1 Instruction Manual Model PH87 Retraction Sensor Assembly Yokogawa Corporation of America 2 Dart Road, Newnan, Georgia U.S.A Tel: Fax: Doc. Nr. L2528 January rd edition
2 2 TABLE OF CONTENTS Overview... 3 Assembly Instructions-Stage # Assembly Installation Instructions- with /B ball valve option... 8 In-Service Installation-with /B ball valve option Sensor Service Sensor Replacement Exploded View of PH Part Numbers and Descriptions PH87 Dimensions and Clearance Requirements... 18
3 3 Sensor Installation Overview NOTE: THE FOLLOWING INFORMATION IS INTENDED TO PROVIDE THE USER WITH A PREVIEW OF WHAT WILL BE REQUIRED TO INSTALL THE MODEL PH87 SENSOR ASSEMBLY. PLEASE REVIEW THE DIAGRAM AND GUIDELINES BEFORE PROCEEDING. Typical Sensor Installation Diagram Shown with Optional /B Ball Valve ASSEMBLY ph/orp SENSOR Sensor Installation Guidelines: [1]. Install the model PH97 sensor in pipe or vessel between 15 and 90 from horizontal. [2]. Connect flex conduit with adequate service loop to junction box. [3]. Connect the sensor cable to the terminals inside the junction box. [4]. Keep cable run between junction box and ph / ORP instrument as short as possible.
4 4 Sensor Assembly Instructions-Stage1 Begin stage #1 Assembly of the model PH87 by following the step-by-step instructions [A] through [F] in order as shown below: [Step B]. slide packing gland, item no.3, onto sheath, item no.4. keep locking nut and ferrules loose but attached to packing gland [Step A]. Install packing gland o-ring, item no. 3d, into packing gland body, item no. 3c. [Step F]. Thread sensor cable through coupler tube, item no. 1, then slide coupler tube into sheath until it contacts sensor. [Step E]. Stretch o-ring, item no. 1a, into groove on coupler tube, item no. 1. [Step D]. Thread Sensor Cable through sheath and slide the sensor into sheath until it stops against flared end Flared end of sheath [Step C]. stretch o-rings, item no. 2, into grooves on sensor PLEASE NOTE REFER to pages 16 and 17 of the sensor assembly and itemized part REFER TO PAGES 15 AND 16 OF THE list for all SENSOR these ASSEMBLY components AND ITEMIZED PART LIST FOR ALL THESE COMPONENTS.
5 Sensor Assembly ly Instructions-Stage1 Instructions Stage1 (Continued) 5 [Step G]. Thread sensor cable through coupler tube, item no. 1, then slide coupler tube into sheath until it contacts sensor. the coupler tube and sensor should be threaded together until both are seated against the ends of the sheath,
6 6 Sensor Assembly Instructions-Stage1 (Continued) [Step H]. REMOVE JUNCTION BOX COVER AND THERAD SENSOR CABLE THROUGH [Step H]. JUNCTION REMOVE BOX ASSEMBLY (JUNCTION JUNCTION BOX BOX AND COVER PROVIDED AND THERAD SENSOR CABLE THROUGH PIPE NIPPLE). THREAD JUNCTION JUNCTION BOX ASSEMBLY BOX ASSEMBLY PIPE NIPPLE INTO (JUNCTION BOX AND PROVIDED SENSOR PIPE NIPPLE). COUPLER THREAD ASSEMBLY, JUNCTION ITEM BOX NO. ASSEMBLY 1.INSTALL PIPE SUITABLE NIPPLE FLEX INTO CONDUIT SENSOR LONG COUPLER ENOUGH ASSEMBLY, TO ALLOW FOR ITEM SERVICING NO. 1.INSTALL OR REPLACEMENT SUITABLE FLEX OF CONDUIT SENSOR. LONG ENOUGH TO ALLOW FOR SERVICING OR REPLACEMENT OF SENSOR. [Step I]. CONNECT THE SENSOR WIRES TO THE 5-TERMINAL STRIP IN THE [Step JUNCTION I]. CONNECT BOX IN THE THE SENSOR FOLLOWING WIRES TO ORDER. THE 5-TERMINAL THE RED TC STRIP WIRE IN THE JUNCTION BOX IN THE FOL- CONNECTS TO TERMINAL #1 AND LOWING ORDER. THE RED TC WIRE THE BLACK TC WIRE TO TERMINAL CONNECTS TO TERMINAL #1 AND #2. THE CONNECT BLACK THE TC WIRE SENSOR TO TERMINAL REFERENCE #2. CONNECT (BLACK) THE TO SENSOR #3, REFER- THE ENCE SOLUTION (BLACK) GROUND TO #3, THE (GREEN) SOLUTION TO GROUND TERMINAL (GREEN) #4 AND TO ph TERMINAL (CLEAR) TO #4 TERMINAL AND ph #5. (CLEAR) THREAD TO THE TERMINAL EXTENSION #5. CABLE THREAD THROUGH THE EXTENSION CONDUIT. ATTACH CABLE THE THROUGH LABELED CONDUIT. WIRES TO ATTACH THE THE TERMINALS LABELED IN WIRES THE SAME TO THE ORDER TERMINALS AS THE IN SENSOR THE SAME WIRES ORDER LISTED AS THE ABOVE. SEN- REPLACE SOR WIRES THE JUNCTION LISTED ABOVE. BOX COVER. SLIDE- THE PROTECTIVE INSULTING BOOT OVER CONNECTORS AND REPLACE THE JUNCTION BOX COVER.
7 7 Sensor Assembly Instructions-Stage1 (Continued) The stage #1 assembly of the ph97 sensor is now complete and should appear as one of the two assemblies shown below Standard STANDARD Assembly ASSEMBLY [STEp J]. Perform a 2-point calibration with ph buffer solutions before proceeding further with the assembly and installation. Note: refer to ph 97 sensor instructions for wiring and calibration procedures. NOTE: REFER TO PH 97 SENSOR INSTRUCTIONS FOR WIRING AND CALIBRATION PROCEDURES.
8 8 Sensor Assembly Instructions-/B Ball Valve Option Begin assembly of the /B ball valve option for model PH87 sensor by following the step-by-step instructions [a] through [d] in order as shown below: PLEASE NOTE REFER to pages 16 and 17 of the REFER sensor TO PAGES assembly 15 AND and 16 OF itemized THE SENSOR ASSEMBLY AND ITEMIZED PART part list for all these LIST FOR ALL THESE COMPONENTS. components [Step a]. Thread the reducer bushing, item no.6, using teflon tape to seal and tighten in place [Step D]. Insure that the valve is in the closed postion as shown below before proceeding to next step. [Step B]. Thread the Nipple, item no.7, into ball valve, item no.6, using teflon tape to seal and tighten in place [Step C]. Thread the Nipple, item no.7, Along with ball valve, item no.6, and reducer bushing, item no.5, into weldalet or appropriate process fitting. use teflon tape to seal threads and tighten in place PLEASE NOTE: /B OPTION IS REQUIRED (OR CUSTOMER SUPPLIED BALL VALUE IS REQUIRED) FOR RETRACTION OF PH87 WHILE PROCESS IS UNDER PRESSURE
9 9 Sensor Assembly Installation Instructions-/B Ball Valve (Continued) Complete /B ball valve option assembly of the model PH87 By following the step-by-step instructions [e] and [f] in order as shown below: [Step E]. Apply teflon tape to pipe threads of packing gland fitting and slide fitting to sensor end of sheath. install stage #1 subassembly into valve taking care not to force the sensor into the ball portion of the valve to prevent breakage of the glass bulb. [Step F]. Securely Tighten packing gland fitting into valve assembly to provide a proper seal. THE ASSEMBLY OF THE THE SENSOR ASSEMBLY IS NOW OF THE COMPL PH87 SENSOR AND SHOULD IS NOW COMPLETE APPEAR A AND SHOWN. SHOULD DO APPEAR NOT TIGHT AS SHOWN. PACKING DO NOT GLAND TIGHTEN LOCKIN PACKING NUT OR GLAND OPEN LOCKING BALL VAL NUT UNTIL OR OPEN SENSOR BALL ASSEMB VALVE UNTIL IS READY SENSOR TO ASSEMBLY GO ON-STR IS READY TO GO ON-STREAM.
10 10 Sensor Assembly In-Service Instructions-/B Ball Valve Option PLEASE NOTE REFER to pages 16 and 17 of the REFER TO PAGES 15 AND 16 OF THE sensor assembly and itemized SENSOR ASSEMBLY AND ITEMIZED PART part LIST FOR list ALL for THESE all COMPONENTS. these components THE FINAL ASSEMBLY IS NOW READY TO BE PLACED INTO SERVICE, WITH PACKING GLAND LOCKING NUT ENGAGED BUT NOT TIGHTENED AND SENSOR IN THE RETRACTED POSITION. [Step B]. INSERT STAGE #1 SUBASSEMBLY INTO PROCESS. THE IN-SERVICE ASSEMBLY SHOULD APPEAR AS SHOWN BELOW. [Step A]. A]. TO TO PLACE PLACE ASSEMBLY INTO INTO SERVICE OPEN OPEN BALL BALL VALVE COMPLETELY. Caution: OPENING VALVE SUBJECTS SENSOR ASSEMBLY TO TO LINE LINE PRESSURE. IF THE IF THE SENSOR IS IS NOT NOT RETRACTED COMPLETLY AGAINST AGAINST STOP, STOP, OPENING OPENING THE THE VALVE VALVE COULD COULD RESULT RESULT IN IN THE THE SENSOR SENSOR SUDDENLY SUDDENLY RETRACTING RETRACTING THE THE REMAINING REMAINING DISTANCE DISTANCE CAUSING DAMAGE TO THE CAUSING DAMAGE TO THE SENSOR ASSEMBLY OR INJURY SENSOR ASSEMBLY OR INJURY TO THE SERVICE PERSON. TO THE SERVICE PERSON. [Step C]. SECURE SENSOR IN PLACE BY TIGHTENING LOCKING NUT AND FERRULES OF GLAND FITTING, ITEM NO.3. DO NOT INSERT TIP OF SENSOR BEYOND INSIDE WALL OF PROCESS PIPE TO AVOID TORQUE AND POSSIBLE BREAKAGE.
11 11 Sensor Service Instructions Before Servicing remove the model ph97 sensor from service following the step-by-step instructions [a] and [b] in order as shown below: [Step B]. [Step FULLY B]. FULLY RETRACT RETRACT STAGE #1 STAGE SUBASSEMBLY #1 SUB- SHAFT. ASSEMBLY SHAFT. [Step A]. LOOSEN LOCKING NUT [Step AND A]. FERRULES LOOSEN OF LOCKING GLAND FITTING, NUT AND ITEM FERRULES NO. 3. OF GLAND Caution: FITTING, THE ITEM SENSOR NO. 3. ASSEMBLY IS Caution: STILL UNDER THE SENSOR PRESSURE. ASSEMBLY LOOSENING IS STILL UNDER THE LOCKING PRESSURE. NUT COULD LOOSENING RESULT THE IN THE LOCKING SENSOR NUT RETRACTING COULD RESULT SUDDENLY IN THE AND SENSOR WITH RETRACTING A GREAT DEAL SUDDENLY OF FORCE. AND THIS WITH COULD A GREAT CAUSE DEAL DAMAGE OF FORCE. TO THIS THE COULD SENSOR CAUSE ASSEMBLY DAMAGE OR TO INJURY THE SENSOR TO THE SERVICE ASSEMBLY PERSON. OR INJURY TO THE SERVICE PERSON.
12 12 Sensor Service Instructions (Continued) Before Servicing remove the model PH97 sensor from service following the step-by-step instructions [C] and [D] in order as shown below: [Step c]. Close ball valve completely. [Step D]. Completely unthread gland fitting, item no.4, from valve assembly 10
13 13 Sensor Service Instructions (Continued) Continue servicing of the model ph97 sensor by following the step-by-step instructions [e] through [g] in order as shown below: [Step G]. TO TEST SENSOR PERFORM A [Step 2-POINT G]. TO TEST CALIBRATION SENSOR WITH PERFORM ph BUFFER A 2-POINT AS CALIBRATION SHOWN WITH BELOW. ph BUFFER AS SHOWN BELOW. IF THE IF SENSOR THE SENSOR TESTS TESTS WITHIN WITHIN ACCEPTABLE ACCEPTABLE LIMITS LIMITS REINSTALL REINSTALL AND PLACE AND PLACE SENSOR SENSOR BACK IN SERVICE BACK BY IN SERVICE FOLLOWING BY FOLLOWING ASSEMBLY ASSEMBLY INSTRUCTIONS. INSTRUCTIONS. IF TESTING IF TESTING PROVES PROVES SENSOR SENSOR T0 BE T0 FAULTY BE FAULTY [Step F]. Before testing the CONTINUE CONTINUE WITH THE WITH SERVICE THE SERVICE INSTRUCTIONS INSTRUcTIONS sensor be sure that it is clean FOR SENSOR REPLACEMENT. FOR SENSOR REPLACEMENT. [Step E]. Remove stage #1 Assembly from valve assembly. With the stage #1 Subassembly completely removed from the valve assembly the sensor may now be tested or replaced.
14 14 Sensor Replacement Instructions Before Servicing remove the model PH87 sensor from service following the step-by-step instructions [a] and [D] in order as shown below: [Step C]. WHILE FIRMLY GRIPING [Step SHEATH C]. While PUSH firmly ON COUPLER griping TUBE sheath - JUNCTION BOX ASSEMBLY TO push on coupler tube junction box UNSEAT SENSOR O-RINGS assembly to unseat sensor o-rings [Step A]. REMOVE JUNCTION BOX COVER AND DISCON- NECT [Step ALL A]. WIRES REMOVE FROM TERMINALS. JUNCTION BOX COVER AND DISCONNECT ALL WIRES FROM TERMINALS. [Step B]. WHILE FIRMLY GRIPING THE [Step SENSOR B]. WHILE AND THE FIRMLY SHEATH GRIPING AT THE THE SAME SENSOR TIME UNTHREAD AND THE SHEATH THEM AT FROM THE THE SAME COUPLER TIME UNTHREAD ASSEMBLY THEM BUT DO FROM NOT COMPLETELY THE COUPLER DISENGAGE ASSEMBLY BUT THREADS. DO NOT COMPLETELY DISENGAGE THREADS. [Step D]. COMPLETELY UNTHREAD SENSOR [Step FROM D]. COUPLER COMPLETELY TUBEUNTHREAD ASSEMBLY. SENSOR FROM COUPLER TUBE ASSEMBLY. CONTINUE WITH SERVICE INSTRUCTIONS CONTINUE AT WITH STEP SERVICE "N". INSTRUCTIONS AT STEP N.
15 15 Sensor Replacement Instructions (Continued) DISASSEMBLE THE ASSEMBLY INTO THE COMPONENT PARTS AS SHOWN IN ONE OF THE ILLUSTRATIONS BELOW. TO INSTALL A NEW SENSOR AND RETURN THE MODEL PH87 TO SERVICE FOLLOW THE ASSEMBLY INSTRUCTIONS STARTING FROM STAGE #1 STEP D (PAGE 4) FIGURE 1: STANDARD ASSEMBLY READY FOR FIGURE SENSOR 1: REPLACEMENT STANDARD ASSEMBLY READY FOR SENSOR REPLACEMENT FIGURE 2: Assembly with sensor mounted FIGURE 2: junction ASSEMBLY box WITH ready for SENSOR MOUNTED JUNCTION sensor replacement BOX READY FOR SENSOR REPLACEMENT
16 16 Sensor Housing Assembly 8 JUNCTION Junction BOX box PLEASE NOTE REFER to pages 15 and 16 of the sensor assembly and itemized part list for all these components ALL O-RINGS REQUIRE PROPER LUBRICATION PRIOR TO SERVICING. WE RECOMMEND USING PARKER O LUBE OR EQUIVALENT. FOLLOW THE LUBRICATION INSTRUCTIONS PROVIDED BY THE MANUFACTURER. 2 Required 9 ph SENSOR ph sensor (SEE (see MODEL model PH97 ph97 SPEC SHEET) spec sheet)
17 17 Sensor Housing Assembly These are the individual part numbers for the items shown in the exploded view on the previous page. Part numbers for a complete assembly are listed on next two pages, on the specific drawing. Item Part Number Description 1 M1234DQ Coupler Assembly, Kynar (PVDF), N=7 M1234TV-A Coupler Assembly, Kynai (PVDF), N=21 These part numbers includes the o-ring 1a, 2 M1234DS O-rings, viton, sensor, size -019, quantity of 10 3 M1234DT Compression Fitting, 1 Tube x 1 MNPT 3a M1234DW Compression Fitting, Nut 3b M1234DX Compression Fitting, Ferrule, Kynar (PVDF) 3c M1234EJ Compression Fitting, Body, Titanium 3d M1234DZ Compression Fitting, O-ring, viton, size M1234EA Sheath, Titanium, N=7 M1234JQ-A Sheath, Titanium, N=21 5 M1234WZ-A Bushing, 1.25 MNPT x 1 FNPT 6 M1234WY-A Ball Valve, Full Port, 1.25 NPT 7 M1234XA-A Nipple, 1.25 NPT 8 M1234EE Junction Box Assembly, Sensor Mount 9 See ph Sensor Specifications
18 18 "C" = MINIMUM CLEARANCE REQUIRED FOR SENSOR REMOVAL "R" = FULLY RETRACTED 3/4" FNPT 15.65" FULLY INSERTED "N" = FULLY INSERTED 3.72" 5.22" 1 1/4" MNPT MODEL NO. N LENGTH Insertion Length R DIMENSION Retraction Distance C DIMENSION Minimum Clearance PH87-07/B PH87-21/B Dart Road, Newnan, GA Tel.: Fax.: PH87 With /B Ball Valve Insertion Assembly: Dimensions & Clearance Requirements
19 19 "C" = MINIMUM CLEARANCE REQUIRED FOR SENSOR REMOVAL "R" = FULLY RETRACTED 3/4" FNPT 9.65" FULLY INSERTED "N" = FULLY INSERTED 3.72" 1 " MNPT MODEL NO. N LENGTH Insertion Length R DIMENSION Retraction Distance C DIMENSION Minimum Clearance PH PH Dart Road, Newnan, GA Tel.: Fax.: PH87 Insertion Assembly: Dimensions & Clearance Requirements
20 YOKOGAWA HEADQUARTERS 9-32, Nakacho 2-chome, Musashinoshi Tokyo 180 Japan Tel. (81) Fax (81) YOKOGAWA EUROPE B.V. Euroweg HD Amersfoort The Netherlands Tel. +31 (0) Fax +31 (0) info@nl.yokogawa.com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA United States Tel. (1) Fax (1) info@yca.com YOKOGAWA ELECTRIC ASIA Pte. Ltd. 5 Bedok South Road Singapore Singapore Tel. (65) Fax (65) webinfo@yas.com.sg Yokogawa has an extensive sales and distribution network. Please refer to the website ( to contact your nearest representative. IM M1234JL-A Subject to change without notice Copyright 2008 Printed in The USA
Instruction Manual. Models FD30 / PB30. Fittings for the DO30G. IM 12J06K04-E-E 6th edition
Instruction Manual Models FD30 / PB30 Fittings for the DO30G 6th edition 2 Table of contents 1. Introduction...3 2. Unpacking and checking...3 3. Warranty and service...3 4. Genral specifications...4 4-1.
More informationGeneral Specifications
General Specifications Model -FTS Differential ph/orp-sensor Overview The -FTS is the newest development in ph sensor technology available from Yokogawa. This sensor combines the measuring technology of
More informationInstruction Manual. Model SC4A 19mm Conductivity Sensors. IM 12D7J4-E-E 6th edition YOKOGAWA
Instruction Manual Model SC4A 19mm Conductivity Sensors YOKOGAWA 6th edition TABLE OF CONTENTS 1. GENERAL... 1 1-1. Type number... 1 1-2. General specifications SC4A-AD... 1 1-3. General specifications
More informationManifolds for DPHarp EJA and EJX Pressure and Differential Pressure Transmitters. Pressure psi (bar) (860) (552) (413) 4.
General Specifications Manifolds for DPHarp EJA and EJX Pressure and Differential Pressure Transmitters www.parker.com Yokogawa Europe has included Parker Hannifin manifolds with our DPharp range of pressure
More informationGeneral Specifications
General Specifications Model DO30G and FD30/PB30 Sensor and fittings for Dissolved Oygen DO The DO30G sensor for dissolved oygen is designed for use in water treatment plants such as sewage treatment works,
More informationGeneral Specifications
General Specifications Manifolds for DPHarp EJA and EJX Pressure and Differential Pressure Transmitters Yokogawa Europe bv has included Sami Instruments manifolds in his DPHarp pressure and differential
More informationGeneral Specifications
General Specifications Model PH20 and FU20 ph/redox and Temperature sensor The PH20 is nicknamed "Tempress" because of the patented compensation for changes in the process temperature and pressure. This
More informationGeneral Specifications
General Specifications Model DO30 and FD30/PB30 Sensor and fittings for Dissolved Oygen DO The DO30 sensor for dissolved oygen is designed for use in water treatment plants such as sewage treatment works,
More informationUser Manual. Model PR20 Retractable fitting for combined ph/redox electrodes. IM 12B6K3-E-E 9th Edition
User Manual Model PR20 Retractable fitting for combined ph/redox electrodes 9th Edition CONTENTS 1. GENERAL 2 1-1.Description 2 1-2.Features 2 1-3.Safety precautions 2 2. UNPACKING AND CHECKING 3 2-1.Unpacking
More informationGeneral Specifications
General Specifications Compact ph Probes Model EXA Series EXA Compact is a versatile series of probes suitable for ph measurements in all common applications in industrial processes and (waste) water treatment
More informationGeneral Specifications
General Specifications Model RAGG Rotameter GS 01R01B06-00E-E This type of Rotameter is designed for measurement of liquids and gases. The conical glass metering tube has a free rotating float. This float
More informationGeneral Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.J January 2011
Product Data Sheet PDS 71-400/rev.J January 2011 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Pre-determined cell constant ensures out-of-the
More informationGeneral Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.I December 2008
Product Data Sheet PDS 71-400/rev.I December 2008 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Pre-determined cell constant ensures out-of-the
More informationGeneral Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.G February 2006
Product Data Sheet PDS 71-400/rev.G February 2006 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Pre-determined cell constant ensures out-of-the
More informationAccessories Guide ph/orp/conductivity Sensors
Jet Cleaners and Scrubber Accessories Isolation Ball Valves Assemblies Compression Fittings and Sheath Sets High Pressure (2500 PSIG) Sensor Housings In-line Flow Cells Nut Lock Union Style Adapters Sensor
More informationInstruction Manual YOKOGAWA. Model FD20 Immersion Fittings. ph/orp. IM 12B6K2-01E-E 7th edition
Instruction Manual Model FD20 Immersion Fittings ph/orp YOKOGAWA IM 12B6K2-01E-E 7th edition TABLE OF CONTENTS 1. INTRODUCTION... 2 2. SPECIFICATIONS... 3 2-1. General specifications... 3 2-2. Functional
More informationInstrumentation Manifolds Traditional C13WT series. For DPharp EJX & EJA-E Differential Pressure and Pressure Transmitters
Instrumentation Manifolds Traditional C13WT series For DPharp EJX & EJA-E Differential Pressure and Pressure Transmitters Introduction About Yokogawa and WIKA About the catalogue Yokogawa and WIKA have
More informationPersonal ph, ph/orp, Conductivity Meters
Personal ph, ph/orp, Meters // Personal ph, ph/orp, Meters Bulletin 12B03D01-01E www.yokogawa.com/an/ Compact and lightweight 220g IP67 Waterproof CE,RCM,KC Various Convenient Data memory Alarm Clock Date/time
More informationCombination ph/redox (ORP) Sensors
Data Sheet Combination ph/redox (ORP) s Series AP00 User-customizable for Insertion, Flow-through, Immersion, Hot-tap Retractor ideal for Original Equipment Manufacturers and general applications Economical
More informationFor additional information, please visit our website at SENSOR SPECIFICATIONS
Instruction Sheet PN 51A-140/rev.E December 2010 Models 140, 141, and 142 Conductivity Sensors For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SENSOR/PROCESS
More informationDigital Vortex Flowmeter. digitalyewflo. Digital Vortex Flowmeter. Series. Bulletin 01F06A02-00E.
Digital Vortex meter digitalyewflo Digital Vortex meter Series Bulletin 01F06A02-00E www.yokogawa.com/ Series Simple body construction for easy installation and high level of safety 2 3 Standard Type /
More informationGeneral Specifications
General Specifications GS 22B01C52-01EN-A M-650, M-750 and M-651 3-Valve Manifolds for Differential Pressure INTRODUCTION The PGI 3-Valve manifolds are designed for mounting on differential pressure transmitters
More informationGeneral Specifications
eneral Specifications Model RAH Rotameter S 01R01B05-00E-E This type of Rotameter is designed for measurement of liquids and gases. The conical glass metering tube has a free rotating float. The mounting
More informationVortex Flowmeter. digitalyewflo. Vortex Flowmeter. Series. Bulletin 01F06A02-00E.
Vortex meter digitalyewflo Vortex meter Series Bulletin 01F06A02-00E www.yokogawa.com/fld Series Simple body construction for easy installation and high level of safety 2 Standard Type / Multi-variable
More informationGeneral Specifications
General Specifications Model PH20, FU20 and FU24 analog 4 in 1 ph sensor A family of wide body sensor is available for application in a wide variety of processes. The sensors share the same valuable features:
More informationCombination ph/redox (ORP) Sensors
Data Sheet SS/AP00_ Combination ph/redox (ORP) s AP00 Series User-customizable for Insertion, Flow-through, Immersion and Hot-tap Retractor ideal for OEMs (Original Equipment Manufacturers) and general
More informationVariable V Area Flowmeter Rotameter The Original
Variable V Area Flowmeter Rotameter The Original precise measurement of liquids and gases robust and universal proven device with an installed base of more than 500,000 worldwide a wide product range many
More informationBolt Type Test Plug - Technical Specifications
EXPANSION SEAL TECHNOLOGIES DC8048 05/98 REV 2 12/04 Bolt Type Test Plug - Technical Specifications Tube ID Shaft Shaft Shaft/Port "A" LI L2 L3 Max. Op. Shaft Hex Hex Nut Approximate Range OD Quantities
More informationInstruction Manual YOKOGAWA. ph/orp. Model FD20 Immersion Fittings. IM 12B06K02-01E-E 9th edition. IM12B06K02-01E-E_ed09.indd 1 07/03/16 09:51
Instruction Manual Model FD20 Immersion Fittings ph/orp YOKOGAWA 9th edition IM12B06K02-01E-E_ed09.indd 1 07/03/16 09:51 2 IM12B06K02-01E-E_ed09.indd 2 07/03/16 09:51 Contents 1. Introduction 5 1.1 Features
More informationGeneral Specifications
General Specifications GS 22B01C53-01EN-A M-675, M-775 and M-678 5-Valve Manifolds for Natural Gas Differential Pressure INTRODUCTION The PGI 5-Valve manifolds are designed for natural gas service and
More informationRosemount 402 and 402VP
Rosemount 402 and 402VP Contacting Conductivity Sensors Instruction Manual LIQ-MAN-402 Rev. M May 2017 hasgkas Essential Instructions Read this page before proceeding! Emerson designs, manufactures and
More informationAP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications
Data sheet DS/AP00 EN Rev. F AP00 Series Measurement made easy A reliable economic solution for a wide range of applications Long life large PTFE junction provides resistance to fouling double junction
More informationRetractable ph/orp Sensors
Model 396R and 396RVP Instruction Manual PN 5100396R January 2001 Retractable ph/orp Sensors ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests
More informationGeneral Specifications
General Specifications Model EXA Series Compact ph Probes PF20, PD20 and PS20 EXA Compact is a versatile series of probes suitable for ph measurements in all common applications in industrial processes
More informationModels 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor
Models 396/396P/396R Product Data Sheet PDS 71-TUpH September 1999 Retraction/Submersion/Insertion ph/orp Sensor MINIMUM SENSOR MAINTENANCE due to patented TUpH reference technology 1 which ensures steady
More informationProcess to Instrument Valves: C13SF Monoflanges For DPharp EJX & EJA-E Pressure Transmitters
Process to Instrument Valves: C13SF Monoflanges For DPharp EJX & EJA-E Pressure Transmitters Introduction Introduction The partnership of Yokogawa and AS-Schneider creates a real added value to our customers.
More informationGeneral Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.K February 2013
Product Data Sheet PDS 71-400/rev.K February 2013 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Factory-measured cell constant ensures out-of-the-box
More informationFirehawk Responder Second Stage Regulator
Firehawk Responder Second Stage Regulator MAINTENANCE AND REPAIR MSA 511 (L) Rev. 4 MSA 2015 Prnt. Spec. 10000005389(I) Mat. 10096394 Doc. 10096394 Replacement Kits and Overhaul Kit P/N Description Firehawk
More informationAnalytical Instruments Two-Electrode Conductivity Sensors
Data Sheet Analytical Instruments Two-Electrode Conductivity Sensors Model TB2 Superior accuracy in low conductivity measurement ranges. Measurement ranges from 0 to 199.9 µs/cm to 0 to 19.99 ms/cm. Resolution
More informationConductivity Sensors. Model 140, 141, 142. Product Data Sheet series/rev.b February 2011
Product Data Sheet 71-140 series/rev.b February 2011 Model 140, 141, 142 Conductivity Sensors RUGGED CONSTRUCTION withstands high temperature and pressure TWO MOUNTING CONFIGURATIONS: screw-in and retractable
More informationRetractable ph/orp Sensors
Instruction Manual PN 51-396R/rev.J February 2011 Model 396R and 396RVP Retractable ph/orp Sensors ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and
More informationShop Press. Maximum Capacity: 25 Tons and 55 Tons SAFETY PRECAUTIONS
Operating Instructions for: Form No. 102481 1826 1845 1846A 1847 1872 1872-220v 211200 576780 60361 61275 D01008AA D01009AA WARNING: To prevent personal injury; Shop Press Maximum Capacity: 25 Tons and
More informationCirculation Kits for CP Pumps
Instruction Sheet P/N 08000C Description This instruction sheet covers two circulation kits for CP Pumps: Stainless steel circulation kit, with Viton O-rings Stainless steel circulation kit, with EPR O-rings
More informationOperation and Maintenance Manual
Operation and Maintenance Manual Piston Actuated Air Operated Valve for c1s/c2s/o1s/o2s model operators 10V SW 10SM 20SM 30VM 40VM 60VM 100VM 150V series valves Parker Autoclave Engineers Instrumentation
More informationSIGNET 2760 DryLoc ph/orp Preamplifier
SIGNET 2760 DryLoc ph/orp Preamplifier 3-2760.090 3-2760.090 (A-3/04) English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confirm chemical compatibility before
More informationLiquid Conductivity Two & Four Electrode Sensors
Industrial Process Control & Monitoring Rugged Two Electrode Conductivity Sensors Unique Four Electrode Conductivity Sensors Versatile Installation Options (Threaded, Quick Change, Retractable) Specialized
More informationDEFENDER SERIES. 5 Gallon, Double Walled, Field Replaceable Spill Container Model Series INSTALLATION, OPERATION, & MAINTENANCE
DEFENDER SERIES 5 Gallon, Double Walled, Field Replaceable Spill Container Model 705-550 Series INSTALLATION, OPERATION, & MAINTENANCE Franklin Fueling Systems 3760 Marsh Rd. Madison, WI 53718 USA Tel:
More informationRetractable ph/orp Sensors
Model 398R & 398RVP Instruction Manual PN 5100398R February 2001 Retractable ph/orp Sensors MODEL 398R MODEL 398RVP (shown in ball valve) ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount
More information351GF-15. Hydrant Coupler INSTALLATION / OPERATION / MAINTENANCE MODEL
INSTALLATION / OPERATION / MAINTENANCE MODEL 351GF-15 Hydrant Coupler 351GF-15 Hydrant Coupler This installation/operation/maintenance guide is designed to provide instructions for the installation operation
More informationConductivity Sensors. Applications. 140 Series Conductivity Sensors Product Data Sheet LIQ_PDS_140_141_142_Conductivity_Sensors January 2014
140 Series Conductivity Sensors Product Data Sheet LIQ_PDS_140_141_142_Conductivity_Sensors January 2014 Conductivity Sensors Rugged construction withstands high temperature and pressure Two Mounting Configurations:
More informationRetractable ph/orp Sensors
Instruction Manual PN 51-398R/rev.G January 2011 Model 398R & 398RVP Retractable ph/orp Sensors MODEL 398R MODEL 398RVP (shown in ball valve) ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount
More informationVibra-Proof Condenser Plugs. The Condenser Plugs That Won t Vibrate Loose.
Vibra-Proof Condenser Plugs The Condenser Plugs That Won t Vibrate Loose. Vibra Proof Condenser Plugs CPI Plugs Perma Plugs P2 High Pressure Plugs Ram Packages Tube Stabilizers Manual Installation Tools
More informationGeneral Specifications
General Specifications Model RAGK Rotameter GS 01R01B07-00E-E This type of Rotameter is designed for measurement of liquids and gases. The conical glass metering tube has a free rotating float. This float
More informationInstruction Manual PN /rev.H March Model 385+ ph/orp Sensor
Instruction Manual PN 51-385+/rev.H March 2006 Model 385+ ph/orp Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet
More informationIso-Flo Coupling Repair
Instruction Sheet P/N 229782C Iso-Flo Coupling Repair Introduction This instruction sheet covers the repair of the male and female couplings used in the Automatic and Manual Iso-Flo HD Voltage Block Systems,
More informationInstallation Introduction im Vertrieb von CAMBOARD Electronics The Flexible Conduit Kit is a 6-foot length of electrical conduit with three 7-½ foot,
im Vertrieb von CAMBOARD Electronics User s Guide Flexible Conduit Kit Replacement Add-On for the Cable Cubby Power Box 68-734-01 Rev. A Printed in the USA 03 03 Installation Introduction im Vertrieb von
More informationPumpAgents.com - Click here for Pricing/Ordering INSTALLATION * Crusader is a registered trademark of Thermo Electron Corporation.
PumpAgents.com - Click here for Pricing/Ordering Model 42730-0000 MARINE ENGINE COOLING WATER PUMPFEATURES FEATURES Body: Impeller: Shaft: Ports: Seal: Bearings: Shipping Weight: APPLICATION Bronze Jabsco
More informationTechnical Manual MSI Hydraulic Adjustable Choke
Technical Manual MSI Hydraulic Adjustable Choke MSI A Division of Dixie Iron Works, Ltd. 300 W. Main St. Alice, TX 78332 www.diwmsi.com (800) 242-0059 Revision A TABLE OF CONTENTS SECTION 1 WARNINGS...
More informationModel 385+ Instruction Manual PN February ph/orp Sensor
Model 385+ Instruction Manual PN 5100385+ February 2001 ph/orp Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet
More informationTerminator ZP/FAK-4L Kit Electrically Heated TubeTrace Bundles
Terminator ZP/ Kit Electrically Heated TubeTrace Bundles INSTALLATION PROCEDURES Listed Cable System 137M Terminator ZP/ The following installation procedures are suggested guidelines for the installation
More informationPerformance Series ph/orp Sensors
Industrial Process Control & Monitoring Patented Axial Ion Path Reference Specialized ph Glass Formulations and ORP Electrodes Proprietary Low-Noise, High Temperature Signal Cable Sensors are Compatible
More informationCombination ph/redox (ORP) Sensors
User Guide IM/AP300_6 Combination ph/redox (ORP) Sensors AP300 Series ABB The Company We are an established world force in the design and manufacture of instrumentation for industrial process control,
More informationInstallation Instructions
c Installation Instructions Sequential Loop -Position Electronic Pressure Regulator Kits This manual provides instructions for Cascade Electronic Pressure Regulator and Indicator System. The kits provide
More informationENDURANCE Conductivity Sensors
Model 400 Series Instruction Manual P/N 5100400 May 1999 ENDURANCE Conductivity Sensors Model 400 Model 401 Model 402 Model 403 Model 404 Screw-in/Insertion Sensors Screw-in/Insertion Sensors Retractable
More informationMODEL BARE REEL SERIES E
MODEL 84275 BARE REEL SERIES E SPECIFICATIONS Maximum working pressure 5000 p.s.i. Material Inlet 1/2 NPT (Female) Material outlet 1/2 NPT (Female) Sheave width 10 Sheave diameter 19-1/4 Drum diameter
More informationSignet DryLoc ph/orp Electrodes
Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature
More informationACCESSORIES & FITTINGS 800 SERIES CONTROL VALVES
ACCESSORIES & FITTINGS 800 SERIES CONTROL VALVES 8 VALVE FITTING PART NUMBERS 2" & 3" PRESSURE REDUCING VALVES* 10645 8970-002 (9526-006) 1/4" tube (9526-006) Flow 8970-001 (9526-001) 8970-010 (9526-010)
More informationLEADER SMART DRAW-OFF
LEADER SMART DRAW-OFF Leader Evaporator Co., Inc. 49 Jonergin Drive Swanton, Vermont 05488 (802) 868-5444 www.leaderevaporator.com Contents INTRODUCTION... 3 EQUIPMENT DESCRIPTION... 3 OPTIONAL SETUP AND
More information7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes
7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes 70-82-25-44 Rev. 1 10/07 Honeywell Process Solutions Copyright, Notices, and Trademarks Copyright 2007 by Honeywell
More informationMP461SC Flex Probes* Installation Instructions
PNEUMERCATOR Liquid Level Control Systems MP461SC Flex Probes* Installation Instructions Model MP461SC Magnetostrictive Flex Probes For use with the following consoles: TMS2000 TMS3000 NOTE: MP461SC PROBES
More informationTUpH Retraction/Submersion/Insertion ph/orp Sensors
Product Data Sheet PDS 71-TUpHVP/rev.B June 2003 Models 396VP, 396RVP, 398VP, & 398RVP TUpH Retraction/Submersion/Insertion ph/orp Sensors SILCORE 3 TECHNOLOGY PROVIDES INCREASED SENSOR LIFE when used
More informationMANUAL BACKUP WRENCH OPERATIONAL AND SPARE PARTS MANUAL
Since 1968 MANUAL BACKUP WRENCH OPERATIONAL AND SPARE PARTS MANUAL CONTENTS MANUAL BACKUP WRENCH OPERATIONAL AND SPARE PARTS MANUAL INTRODUCING HYTORC 3 SECTION I OPERATING INSTRUCTIONS 4 1-1 Link Selection
More informationMP56xSA Flex Probes* Installation Instructions
PNEUMERCATOR Liquid Level Control Systems MP56xSA Flex Probes* Installation Instructions Model MP561SA, MP562SA, MP563SA And MP564SA Magnetostrictive Flex Probes Inventory Management Of Tanks Up To 70
More informationChemical Injection Pump
TEXSTEAM MX3 SERIES Chemical Injection Pump TEXSTEAM INC. Chemical Injection Pump Product Features Up to 62 gallons (235 liters) per day Lightweight Discharge pressures to 10,000 PSI (690 bars) Adjustable
More informationMicro Motion F-Series Coriolis Flow and Density Sensors. Installation Manual , Rev CC December 2010
Micro Motion F-Series Coriolis Flow and Density Sensors Installation Manual 20002298, Rev CC December 2010 Safety and approval information This Micro Motion product complies with all applicable European
More informationMODEL 1385 & 1385-H Professional Rotary Drum Pump
NOV - 2010 Form 404168 Section - B5 Page - 47 OWNER/OPERATOR RESPONSIBILITY All instructions and warnings contained in this manual shall be read and understood by the owner/operator prior to operating
More informationBEFORE USING THE PRODUCT
HL-17015-5 BEFORE USING THE PRODUCT Thank you for purchasing an Oriental Motor product. Please read this BEFORE USING THE PRODUCT and the separate manual entitled OPERATING MANUAL thoroughly to ensure
More informationOPERATING INSTRUCTION MANUAL
OPERATING INSTRUCTION MANUAL Model 3422-series Compression Fitting Style Contacting Conductivity/Resistivity Sensors (Enhanced Performance Design) Worldwide Sales: GLI International, Inc. 9020 West Dean
More informationOPERATING INSTRUCTIONS. HTL-DSX Dedicated Square Drive Wrench
OPERATING INSTRUCTIONS HTL-DSX Dedicated Square Drive Wrench CONTENTS Contents 2 Safety 2-3 Warranty 3 Connecting the tool 3-4 Setting the torque 4-5 Operating the wrench 5-6 Maintenance 6 Troubleshooting
More informationPerformance Series ph/orp Sensors
Industrial Process Control & Monitoring Patented Axial Ion Path TM Reference Specialized ph Glass Formulations and ORP Electrodes Proprietary Low-Noise, High Temperature Signal Cable Sensors are Compatible
More informationVALVE AND PLUMBING KIT 2409 LOADER CUB CADET & KIOTI TRACTORS
ASSEMBLY MANUAL Keep With Operator s Manual VALVE AND PLUMBING KIT 2409 LOADER CUB CADET & KIOTI TRACTORS TRACTOR MODELS CUB CADET 8404, 8454 KIOTI DK45, DK50 ROPS X X TRACTOR AND VALVE KIT GENERAL INFORMATION
More informationEXPANSION SEAL TECHNOLOGIES DC /01 REV 2 06/05 PAGE 1 OF 4
EXPANSION SEAL TECHNOLOGIES DC9567 11/01 REV 2 06/05 PAGE 1 OF 4 OPERATING PROCEDURES FOR GRIPTIGHT HIGH PRESSURE TEST PLUGS WITH ALTERNATE SEAL MATERIALS Description: GripTight plugs supplied with alternate
More informationInstallation Instructions
c Installation Instructions -Position Electronic Pressure Regulator Kits This manual provides instructions for Cascade Electronic Pressure Regulator and Indicator System. The kits provide selectable clamp
More informationDisassembly and Reassembly for WG Series Worm Gear Operator. User Instructions Part Number: XXXXXX, Rev. 0 Release: February 2017
Disassembly and Reassembly for WG Series Worm Gear Operator User Instructions Part Number: XXXXXX, Rev. 0 Release: February 2017 User Instructions Doc. Number: XXXXXXXX Table of Contents February 2017,
More information888 SMITH SYSTEM USER S GUIDE WARNING: Read and follow all directions for each step to insure proper assembly of this product.
888 SMITH SYSTEM USER S GUIDE WARNING: Read and follow all directions for each step to insure proper assembly of this product. CLASS H PART # 733970 REV. B Version: 88809 Revision: 0/09/03 TABLE OF CONTENTS
More informationINSTRUCTION MANUAL MODEL S80 INTELLIGENT SENSORS
INSTRUCTION MANUAL MODEL S80 INTELLIGENT SENSORS PREFACE Purchasing products from Electro Chemical Devices, Inc. provides you with the finest liquid analytical instrumentation available. If this is your
More informationSignet ph/orp Electrodes
Signet 2724-2726 ph/orp Electrodes Compatible with ALL Signet ph/orp Instruments Features Flat Glass Protected Bulb Patented reference design for exceptional performance * Mounts in Signet standard fittings
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. TBX5 ph, Redox (ORP) sensors with diagnostics
BB MESUREMENT & NLYTICS DT SHEET TBX ph, Redox (ORP) sensors with diagnostics Measurement made easy The most durable ph / Redox (ORP) sensors in the world Next Step Solid State reference eliminates poisoning,
More informationDRAFT. Installation, Operation and Maintenance Instructions for Emerson KBG-V Series
Part Number: KBG-V.IOM.1001, Rev. 0 Release: February 2017 Installation, Operation and Maintenance Instructions for Emerson KBG-V Series DRAFT Part Number: KBG-V.IOM.1001, Rev. 0 Table of Contents February
More informationMERCER VALVE 8100 SERIES THINK...MERCER FIRST MERCER VALVE CO., INC. AUTO SEAT TECHNOLOGY
MERCER VALVE CO., INC. AUTO SEAT TECHNOLOGY 8100 SERIES MERCER VALVE THINK...MERCER FIRST 8100 Series Product Overview The Mercer Valve 8100 Series Pressure Relief Valve is State of the Art in soft seat,
More informationRating when used as a weight distributing hitch with spring bars:
January 2007 INSTALLATION INSTRUCTIONS 70210 70220 70230 Bolt Together Weight Distributing Hitch System Rating when used as a weight distributing hitch with spring bars: Part Number Max. Tongue Wt. Max
More information* * Rev. 9 01/18 Product Manual
English Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. 9 01/18 Product Manual Description The Signet 2760 ph/orp Preamplifiers with DryLoc connectors are part of Signet's growing line
More informationSignet DryLoc ph/orp Electrodes
Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature
More informationHYDRAULIC SPREADER. Max. Pressure: 10,000 PSI (700 BAR) Unit Weight: HS2000 = 4.8 LBS. (2.17 Kg) HS3000 = 22 LBS. (9.98 Kg)
Form No. 103525 Operating Instructions for: HS2000 HS3000 HYDRAULIC SPREADER Max. Pressure: 10,000 PSI (700 BAR) Unit Weight: HS2000 = 4.8 LBS. (2.17 Kg) HS3000 = 22 LBS. (9.98 Kg) Definition: Hydraulic
More informationOPERATION AND MAINTENANCE MANUAL
WREN IBT SERIES HYDRAULIC TORQUE WRENCHES IBT SQUARE DRIVE SERIES OPERATION AND MAINTENANCE MANUAL FOR WREN Products: POINT 75, 1IBT, 3IBT, 5IBT, 8IBT, 10IBT, 20IBT, 25IBT, 35IBT, 50IBT SQUARE DRIVE HYDRAULIC
More informationReplacing the Meter Body / Process Head Parts
Replacing the Meter Body / Process Head Parts ST 700 and ST 800 Pressure Transmitters and SMV800 Multivariable Transmitter - Instruction sheet Document Number: 34-ST-33-70 Effective Date: 5/27/2016 (Process
More informationInstallation Instructions Skylight FlexShades by Draper
Caution 1 Read entire procedure before installing. Keep instructions for future reference. 2 Tools needed for installation: Cordless drill/screwdriver, with flathead, phillips, and 1 /4" nut driver attachments
More informationHead Removal Detection Function for Safer Performance
HS5D Miniature Interlock Switches Head Removal Detection Function for Safer Performance (00405) New Generation Miniature Interlock Switches IDEC s new unique Head Removal Detection Function is a new generation
More informationType N550 Snappy Joe Emergency Shutoff Valves
Instruction Manual MCK-1149 Type N550 March 2010 Type N550 Snappy Joe Emergency Shutoff Valves Failure to follow these instructions or to properly install and maintain this equipment could result in an
More informationPerformance Series ph/orp Sensors
Industrial Process Control & Monitoring Patented Axial Ion Path TM Reference Specialized ph Glass Formulations and ORP Electrodes Proprietary Low-Noise, High Temperature Signal Cable Sensors are Compatible
More information