CEMTEK Instruments- NH3 Analyzer Quick Start Manual

Size: px
Start display at page:

Download "CEMTEK Instruments- NH3 Analyzer Quick Start Manual"

Transcription

1 CEMTEK Instruments- NH3 Analyzer Quick Start Manual November 2013

2 Table of Contents Section 1- General Overview 1.1- Overview Important Precautions Theory of Operation Performance Specifications Features Mechanical Specifications Electrical Specifications Section 2- Configuration and Operation 2.1- Pre-Setup Details Initial Setup Procedure Touch Screen Menu Navigation Calibration Modifying/Checking Analyzer Diagnostics Preventative Maintenance Section 3- System Description 3.1- Warnings/Alarms Troubleshooting Important System Components Appendix Controller Enclosure Mounting Dimensions i Controller Enclosure Layout ii Probe Enclosure Layout. iii Full System Layout iv ADAM wiring Layout v Cable Interconnect Wiring vi-vii Air Clean Up System Details.....viii-xi Contact Information xii 1

3 1.1- Overview The CEMTEK Instruments NH3 Hot/Wet Extractive NH 3 Analyzer is designed specifically for use in Combustion Process applications for Low NO x Burner Tuning and SCR Inlet/Outlet monitoring. NH 3 is measured using a Tunable Diode Laser. The system requires no sample conditioning and is simply heated throughout the sample s flow scheme to ensure NH3 is not lost from the sample gas Important Precautions Use dry, oil-free instrument air only Caution: This instrument will be damaged if used with instrument air that is not dry and oil-free. Ensure instrument air has 30 F dew point and has been filtered to remove all oil and particulates. Safety Notice This instrument operates from potentially lethal line voltage. In addition, certain components operate at high temperatures and can cause serious burns. Observe all precautions when using this device, and particularly be sure that all devices connected to the instrument are safely wired and properly grounded. Always disconnect power to the instrument before making any wiring adjustments. Caution: The analyzer should not be operated without each enclosure door shut and the fan fully functioning Theory of Operation With NH3 being a difficult compound to measure due to the physical properties created in combustion monitoring gas applications, our system looks to keep the sample gas at or above 190C throughout each component that our sampling system consists of, including most importantly, our high temperature tunable diode laser bench. In addition, we have a shortened sample flow scheme to lessen the chance of having our sample gas compromised, ensuring correct readings. 2

4 1.4- Performance Specifications Measurement technology Measurement range (adjustable) Zero noise Zero calibration drift Span noise Span calibration drift Linearity error Response time NH3 Tunable Diode Laser PPM PPM Better than ± 1 PPM < 1% of reading Better than ±2% of reading < ±2% of high calibration value across range from zero to full scale T90 < 30 seconds 1.5- Features Proven TDL technology for detection, accuracy, precision & long life span Simple Hot/Wet Extractive gas sampling system Automated Probe Enclosure climate control Front panel Touch Screen Display and Control Automatic Calibration mode option Automatic Calibration check option 4-20 ma analog output and Good Data Digital Circuit 1.6- Mechanical Specifications Probe Enclosure: NEMA 4 mounted directly to the flange 16 W x 12 D x 20 H Probe: SS with 3 ANSI 150lb flange Controller: Wall Mount NEMA4-12 W x 10 D x 16 H Total Weight: ~70 lbs depending on probe 3

5 1.7- Electrical Specifications AC input (supplied by customer): - Gauge wire: AWG VAC Output Contacts (supplied by customer): - Gauge wire: AWG Minimum rating 12V, 60mA Maximum rating 12V, 3.5A 2.1- Pre-Setup Details Getting started using your new CEMTEK Instruments analyzer should be quick and easy. In this section we have included the information needed for standard installations only. To ensure the quickest and most reliable startup, please follow the subsequent steps. The system consists of 3 major components: The Controller Enclosure which houses the majority of the electronics and pneumatic controls. The Probe Enclosure which contains the heated pump, laser bench, heated enclosure tube section, and heated probe. These components make up the sampling system. Enclosure Interconnects- Power cable, Signal cable, Calibration line, and Air line. 4

6 Ensure all required services such as power supply, instrument air, and signal cabling are available on-site prior to installation. * Once the sample location has been identified, mount the probe at the sample port. * Next mount the controller appropriately so that the sample line can be run between the controller and probe enclosure without danger of being damaged upon completion of installation. ** A basic layout of the NH3 Analyzer System can be seen on the following page ** 5

7 6

8 2.2- Initial Setup Procedure 1. Connect outer controller gas lines 1. Make sure that the analyzer is NOT powered on. 2. Connect instrument air and span calibration gas to inlets shown below via 3/8 and 1/4 Swagelok fittings respectively. a. Instrument air: must be dry, oil-free air, 8-10 SLPM at 60-70PSI. b. Calibration gas: Should be at 8-10 SLPM. * For the system, here are all the input pneumatic connections. All located going into the Controller Enclosure. 7

9 2. Ensure inner controller components and pneumatic lines are connected 1. Check to see if all internal wires are landed and in their correctly associated terminals and that there are no loose or exposed wires seen. 3. Connect analog and digital customer connections into correct ADAM module terminal locations 1. Connect analog and digital output signals that are brought in through the conduit labeled signal cable. (See below) Analog connector: ADAM-4021 (4-20mA Output) +IOUT, -IOUT Digital connector: Good Data (NORMALLY OPEN) - Good Data defined as analyzer is pulling sample with pump on, no temperature alarms, no pneumatic valves open. RL6 NO, RL6 COM (CLOSED = Bad Data) Analog Connection Digital Connection 8

10 2. Connect the 120 VAC power supply input (from customer) through the other ¾ conduit labeled AC IN into the proper locations on the power distribution terminal blocks. Analyzer Circuit Breaker: Switch up/down supplies/severs power to system. 4. Connect Enclosure Interconnect Cable and Gas lines 1. Connect the signal cable and power cable to their respective bulkheads. 2. Connect the 3/8 Air line and ¼ Calibration gas line to their respective tubing bulkheads. 9

11 5. Apply power and begin application specific analyzer setup 1. Apply power to the system by either flipping the circuit breaker and/or plugging the external power supply in. 2. After half a minute the NH3 concentration will be shown on the lock screen. This is the screen that should be seen once analyzer has gone through initial power up procedure. 3. Next go into the Temperature Setup screen and turn all heaters on. Let each heater begin to regulate at their associated temperature set points. The System Temperatures screen will show what each temperature set point is and what the high and low temperature alarms are set at. Adjust the temperature limits and set points accordingly in terms of your analyzer s application. Note: If set point temperatures are changed, check/adjust flow balance regulators and recalibrate analyzer. 10

12 The Main Menu Screen for analyzer set up and control. 4. Set the Full-scale value for the analyzer Full-scale value for the NH 3 channel can be set at any integer value between 1 and 500 ppm. To adjust the output scaling, click on the Other Parameters button via the menu screen and then set the Full Scale to the appropriate value. Note: The analog output is scaled to the NH 3 full-scale of the analyzer. For example, if the full-scale of the analyzer is set to 50 ppm, then 20 ma will represent 50ppm. (0ppm=4mA) 5. Manually calibrate the analyzer After installation and at least a 1-hour warm up period, the analyzer can be calibrated following the procedure outlined in the "Calibration" Section of

13 2.3- Program Screen Navigation - When the analyzer is functioning the touch screen display makes it easy for the user to traverse to many different areas dealing with analyzer controls and diagnostics. Here is the program s runtime screen hierarchy. 12

14 2.4- Calibration After installation and at least a 1-hour warm up period, the NH3 channel can be calibrated via the Calibration Procedure as listed below: Initial Calibration: 1) Follow the Calibration Procedure outlined below which is executed from the Manual Calibration screen. 2) By utilizing the Trending NH3 Chart, the response time for zero and span gases can be timed, which assist in approximating the amount of time calibration gas should be run on the analyzer. 3) It may be necessary to do a second calibration to stabilize the analyzer after successfully completing the first procedure fully. Auto Calibration: 1) Setting up an auto calibration sequence can be accomplished through the Auto Cal Setup screen. Follow the steps as stated in section 2.5 Modifying/Checking Auto Cal Settings. Notes on Calibration Frequency * For normal process applications following the maintenance and calibration procedures listed in Section 2.6 will provide a level of accuracy acceptable for process monitoring. * For applications where the NH 3 value is being used as a feedback number for SCR control then calibrating at least on a weekly basis is recommended. * For applications whereby the NH 3 numbers are being utilized as a means of proving the process system s ability to meet the permitted limits (without the requirement for certification) calibration drift of the analyzer should be monitored- When readings obtained are outside the acceptable limits, a recalibration should be performed. 13

15 Calibration Procedure In order to calibrate the analyzer s zero and span values the user must click on the manual calibration button from the home screen. NOTE! * Before calibration of the analyzer ensure that the calibration gas bottle value saved is correct. * Make sure that the associated bottle value is within the range of the fullscale output value set. The screen displayed above shows the following: The calibration gas bottle value for its span calibration. The software s stored response value, which is shown and saved to provide a correctly calibrated Live NH3 PPM reading. The Raw Laser PPM value which is the raw response as outputted from TDL bench. 14

16 In order to perform a manual calibration, follow the procedure below starting with manually calibrating the zero gas (OFFSET), followed by performing a calibration with the span gas (SPAN) 1. Open the zero or span valve to initiate either the zero or span calibration of the analyzer. (make sure only one valve is open at a time) 2. Once the live response has settled (use the Trending NH3 screen to help determine this) you may select Calibrate Span/Zero. This will change the stored response value stored in the analyzer s software which is used to calculate the Live Response of the real-time sample gas. NOTE! Make note of the time it takes for the value to settle, as these will be used to set the automatic calibration sequence. (See Auto Calibration Setup Section 2.5) 3. After completing the calibration of the analyzer close the calibration gas valves and check to ensure the sample pump is on and Good Data is trending. NOTE! For the most accurate readings calibrate the zero and span of a gas back to back. 15

17 2.5- Modifying/Checking System Diagnostics Modifying/Checking Full Scale & NH3 Trend Chart- Menu Screen -> Trending NH3-> From this screen you can: Modify or check the trending graph s scale. See how the Live NH3 ppm has been trending. Navigate to the Manual Calibration or Chart Options screens. 16

18 Modifying/Checking Calibration values- Main Menu-> Manual Cal -> From this screen you can: Modify or check bottle or stored response values associated with the span and zero of the sampling gas. Open or close the span or zero gas valves. Capture the Span or Zero response, also known as manually calibrating the analyzer. View the raw or live response readings. Go to Trending NH3 screen. 17

19 Modifying/Checking Chart Options- Main Menu-> Trending NH3-> More Chart Options From this screen you can: Modify or check values associated with the number of rolling data points set for the Live Reading NH3 calculation. Modify or check the Charts time range hours and/or minutes. 18

20 Modifying/Checking Auto Calibration Settings Main Menu-> Auto Cal Settings-> From this screen you can: Modify or check the upcoming hour (0-24) to begin 1 st Auto Cal Run. Modify or check the number hours between Auto Cals. Check the time and date in which the next Auto Calibration will run. Modify or check the Duration in minutes that the span and zero valves will be open. Enable or Disable the Auto Calibration mode. 19

21 Modifying/Checking System Temperatures- Main Menu-> System Temperatures-> From this screen you can: Modify or check the over values under values, and/or set point values of each component s temperature limits. Check the Live Temperature Readings of each heater component. Check the Ambient Outside air temperature and inside controller temperature. Turn on/off each heater. 20

22 Modifying/Checking System Parameters- Main Menu-> Other Parameters-> From this screen you can: Modify or check the heater regulating temperature range, temperature set points, offset value, span value, and Full Scale output value. Check Heater Power Relay functioning. Go to Software Tags screen. 21

23 Modifying/Checking Software Tags- Main Menu-> Other Parameters-> Software Tags From this screen you can: Check Laser bench parameters. Override the pump, laser controls, and Good data variables. Change Cooler Valve Set point, Cal Wait Time, or Laser Range Filter values. 22

24 2.6- Preventative Maintenance Once the analyzer is correctly configured and functioning as intended, it is crucial to implement periodic checks to avoid problems causing analyzer down time. The type and amount of maintenance required varies upon application. It is dependent on the concentration of the measured gas, particulate loading, moisture, and other sorts of relevant flue gas constituents. Yearly basis: Check/Clean/Replace as necessary the calibration shroud and probe tip-end filter. Check Linearity of TDL bench by running a mid-range gas value after a calibration procedure is performed. -Contact CEMTEK if sensor linearity procedure must be done. Replace/Check sample lines for contamination. Quarterly basis: Check the amount of air/gas flow while instrument air/cal gas is flowing through the analyzer. Adjust the flow balance regulators as necessary. Check/Clean the Exhaust orifice at the probe enclosure. Monthly/weekly basis (application dependent): Check for alarms and system abnormalities. Check Air and Calibration gas system pressures. Run analyzer calibration check. Manually recalibrate the analyzer if necessary. ** On the following page are acceptable log sheets that can be used to carry out monthly/weekly, quarterly, and yearly preventative maintenance checks. 23

25 NH3 Analyzer MONTHLY/WEEKLY CHECKS Alarms? YES NO Plant inlet air pressure(psi) psi psi OK Air cleanup panel psi(exit) psi psi OK Calibration Gas Bottle pressure >150psi psi OK *must recalibrate analyzer if pressures and/or flows are changed. NH3 zero air check value (PPM) 0 Pass/Fail NH3 zero stored % value % NH3 span check value (%) Bottle Val: Pass/Fail NH3 span stored % value % NH3 Analyzer QUARTERLY CHECKS CALIBRATION GAS FLOWS: NH3 span gas flow 3 LPM lpm OK NH3 zero gas flow 3 LPM lpm OK ANALYZING SAMPLE GAS: Calibration gas flow 0 LPM lpm OK Probe Temperature see set point C OK Enclosure Section Temperature see set point C OK Pump Temperature see set point C OK Enclosure Temperature see set point C OK Check/Replace Probe Filter Check/Clean Pump Exhaust Check/Clean Enclosure Section line OK OK OK *must recalibrate analyzer if flows, parts, or temperatures are altered. 24

26 8000 NOX-O2 PROBE YEARLY CHECKS Check/Clean all lines for contamination Check/Clean as necessary the calibration shroud and probe tipend filter Check Linearity of TDL bench by running a mid-range gas value of NH3 after a calibration procedure is performed. Contact CEMTEK if sensor linearity procedure must be done OK OK OK *must recalibrate analyzer if sensor parts or lines are cleaned. 25

27 3.1- Warnings/Alarms Warnings and Alarms are displayed atop each screen (as shown below). PUMP OFF- There is no sample being drawn through the analyzer. The sample pump may be turned off or a temperature alarm thrown. TEMP ALRM- Alarm displayed when a heater temperature is either over or under a specified limit or not turned on. See touch screen menu navigation section to view alarm details. Many need to troubleshoot. VALVE OPEN- States that the Span or Zero gas valve is open, should be seen during manual or automatic calibrations. See touch screen menu navigation section to close and open this valve. AC- States that system has been placed in Auto Calibration Mode. See Auto Calibration Section on how to configure this mode. M- States that the system has been placed in Maintenance mode. Must be placed in this mode to signal Good Data is NOT being transmitted at the time and to place the system in/out of Auto Calibration mode. 26

28 3.2- Troubleshooting Once the analyzer has been set up correctly through the Initial Setup Procedure and has been functioning acceptably beyond start up issues, situations in which the analyzer is not functioning properly may arise. Below are descriptions of possible issues that may be experienced and the correct steps to take in troubleshooting them. Analyzer Touch screen not powering on: Blank Touch screen - Check wiring to analyzer and into back of HMI display. - Check analyzer circuit breaker switch. - Replace screen. Unresponsive NH3 reading: When a change in NH3 ppm value should be seen (such as when switching from zero to span gas during a calibration check) - Check flow scheme for leaks and loose fittings. - Make sure there is some type of range between the stored % response of the zero and span calibration values on the manual calibration page. - Validate proper calibration gas input conditions as specified in the initial set up section. - Check that the pneumatic block solenoid valves are all functioning correctly. a. While monitoring the flow meters, manually open/close the span and zero valves via the touch screen on the calibration page. - Check state of filters in filter assembly. -Check pump exhaust flow at probe. a. If there is not sufficient air flow out of the pump there may be a problem with your pump or a clogged line (within sampling system up at probe). Blow out all exhaust lines. 27

29 Over/Under Temperature Alarm: Will be seen atop the main screen - Go to system temperature page and confirm set points for alarms and regulating points are correct. - Check Heater lines to main board wiring connections. - Check ADAM relay heater wiring connections. - Check Probe Enclosure wiring connections. 28

30 3.3- Important System Components PROBE ENCLOSURE COMPONENTS 1. Heated TDL Bench Rectangular black TDL bench assembly mounted inside Probe Enclosure in between the Enclosure Heated Line Section and Heated Sample Pump. The bench outputs a RS-485 protocol which communicates with the Advantech Touchscreen through signal cabling. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 24VDC into bench serial port and 120VAC into bench heater power supply supplying 2 12VDC lines running into green bench heater connector 2. Heated Extraction Probe Assembly Probe Assembly connected via flange protruding out of the Probe Enclosure into the gas stream that the analyzer is looking to pull sample from. The probe assembly consists of sampling orifice, probe filter, coil heater, and filter bushing assembly. 120VAC Coil Heater 29

31 3. Enclosure Heated Line Section Heated line that is temperature controlled located between the heated filter bushing assembly and heated TDL Bench. 120 VAC line heater 4. Heated Sample Pump Heated Sample Pump that is temperature controlled and responsible for drawing a heated sample gas through the analyzer s sampling system. 120 VAC heater power and 120 VAC pump power 30

32 5. Vortex Air Cooler The Vortex Air Cooler assists in controlling the Probe Enclosure s temperature set point. It will produce a cooled stream of air when a pressurized air stream is fed through its port via a controlling solenoid valve. 6. Enclosure Fan Assembly Cross ventilating fan that pulls air through Probe Enclosure to regulate the enclosure s temperature to a specified set point. Fan works in conjunction with lower louver plate vent. 120VAC fan power 31

33 CONTROLLER ENCLOSURE COMPONENTS 1.24 Vdc Power Supply This component supplies the Vdc power to the entire system. The input voltage supply is typically 120 Vac. And the output voltage of the component is 24Vdc. Located in the controller. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 120VAC power input 2. ADAM 4069 Module These two components provide the logical switching between different states of the programmed heating/switching functions within the system. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 24VDC power input 32

34 3. ADAM Module The module that is responsible for receiving each thermocouple pair from system heater components and relaying their signals to the Advantech Touchscreen. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 24VDC power input 4. ADAM 4021 Module ADAM module that is programmed to transmit a 4-20mA output signal of the NH3 ppm reading. Module communicates with the Advantech Touchscreen. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 24VDC power input 33

35 5. Advantech Touch Screen HMI Touchscreen HMI that is responsible for configuring each analyzer s programmable settings. Found on the outer face of the Controller Enclosure s hinged front door. See Component Manuals folder on disc manual for more information concerning troubleshooting and servicing this component further. 24VDC power input 6. Calibration Assembly Assembly that consists of the calibration flow meter, span/zero calibration valves, and various stainless steel fittings located in the controller enclosure. 120VAC valve power 34

36 7. System Air Pressure Regulator A pressure regulator that is responsible for setting and adjusting the incoming system air pressure. 8. Power Distribution DIN rail Assembly DIN rail which system circuit breaker, power supplies, and terminal blocks are mounted on. 35

37 APPENDIX APPENDIX CONTROLLER ENCLOSURE MOUNTING DIMS i

38 APPENDIX CONTROLLER ENCLOSURE LAYOUT ii

39 APPENDIX PROBE ENCLOSURE LAYOUT iii

40 APPENDIX Full System Layout iv

41 APPENDIX ADAM Wiring Layout v

42 APPENDIX Cable Interconnect Wiring Layout vi

43 APPENDIX vii

44 APPENDIX Cemtek Air Clean Up System (Box) - Type 1 viii

45 APPENDIX Cemtek Air Clean Up System (Panel) - Type 2 ix

46 APPENDIX Cemtek Air Clean Up System (with Heatless Dryer) - Type 3 x

47 APPENDIX Air Clean Up System Information Scrubber 1- Activated Carbon (black) Scrubber 2- Activated Aluminum (white) Scrubber 3- Purafil Media (purple) *See Supplied Material Safety Data Sheets for more information on these substances LB canisters contain each of these media - May need to be changed out due to the quality of the compressor plant air used (input air should be clean and dry). - Based on each specific application; media and air clean up components should be monitored accordingly. xi

48 APPENDIX Contact Information: CEMTEK INSTRUMENTS Telephone: (732) hr. Emergency Service: xii

Section 55 Chapter 6

Section 55 Chapter 6 Section 55 Chapter 6 REMOTE HYDRAULICS CONTROLLER Calibration and Fault Codes 6-12880NH TABLE OF CONTENTS REMOTE HYDRAULICS CONTROLLER CALIBRATION... 55-5 Requirements For Calibration... 55-5 Aux Set Main

More information

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATOR SYSTEM START-UP STEP 1 Walk around the skid and inspect for damage. Then check that all electrical

More information

O 2 CO CO 2 NO NO 2 NO x

O 2 CO CO 2 NO NO 2 NO x continuous emissions monitoring O 2 CO CO 2 NO NO 2 NO x FGA 900 Series Combustion & Environmental Monitoring FGA Continuous Emissions & Combustion Efficiency Monitoring The FGA range of analyzers use

More information

A6Z OPERATING MANUAL

A6Z OPERATING MANUAL A6Z OPERATING MANUAL TABLE OF CONTENTS Introduction... p. 2 Features... p. 2 Description... p. 3 Theory of Operation... p. 3 Installation... p. 4 Electrical Connections... p. 5 Options... p. 6 Warranty.p.

More information

A3Z OPERATING MANUAL

A3Z OPERATING MANUAL A3Z OPERATING MANUAL TABLE OF CONTENTS Introduction... p. 2 Features... p. 2 Description... p. 3 Theory of Operation... p. 3 Installation... p. 4 Electrical Connections... p. 5 Options... p. 6 Warranty...

More information

Observe all necessary safety precautions when controlling the soft starter remotely. Alert personnel that machinery may start without warning.

Observe all necessary safety precautions when controlling the soft starter remotely. Alert personnel that machinery may start without warning. Introduction OPERATING INSTRUCTIONS: MCD REMOTE OPERATOR Order Codes: 175G94 (for MCD 2) 175G361 + 175G9 (for MCD 5) 175G361 (for MCD 3) 1. Introduction 1.1. Important User Information Observe all necessary

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the CO level and outputs a 4-20 ma signal. The standard product features a 2- wire loop-powered output. Optionally, the device

More information

ZMT & ZFG2 Zirconia Oxygen Analyzer Systems. Superior technology and quality from the world leader in oxygen measurement

ZMT & ZFG2 Zirconia Oxygen Analyzer Systems. Superior technology and quality from the world leader in oxygen measurement Data sheet DS/ZMT/LO EN Rev. J ZMT & ZFG Superior technology and quality from the world leader in oxygen measurement Integral reference-air supply option avoids necessity for additional equipment Dual

More information

Section 3 Technical Information

Section 3 Technical Information Section 3 Technical Information In this Module: Engine identification Modes of operation Battery charging and heat manage operation Service and repair procedures Maintenance requirements Engine Identification

More information

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL Content 1. Introduction...2 2. Important safety instruction...3 3. System component...4 4. Engineering design overview...5 5. Installation...6

More information

Vission 20/20 micro-controller. Operation and service manual

Vission 20/20 micro-controller. Operation and service manual Vission 20/20 micro-controller Operation and service manual Section Title Table of Contents Section Number How To Use This Manual...TOC-8 Section 1 Operational Flow Charts Requirements to Start Compressor...1-1

More information

HGM-MZ Multi-Zone Monitor Annual Maintenance And Troubleshooting Guide

HGM-MZ Multi-Zone Monitor Annual Maintenance And Troubleshooting Guide HGM-MZ Multi-Zone Monitor Annual Maintenance And Troubleshooting Guide Service, Testing and Maintenance procedures BACHARACH Inc. HGM-MZ Routine Annual Maintenance And Operating Parameter Verification

More information

Vission 20/20 micro-controller. Operation and service manual Version 2.6

Vission 20/20 micro-controller. Operation and service manual Version 2.6 Vission 20/20 micro-controller Operation and service manual Version 2.6 Important Message READ CAREFULLY BEFORE OPERATING YOUR COMPRESSOR. The following instructions have been prepared to assist in operation

More information

IMI vibration switch USER MANUAL INSTALLATION - OPERATION - MAINTENANCE

IMI vibration switch USER MANUAL INSTALLATION - OPERATION - MAINTENANCE USER MANUAL IMI vibration switch INSTALLATION - OPERATION - MAINTENANCE Z0929039_A ISSUED 03/2017 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. contents Overview General

More information

Safe & Secure Series Electric Actuator with Internal Battery Back-up Installation, Operation & Maintenance Manual

Safe & Secure Series Electric Actuator with Internal Battery Back-up Installation, Operation & Maintenance Manual Safe & Secure Series Electric Actuator with Internal Battery Back-up Installation, Operation & Maintenance Manual For Use with: SNS4, SNS6, SNS10 & SNS15 Models Additional supplements may be needed for

More information

LTX RF LEVEL SENSOR. Instruction Manual

LTX RF LEVEL SENSOR. Instruction Manual LTX RF LEVEL SENSOR Instruction Manual FOR MODELS LTX01, LTX02, LTX05 Intempco Document No: LTX - M01 Rev. 1 Issue Date: April 2005 LTX01 RF LEVEL SENSOR USER MANUAL Software Rev : Rev. Date : June 2004

More information

Flight Systems. Replacement for KASSEC DESCRIPTION

Flight Systems.   Replacement for KASSEC DESCRIPTION DESCRIPTION The is a universal generator controller that will start, stop, and provide engine protection for most generators. Universal replacement for both the 90353 and 90354 KASSEC Compatible with most

More information

A1P OPERATING MANUAL

A1P OPERATING MANUAL A1P OPERATING MANUAL TABLE OF CONTENTS Introduction... p. 2 Features... p. 2 Description... p. 3 Theory of Operation... p. 3 Installation... p. 4 Electrical Connections... p. 5 Options... p. 6 Warranty...

More information

Type 2000 Transducer Product Instructions

Type 2000 Transducer Product Instructions Type 2000 Transducer Product Instructions The Type 2000 is an electro-pneumatic device that regulates an unregulated supply pressure down to an electronically-controlled output pressure. There are two

More information

UL/NEMA Type 1 & Type 12 FRENIC-HVAC Combination VFD

UL/NEMA Type 1 & Type 12 FRENIC-HVAC Combination VFD UL/NEMA Type 1 & Type 12 FRENIC-HVAC Combination VFD Safety Precautions Read this manual thoroughly before proceeding with installation, connections (wiring), or maintenance and inspection. Ensure you

More information

dv Sentry TM 208V 600V INSTALLATION GUIDE Quick Reference ❶ How to Install Pages 6 14 ❷ Startup/Troubleshooting Pages WARNING

dv Sentry TM 208V 600V INSTALLATION GUIDE Quick Reference ❶ How to Install Pages 6 14 ❷ Startup/Troubleshooting Pages WARNING dv Sentry TM 208V 600V INSTALLATION GUIDE FORM: DVS-IG-E REL. January 2018 REV. 003 2018 MTE Corporation High Voltage! Only a qualified electrician can carry out the electrical installation of this filter.

More information

Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002

Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002 Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002 Revision History Issue Date Reason 001 6/15/2016 Initial Issue 002 11/17/2016 Correct relay labels Copyright Siemens AG 2015 All

More information

Analog Condensate Monitor Instruction & Maintenance Manual

Analog Condensate Monitor Instruction & Maintenance Manual Analog Condensate Monitor Instruction & Maintenance Manual Advantage Controls, Inc. PO Box 1472 Muskogee, OK 64402 Phone: 800-743-7431 Fax: 888-686-6212 www.advantagecontrols.com EE-13 email: aci@advantagecontrols.com

More information

Fuel Cell Lab Manual. Non Conventional Energy Systems Facility. Non Conventional Energy Systems Facility. Boiler House

Fuel Cell Lab Manual. Non Conventional Energy Systems Facility. Non Conventional Energy Systems Facility. Boiler House Fuel Cell Lab Manual Non Conventional Energy Systems Facility Boiler House Department of Mechanical Engineering IIT Kanpur System Description The Nexa system provides up to 1200 watts of unregulated DC

More information

Lingenfelter NCC-002 Nitrous Control Center Quick Setup Guide

Lingenfelter NCC-002 Nitrous Control Center Quick Setup Guide Introduction: Lingenfelter NCC-002 Nitrous Control Center Quick Setup Guide The NCC-002 is capable of controlling two stages of progressive nitrous and fuel. If the NCC-002 is configured only for nitrous,

More information

High Pressure Leak Diagnostics System Part No & NA USER MANUAL LEAK DETECTION SERVICE

High Pressure Leak Diagnostics System Part No & NA USER MANUAL LEAK DETECTION SERVICE Cool Smoke HP High Pressure Leak Diagnostics System Part No. 500-0150 & 500-0150NA USER MANUAL LEAK DETECTION SERVICE 1324 Blundell Rd. Mississauga ON Tel. 905.615.8620 Fax. 905.615.9745 Introduction Congratulations!

More information

CENTRIFLOW METER and ELECTRONICS WIRING and INSTALLATION MANUAL REV 01/16 ORIGINAL LANGUAGE Copyright 2016 Eastern Instrument Laboratories, Inc. All Rights Reserved. TABLE OF CONTENTS SAFETY Safe Operation...

More information

CO 2 Room Sensor in the BAPI-Stat 3 Enclosure with Common Ground Configuration

CO 2 Room Sensor in the BAPI-Stat 3 Enclosure with Common Ground Configuration Identification and Overview The BAPI CO 2 Sensor is an accurate and reliable way of incorporating demand controlled ventilation into a building s HVAC strategy. It measures the CO 2 in a range of 0 to

More information

DSL-240 MkIII Single Pass Dust Monitor Measures mg/m 3 using DDP

DSL-240 MkIII Single Pass Dust Monitor Measures mg/m 3 using DDP Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions

More information

¼ to 1HP Chiller Service Manual

¼ to 1HP Chiller Service Manual ¼ to 1HP Chiller Service Manual Before performing any service to the unit, reset the controller to its default settings. It is our experience that many problems can be resolved by doing so. See Sec. 3.1

More information

Model FVP General Specification. Flanged Vortex Plate. Product Features

Model FVP General Specification. Flanged Vortex Plate. Product Features Flanged Vortex Plate Product Features Our Inline meters can measure Steam, Gases or Liquids. Measures two times lower than any other Vortex Meter. Heavy Duty & Maintenance Free Design C&C Machined from

More information

AVL Particle Measurement System Aviation

AVL Particle Measurement System Aviation AVL Particle Measurement System Aviation Measurement of non-volatile Particulate Matter mass and number emissions from aircraft turbines according to the AIR6241 MARKET REQUIREMENTS Non-volatile particulate

More information

Optidew Vision Optical Dew-Point Meter

Optidew Vision Optical Dew-Point Meter Optidew Vision Optical Dew-Point Meter Chilled Mirror Instruments A precision hygrometer, combining drift free measurement with the flexibility of a remote sensor. Ideal for laboratory and industrial measurements,

More information

CLA-VAL e-drive-34. User Manual. Motorised Pilots. CLA-VAL Europe LIN072UE - 04/16

CLA-VAL e-drive-34. User Manual. Motorised Pilots. CLA-VAL Europe LIN072UE - 04/16 User Manual CLA-VAL Europe www.cla-val.ch cla-val@cla-val.ch 1 - LIN072UE - 04/16 Table of Contents 1 Introduction... 3 1.1 Precautions Before Starting... 3 1.2 Troubleshooting... 3 1.3 General Disclaimer...

More information

Planting. Press to Highlight Configuration. Gauge Wheel Load Alarm. Value must be entered for your specific planter and parallel arm combination

Planting. Press to Highlight Configuration. Gauge Wheel Load Alarm. Value must be entered for your specific planter and parallel arm combination To create a configuration, make the following button presses to start the Configuration Wizard and then follow the instructions given on the display. Planting Start of Configuration Wizard Enter Settings

More information

QPV1 Electronic Pressure Regulator INSTALLATION AND MAINTENANCE INSTRUCTIONS

QPV1 Electronic Pressure Regulator INSTALLATION AND MAINTENANCE INSTRUCTIONS INSTALLATION AND MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QPV1 is a high resolution electronic pressure regulator. The QPV1 provides a regulated output pressure that is proportional to

More information

Low Profile Insertion Vortex

Low Profile Insertion Vortex Low Profile Insertion Vortex Product Features The same insertion meter can measure Steam, Gases or Liquids. Standardize on an insertion bar that can measure line sizes 2 24 Heavy Duty & Maintenance Free

More information

INTECH Micro 2300-RTD6

INTECH Micro 2300-RTD6 INTECH Micro 2300-RTD6 6 Channel RTD Input Station Overview. The Intech Micro 2300 Series is a system of modular I/O Remote Stations, that add an even lower cost option to Intech s already extensive intelligent

More information

Installation, Operation, and Maintenance Manual CEMLINE CORPORATION

Installation, Operation, and Maintenance Manual CEMLINE CORPORATION $5.00 Installation, Operation, and Maintenance Manual CEMLINE CORPORATION CEM-TROL Solid State Control Module and Electronic Control Valves P. O. Box 55, Cheswick, PA, 15024 Phone: (724) 274-5430 FAX (724)

More information

KANEAUTO 4-4 & KANE AUTO 5-4

KANEAUTO 4-4 & KANE AUTO 5-4 KANEAUTO 4-4 & KANE AUTO 5-4 Hand Held Gas Analyser Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel: +44 (0) 1707 375550 Fax: +44 (0) 1707 393277 E-mail: info@kane.co.uk

More information

Horizon Troubleshooting Flowchart

Horizon Troubleshooting Flowchart Horizon Troubleshooting Flowchart Table of Contents Warnings..................................................... 2 Screen........................................................3 Temp Sensor...................................................4

More information

OCX 8800 Oxygen/Combustibles Transmitter

OCX 8800 Oxygen/Combustibles Transmitter Instruction Manual OCX 8800 Oxygen/Combustibles Transmitter Hazardous Area OCX 8800 Oxygen/Combustibles Analyzer with Integral Electronics http://www.raihome.com Instruction Manual OCX 8800 Table of Contents

More information

DSL-220 MkIII Single Pass Opacity Monitor Measures 0-100% Opacity

DSL-220 MkIII Single Pass Opacity Monitor Measures 0-100% Opacity Ideal for monitoring opacity and smoke levels in the exhaust gas of industrial combustion or air filtration processes. In situ measurement directly in exhaust gas flow using the widely accepted light transmission

More information

Heating Cable. Description

Heating Cable. Description ITLS Base Panel Heat Tracing Control Panel for Ordinary Areas 10 VGA Touch Screen HMI 40 Amps/Loop @ 100 600 VAC 6 Loops to 72 Loops NEMA 4 or NEMA 4X Enclosure SCR Control Optional Wireless Temperature

More information

CO 2 Room Sensor in the BAPI-Stat Quantum Prime Style Enclosure

CO 2 Room Sensor in the BAPI-Stat Quantum Prime Style Enclosure Identification and Overview The BAPI CO 2 Sensor is an accurate and reliable way of incorporating demand controlled ventilation into a building s HVAC strategy. It measures the CO 2 in a range of 0 to

More information

Dusty / Dusty Ex. Low-Cost Broken Bag Detection. Operating Instructions. SWR engineering Messtechnik GmbH

Dusty / Dusty Ex. Low-Cost Broken Bag Detection. Operating Instructions. SWR engineering Messtechnik GmbH EN Dusty / Dusty Ex Operating Instructions Low-Cost Broken Bag Detection SWR engineering Messtechnik GmbH CONTENTS Page 1. Introduction.............................................................. 3 1.1

More information

IOM A-0984E IP68-420DC AUTOMATED PILOT ACTUATOR Installation and Operation Manual

IOM A-0984E IP68-420DC AUTOMATED PILOT ACTUATOR Installation and Operation Manual IOM A-0984E IP68-420DC AUTOMATED PILOT ACTUATOR Installation and Operation Manual Please ensure to read and understand the contents of this manual. TABLE OF CONTENTS 1 HEALTH AND SAFETY: READ FIRST...3

More information

MATRIX LLC LEVEL CONTROL PANEL

MATRIX LLC LEVEL CONTROL PANEL MATRIX LLC LEVEL CONTROL PANEL Installation Operation Maintenance Manual 1-26-09 REV-0.3 1 INDEX Introduction. 3 Product Configurations..... 4 I/O Descriptions. 8 Operation o Intercooler, Accumulator,

More information

Combined Ventilation Controller RVWS-T-224HA

Combined Ventilation Controller RVWS-T-224HA Combined Ventilation Controller RVWS-T-224HA 8-stage Control for Power/Natural Applications 2 variable speed stages, 2 curtain winch stages, 2 fixed speed ventilation stages, 1 thermo/mister cycle stage

More information

mass flowmeter for compressed air

mass flowmeter for compressed air mass flowmeter for compressed air VARIOMASS ECO pressure and temperature compensated easy and cost effective installation of the sensor no pressure loss caused by the sensor no moving parts, i.e. maintenance

More information

DCC-3000 Climate Control for Vintage Air GEN-IV systems

DCC-3000 Climate Control for Vintage Air GEN-IV systems INSTALLATION AND OPERATOR S MANUAL FOR DCC-3000 Climate Control for Vintage Air GEN-IV systems PARTS INCLUDED WITH THIS SYSTEM Vent sensor housings: 2 1 / 2 housings (x2) 2 housings (x2) Installation/operator

More information

Model: AEM14 Analog Engine Monitor

Model: AEM14 Analog Engine Monitor Model: AEM14 Analog Engine Monitor Installation and Setup Manual Version 1 Table of Contents Monitor Overview DMK Engine Monitor Kit Section 1: Initial Setup 1.1 Internal Settings Switches Figure 1. AEM14

More information

CENTRIFEEDER ELECTRONICS

CENTRIFEEDER ELECTRONICS CENTRIFEEDER ELECTRONICS FOR USE WITH CENTRIFEEDER with Integrated Vibratory Control REV 08/13 ADDENDUM VERSION 6.22 SOFTWARE Copyright 2011 Eastern Instrument Laboratories, Inc. All Rights Reserved. TABLE

More information

Inlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00

Inlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00 Inlet Controller TC5-ITA USER'S MANUAL M 890-00047 rev. 02 K 895-00458 rev. 00 TABLE OF CONTENTS PRECAUTIONS... 3 FEATURES... 4 LOCATION OF THE CONTROLS... 5 Status Leds...5 Internal Switches...6 INSTALLATION

More information

Solarity with InfoTech Software Solarity, powered by Genisys technology, is a powerful 4-channel scope with the functionality required to analyze the most sophisticated vehicles. InfoTech Software provides

More information

THC Heated FID 3-300A

THC Heated FID 3-300A Rack Mount/Table Top TVOC Analyzer THC Heated FID 3-300A Space saving 19"/3PU space saving rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of

More information

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus:

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus: 1.0 Introduction: This probe is designed to provide an additional help to automotive technicians in trouble shooting of electrical circuits problems in the car. Apart from using the normal multi tester,

More information

Specification: Industrial Mixing Center

Specification: Industrial Mixing Center Specification: Industrial Mixing Center Model: IMC40 Article I. General Specifications: Section 1.01 The assembly shall be pre-piped with performance matched components and pressure tested before delivery.

More information

POWERLINE 2000 Energy Management System TM

POWERLINE 2000 Energy Management System TM Display Panel TM The PowerLine 00 EMS is a specialized power distribution and energy management system intended to be used in recreational vehicles. The Control Module is housed in the standard main distribution

More information

Device Description User Manual Logix MD+ Positioners with HART

Device Description User Manual Logix MD+ Positioners with HART Device Description User Manual Logix MD+ Positioners with HART CONTENTS DD MENU CHART... 3 GENERAL INFORMATION... 6 INTRODUCTION... 6 QUALIFIED PERSONNEL... 6 USING THIS DOCUMENT... 6 TERMS CONCERNING

More information

Model 7400 OPERATOR MANUAL

Model 7400 OPERATOR MANUAL Model 7400 OPERATOR MANUAL DORAN SCALES, INC. 1315 PARAMOUNT PKWY. BATAVIA, IL 60510 1-800-262-6844 FAX: (630) 879-0073 http://www.doranscales.com MANUAL REVISION: 1.0 MAN0198 10/3/2005 INTRODUCTION Introducing

More information

USERS MANUAL MCD REMOTE OPERATOR

USERS MANUAL MCD REMOTE OPERATOR USERS MANUAL MCD REMOTE OPERATOR Order Code: 175G9004, 175G3061 Contents Contents Introduction...2 Important User Information...2 General Description...2 Symbols Used in this Manual...2 Installation...3

More information

NT HIGH-TEMPERATURE ELECTRONIC FLOWMETER, MODEL 4401

NT HIGH-TEMPERATURE ELECTRONIC FLOWMETER, MODEL 4401 P/N 1400 (Rev. A 08/12) NT HIGH-TEMPERATURE ELECTRONIC FLOWMETER, MODEL 4401 User Guide Table of Contents Introduction... 2 Dimensions... 3 Installation... 5 Provided Equipment... 5 Operating Environment...

More information

User s Manual. ACH550-CC/CD Packaged Drive with Classic Bypass Supplement for ACH550-UH HVAC User s Manual

User s Manual. ACH550-CC/CD Packaged Drive with Classic Bypass Supplement for ACH550-UH HVAC User s Manual User s Manual ACH550-CC/CD Packaged Drive with Classic Bypass Supplement for ACH550-UH HVAC User s Manual ii ACH550-CC/CD Packaged Drive with Classic Bypass ACH550 Drive Manuals GENERAL MANUALS ACH550-UH

More information

3 Phase Smart Controller

3 Phase Smart Controller 3 Phase Smart Controller Installation and Owner s Manual STP-SCIII 208-230 VAC, 60Hz, 120 Volt Coil Franklin Fueling 3760 Marsh Rd. Madison WI 53718 USA Tel: +1 608 838 8786 800 225 9787 Fax: +1 608 838

More information

PowerOhm Installation Manual for LG ATV Series Braking Modules

PowerOhm Installation Manual for LG ATV Series Braking Modules PowerOhm Installation Manual for LG ATV Series Braking Modules IMPORTANT: These instructions should be read thoroughly before installation. All warnings and precautions should be observed for both personal

More information

Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002

Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002 Maxum II Modular Oven Automatic Purge Control Unit Product Update PU-002 Revision History Issue Date Reason 001 6/15/2016 Initial Issue Copyright Siemens AG 2015 All rights reserved atable of Contents

More information

PowerOhm Installation Manual for BM R Series Braking Modules

PowerOhm Installation Manual for BM R Series Braking Modules PowerOhm Installation Manual for BM R Series Braking Modules IMPORTANT: These instructions should be read thoroughly before installation. All warnings and precautions should be observed for both personal

More information

EC 200 EC 300 GE 360 GE 400 GE 600 L-S BM

EC 200 EC 300 GE 360 GE 400 GE 600 L-S BM DIESEL SPACE HEATER - SERVICE MANUAL INDEX 1. CONTROLS AND COMPONENTS 2. FLAME CONTROL CYCLES 3. MAINTENANCE SCHEDULE 4. REPAIR PROCEDURES 1. FAN MOTOR ASSEMBLY 2. FUEL FILTER ASSEMBLY 3. FUEL PUMP ASSEMBLY

More information

Hot Tap Insertion Vortex

Hot Tap Insertion Vortex Hot Tap Insertion Vortex Product Features The same insertion meter can measure Steam, Gases or Liquids. Standardize on an insertion bar that can measure line sizes 2 48 Heavy Duty & Maintenance Free Design

More information

Model 8000XL OPERATOR MANUAL

Model 8000XL OPERATOR MANUAL Model 8000XL OPERATOR MANUAL DORAN SCALES, INC. 1315 PARAMOUNT PKWY. BATAVIA, IL 60510 1-800-262-6844 FAX: (630) 879-0073 http://www.doranscales.com MANUAL REVISION: 1.0 MAN0191 10/3/2005 INTRODUCTION

More information

DynoTune Wideband Gauge

DynoTune Wideband Gauge DISPLAY: RED GREEN BLUE DynoTune Wideband Gauge FACE: BLACK WHITE BEZEL: BLACK SILVER PACKAGE: ROUND SQUARE The DynoTune A/F Gauge will display the air/fuel ratio output from the LC-1 Wide-Band controller.

More information

Automotive Application ET01 Software Revision A 12/06

Automotive Application ET01 Software Revision A 12/06 Automotive Application ET01 Software Revision A 12/06 INTRODUCTION... 2 FUNCTIONAL DESCRIPTION... 3 INSTALLATION... 4 COMPONENT PLACEMENT... 4 PLUMBING AND WIRING... 5 MSBC OPERATION (ET-01)... 14 TIMED

More information

Model 200L. Instruction Manual Dual Dilution Probe Controller Part Number Dec2007

Model 200L. Instruction Manual Dual Dilution Probe Controller Part Number Dec2007 Model 200L Instruction Manual Dual Dilution Probe Controller Part Number 14638 20Dec2007 2007 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change.

More information

SPN 4364 (ACM)(EPA10); SPN 4364 (ACM)(GHG14) January Publication Number / Title Platform Section Title Change SPN 4364/FMI 18 - EPA10

SPN 4364 (ACM)(EPA10); SPN 4364 (ACM)(GHG14) January Publication Number / Title Platform Section Title Change SPN 4364/FMI 18 - EPA10 01 01-18 1 01 01-18 SUBJECT DATE SPN 4364 (ACM)(EPA10); SPN 4364 (ACM)(GHG14) January 2018 Additions, Revisions, or Updates Publication Number / Title Platform Section Title Change DDC-SVC-MAN-0084 DD

More information

Physical Property Analyzer PPA4100 Series. On-Line RVP Analyzer

Physical Property Analyzer PPA4100 Series. On-Line RVP Analyzer Physical Property Analyzer PPA4100 Series On-Line RVP Analyzer More accurate, on-line measurement of Reid Vapor Pressure will save millions of dollars in gasoline blending... the PPA4100 is more than an

More information

Rosemount TM OCX 8800 Oxygen and Combustibles Transmitter

Rosemount TM OCX 8800 Oxygen and Combustibles Transmitter Instruction Manual April 2017 Rosemount TM OCX 8800 Oxygen and Combustibles Transmitter General Purpose Probe with Integral Electronics and Local Operator Interface Page/Section Throughout Summary HIGHLIGHTS

More information

Morgan Schaffer. Calisto Monitors. Maintenance. Presented by: Eric Thibodeau. Technical Support Specialist

Morgan Schaffer. Calisto Monitors. Maintenance. Presented by: Eric Thibodeau. Technical Support Specialist Morgan Schaffer Calisto Monitors Maintenance Presented by: Eric Thibodeau Technical Support Specialist Flushing of the oil lines Purging the oil lines Step 1: All valves are closed Calisto Pump OFF OUTLET

More information

Model Q45H/ Residual Chlorine Monitor...

Model Q45H/ Residual Chlorine Monitor... Model Q45H/62-63 Residual Chlorine Monitor... ATI s Model Q45H is designed for on-line monitoring and control of chlorination systems. It is available in two versions: a free chlorine monitor for drinking

More information

Vortex Plate. Model VP General Specification. Product Features

Vortex Plate. Model VP General Specification. Product Features Vortex Plate Product Features Our Inline meters can measure Steam, Gases or Liquids. Heavy Duty & Maintenance Free Design C&C Machined from one piece of solid stainless steel Absolutely NO o-rings or seals

More information

OnCommand Troubleshooting Guide Hayward Industries

OnCommand Troubleshooting Guide Hayward Industries OnCommand Troubleshooting Guide 2010 Hayward Industries Table of Contents Safety Precautions Page 1 Overview Pages 2-5 Software Troubleshooting Page 6 Local Display Pages 7-8 Relays Pages 9-10 Heaters

More information

Electric Actuated Ball Valves 150 lb Flanged Stainless Steel, Full Port 1/2 to 4 inch Pipe Sizes

Electric Actuated Ball Valves 150 lb Flanged Stainless Steel, Full Port 1/2 to 4 inch Pipe Sizes Electric Actuated Ball Valves 150 lb Flanged Stainless Steel, Full Port 1/2 to 4 inch Pipe Sizes SERIES Features Direct mount Full Port 16 Stainless Steel ball valve 2-piece ANSI Class 150 flanged body

More information

Subject Underhood G System Error Codes and Symptoms System or Parts affected

Subject Underhood G System Error Codes and Symptoms System or Parts affected System or Parts affected Index Underhood70G (V90Gxxx) System or Parts affected... 1 Overview... 1 Identifying your System... 1 Retrieving Logged Error Messages... 1 Error Messages... 3 Error Message Table...

More information

1100 Series Piston Type Differential Pressure Gauges

1100 Series Piston Type Differential Pressure Gauges 1100 Series Piston Type Differential Pressure Gauges 1. Safety Before installing, check the Series Number and verify compatibility to the process media and temperature in contact with the wetted parts.

More information

MODEL 905V OPERATING INSTRUCTIONS

MODEL 905V OPERATING INSTRUCTIONS MODEL 905V OPERATING INSTRUCTIONS Quantek Instruments 183 Magill Drive Grafton, MA 01519 Tel: (508) 839-3940 Fax: (508) 819-3444 Email: sales@quantekinstruments.com GENERAL DESCRIPTION These instructions

More information

CAMS. Combustion Airflow Management System. Proven solutions for a tough industry. Precision Airflow Measurement An ONICON Brand

CAMS. Combustion Airflow Management System. Proven solutions for a tough industry. Precision Airflow Measurement An ONICON Brand CAMS Combustion Airflow Management System Proven solutions for a tough industry Precision Airflow Measurement An ONICON Brand CAMS TM Combustion Airflow Management System Product Description The Air Monitor

More information

Automatic Transfer Switch FT-10 Network Control Communications Module (CCM-T) Kit

Automatic Transfer Switch FT-10 Network Control Communications Module (CCM-T) Kit Instruction Sheet 10-2004 Automatic Transfer Switch FT-10 Network Control Communications Module (CCM-T) Kit 541 0811 PURPOSE OF KIT A CCM-T is used to monitor and control an automatic transfer switch.

More information

BAYLAN AUTOMATED TEST BENCH

BAYLAN AUTOMATED TEST BENCH BAYLAN AUTOMATED TEST BENCH At this test bench, the electromagnetic flow meters are used with the measurement method. Calibrations of flowmeters are made with the help of the water collection tank and

More information

Ventilation Control. Ventilation Control 12

Ventilation Control. Ventilation Control 12 Ventilation Control One of our major markets is building automation where our sensors control individual fans, dampers, valves, etc. A common application is controlling ventilation in rooms with varying

More information

SS3 Linear / Rotary. Smart Valve Positioner. Smart performance with innovative and ever-strong coil drive even under harsh working environments

SS3 Linear / Rotary. Smart Valve Positioner. Smart performance with innovative and ever-strong coil drive even under harsh working environments Smart Valve Positioner SS3 Linear / Rotary Smart performance with innovative and ever-strong coil drive even under harsh working environments Easy and quick auto-calibration Detecting RA (reverse acting)

More information

RF Point Level Control with Sensor Monitor

RF Point Level Control with Sensor Monitor The 681 Point Level Control utilizes DPDT relays to provide switching for peripheral devices (such as pumps) in level applications. A sensor attached to the control acts as an antenna to transmit the process

More information

TECHNICAL PAPER 1002 FT. WORTH, TEXAS REPORT X ORDER

TECHNICAL PAPER 1002 FT. WORTH, TEXAS REPORT X ORDER I. REFERENCE: 1 30 [1] Snow Engineering Co. Drawing 80504 Sheet 21, Hydraulic Schematic [2] Snow Engineering Co. Drawing 60445, Sheet 21 Control Logic Flow Chart [3] Snow Engineering Co. Drawing 80577,

More information

Figure 1: WPR2 Dimensions and Hardware

Figure 1: WPR2 Dimensions and Hardware Installation and Operation Instructions WPR2 Series (Max. Line Pressure < 300 PSI) Wet to Wet Differential Pressure Ø0.200" 2.60" 4.47" 5.47" 5.28" 3.00" Precautions Figure 1: WPR2 Dimensions and Hardware

More information

SB 2000 PUSH TO SEARCH NEXT STAG E. Aerotech, Inc. FORM: QM 1320

SB 2000 PUSH TO SEARCH NEXT STAG E. Aerotech, Inc. FORM: QM 1320 Inlet Controller SB 2000 USER'S MANUAL AUTO OPEN MANUAL PUSH TO SEARCH NEXT STAG E CLOSE Aerotech, Inc. FORM: QM 1320 4215 Legion Dr. Mason, MI 48854-1036 USA Rev. 3, Sept. 1997 Ph. (517) 676-7070 Fax

More information

VOC Duct and Rough Service Sensor

VOC Duct and Rough Service Sensor Overview and Identification Humans respirate Volatile Organic Compounds (VOCs) as well as CO 2. The BAPI sensor measures these VOCs and indicates space occupancy just as well as a CO 2 sensor. The advantage

More information

Sidel Therm & CO 2 Counter Instruction & Installation Manual

Sidel Therm & CO 2 Counter Instruction & Installation Manual Sidel Therm & CO 2 Counter Instruction & Installation Manual P.O. Box 1868 Atascadero, CA 93423 800-668-5003 Fax: 805-464-0504 Sid Abma, President www.sidelsystems.com Table Of Contents System Overview..............................

More information

PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS. v1.1 Revised Nov 29, 2016

PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS. v1.1 Revised Nov 29, 2016 PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS v1.1 Revised Table of Contents 1 Common Alarms and Warnings... 1 2 Common Issues... 6 2.1 Communication problems... 6 2.1.1 Controller communication

More information

DIESEL Engine Fire Pump Controllers Features

DIESEL Engine Fire Pump Controllers Features 1-1 Printer / Recorder The industrial grade thermal printer is housed in a rugged steel enclosure within the controller. The on/off switch, feed and reset buttons are front accessible. A bi-color status

More information

CEMLINE CORPORATION P. O. BOX 55 CHESWICK, PENNSYLVANIA Phone: (724) FAX (724)

CEMLINE CORPORATION P. O. BOX 55 CHESWICK, PENNSYLVANIA Phone: (724) FAX (724) Installation, Operation, and Maintenance Manual CEMLINE CORPORATION Electronic Controls for Water Heaters with RWD45U Controller CEMLINE CORPORATION P. O. BOX 55 CHESWICK, PENNSYLVANIA 15024 Phone: (724)

More information

BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD)

BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD) LAKEWOOD INSTRUMENTS MODEL 260 SERIES BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD) INSTALLATION & OPERATION MANUAL SERIAL #: Lakewood Instruments 7838 North Faulkner Road, Milwaukee, WI

More information