Firmware Version : 1.2a Limited Warranty & Limitation of Liability Clare Instruments Inc, guarantees this product

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

Model T2642 Wall Mount. Television Wall Mount with Tilt Option

Model T2642 Wall Mount. Television Wall Mount with Tilt Option Model T2642 Wall Mount Television Wall Mount with Tilt Option Getting Started Introduction Congratulations on the purchase of your new Audio Solutions T2642 Television Wall Mount. For maximum benefit,

More information

User Manual. T6 Tachometer. Online: Telephone: P.O. Box St. Petersburg, Florida 33736

User Manual. T6 Tachometer. Online:   Telephone: P.O. Box St. Petersburg, Florida 33736 User Manual T6 Tachometer Online: www.phareselectronics.com Telephone: 727-623-0894 P.O. Box 67251 St. Petersburg, Florida 33736 Table of Contents Overview... 1 Description... 1 Wiring... 1 T6 Tachometer

More information

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to AMB-50 5000V Insulation Resistance Tester Users Manual For detailed specifications and ordering info go to www.testequipmentdepot.com AMB-50 5000V Insulation Resistance Tester Users Manual English June

More information

Model FM2642 Wall Mount. Full Motion Television Wall Mount

Model FM2642 Wall Mount. Full Motion Television Wall Mount Model FM2642 Wall Mount Full Motion Television Wall Mount Getting Started Introduction Congratulations on the purchase of your new Audio Solutions FM2642 Television Wall Mount. For maximum benefit, please

More information

User Guide. Model Insulation Tester / Megohmmeter

User Guide. Model Insulation Tester / Megohmmeter User Guide Model 380260 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380260 provides three test ranges plus continuity

More information

Model AS-RC3260 TV Cart. Rolling Cart for Audio Mount System & Flat Panel TVs

Model AS-RC3260 TV Cart. Rolling Cart for Audio Mount System & Flat Panel TVs Model AS-RC3260 TV Cart Rolling Cart for Audio Mount System & Flat Panel TVs GETTING STARTED Introduction Congratulations on the purchase of your new Helios AS-RC3260 Rolling Cart. For maximum benefit,

More information

362 Clamp Meter. Users Manual

362 Clamp Meter. Users Manual 362 Clamp Meter Users Manual PN 3622678 August 2010 2010 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product names are trademarks

More information

233 True-rms Remote Display Digital Multimeter Safety Information

233 True-rms Remote Display Digital Multimeter Safety Information 233 True-rms Remote Display Digital Multimeter Safety Information Go to www.fluke.com to register your product and find more information. A Warning identifies conditions and procedures that are dangerous

More information

PHASERS. Operating and Instruction Manual. a n d A C C E S S O R I E S

PHASERS. Operating and Instruction Manual. a n d A C C E S S O R I E S VOLTMETERS PHASERS a n d A C C E S S O R I E S Operating and Instruction Manual HD ELECTRIC COMPANY 1 4 7 5 L A K E S I D E D R I V E WA U K E G A N, I L L I N O I S 6 0 0 8 5 U. S. A. PHONE 847.473.4980

More information

MANUAL PCE-IT111

MANUAL PCE-IT111 www.pce-industrial-needs.com Tursdale Technical Services Ltd Unit N12B Tursdale Business Park Co. Durham DH6 5PG United Kingdom Phone: +44 ( 0 ) 191 377 3398 Fax: +44 ( 0 ) 191 377 3357 http://www.industrial-needs.com/

More information