University of York NO x instrument installation on the NERC Dornier 228 aircraft.

Size: px
Start display at page:

Download "University of York NO x instrument installation on the NERC Dornier 228 aircraft."

Transcription

1 University of York NO x instrument installation on the NERC Dornier 228 aircraft. Report prepared for: NERC Airborne Research and Survey Facility Firfax Building, Meteor Business Park Cheltenham Road East Gloucester, GL2 9QL UK Report prepared by: Air Quality design, Inc W I-70 Frontage Road North, #105 Wheat Ridge, Colorado USA May 2013

2 Contents DESCRIPTION OF THE EQUIPMENT INSTALLED ELECTRICAL LOAD ANALYSIS HAZARD ASSESSMENT OF THE INSTALLED EQUIPMENT STRESS REPORT FOR THE INSTALLED EQUIPMENT FLAMMABILITY REPORT/ANALYSIS ENVIRONMENTAL TESTING SUMMARY INSTRUCTIONS FOR CONTINUED CHECKS, MAINTENANCE AND PERIODIC SERVICE ii

3 Description of the equipment installed. Instrument functional description The University of York NO x instrument, manufactured by Air Quality Design, Inc (AQD - Wheat Ridge, CO, USA), is a two-channel chemilumescence-based nitric oxide (NO) detector. NO is measured by the chemiluminescence resulting from the reaction of trace amounts of NO in the sample air with percent levels of ozone to form NO 2 (nitrogen dioxide) in an excited state. The NO 2 relaxes by emitting a red photon which is counted using a dry-ice cooled photomultiplier tube. The instrument includes an inlet-based photolytic converter that converts NO 2 to NO, allowing simultaneous measurement of NO and NO 2 (taken together as NO x ). The component parts of the instrument are described below and are shown graphically in Figure 1. NO detector The NO detector contains the dry-ice cooled photomultiplier tube housing, reaction chambers, detector electronics and vacuum valves. The instrument chassis is designed for 19 inch rack-mountability and is manufactured from inch 5052 aluminum with flanges attached to a inch 6061-T6 aluminum front panel and 0.09 inch rear panel. The NO detector is a 5U (8.75 inch) high box, 24 inches deep, and weighs 59 pounds. Ozonizer The Ozonizer contains two corona-discharge ozone generators and associated flow and pressure control elements. The ozonizer uses pure oxygen as a reagent gas and produces about 3% ozone in a flow of 100 sccm (standard cubic centimeters per minute) per channel. The instrument chassis is designed for 19 inch rackmountability and is manufactured from inch 5052 aluminum with flanges attached to a inch 6061-T6 aluminum front panel and 0.09 inch rear panel. The ozonizer is a 2U (3.5 inch) high box, 18 inches deep, and weighs 16 pounds. Inlet system The Inlet system contains a UV-LED based photolytic converter used to convert ambient NO 2 to NO for subsequent measurements, the system flow and pressure control elements, a data acquisition device, and system calibration components. The instrument chassis is designed for 19 inch rack-mountability and is manufactured from inch 5052 aluminum with flanges attached to a inch 6061-T6 aluminum front panel and 0.09 inch rear panel. The Inlet system is a 3U (5.25 inch) high box, 18 inches deep, and weighs 30 pounds. Computer tray The computer tray contains a laptop computer used to control the instrument and a serial communications device. The tray is designed for 19 inch rack-mountability and is manufactured from inch cold-rolled steel with welded corners. The computer tray is a 1U (1.75 inch) high tray, 17 inches deep, and weighs 12.5 pounds. AC power distribution box The NERC-built AC power distribution box contains inverters to converter the aircraft 28 VDC power to 230 VAC power for instrument use. The instrument chassis is designed for 19 inch rack-mountability and is manufactured from inch 5052 aluminum with flanges attached to a inch 6061-T6 aluminum front 1

4 panel and 0.09 inch rear panel. The AC power distribution box is a 3U (3.5 inch) high box, 20 inches deep, and weighs 32 pounds. Pump Pallet The Pump Pallet is a 24 x 30 x ¼ inch 6061-T6 aluminum plate to which is mounted the instrument vacuum pumps and gas cylinders. The pump pallet is designed to be mounted directly to the aircraft seat-rails and weighs 94 pounds. 2

5 Figure 1. Sketch of the University of York NO x instrument installation configuration on the NERC Dornier 228. The various installed components are identified in the table below. ID Description 1 Computer tray: sliding tray containing the system laptop computer and serial communication device 2 Inlet System: Contains a photolytic NO 2 converter, calibration components, and flow and pressure control devices. 3 NO detector: contains a two channel NO detector with dry ice cooled detectors 4 Ozonizer: two channel corona discharge ozone generator (reagent gas for the NO detector). 5 NERC-built AC power distribution box 6 NERC-provided instrument rack 7 Oxygen cylinder forward stop. 8 Calibration cylinder forward stop. 9 Edwards 20i scroll vacuum pump. 10 N/A 11 NO in N2 calibration gas 12 Oxygen cylinder 13 Rear gas cylinder stop 14 Thomas diaphragm pump. 15 Instrument slide (2 per each rack-mounted box except for the AC power distribution box). 16 Pump mounting plate. 3

6 Electrical load analysis. Electrical loads The electrical load requirements for NO x instrument are summarized in Table 1. Values are based on the manufacturer specified power consumption values for the component electrical parts in each instrument chassis. The figures assume 80% efficient AC-DC power supplies (consistent with manufacturer specifications for a convection cooled system (worst case) Table 1. NO x instrument AC power consumption inventory. Component 4 Power consumption, Watts Power requirement, Amps at 240 VAC NO detector with dry ice cooler NO detector with optional/removable thermal electric cooler (TEC) Ozonizer Inlet system Edwards vacuum pump Thomas vacuum pump Totals* *- totals assume use of the optional TEC PMT cooler. The normal mode of operations is with the dry ice cooler. Wire, cable, and connectors All of the wire and connectors used in the instrument are of the type and size specified in the NERC document Engineers Guidelines for the Design and Preparation of Airborne Scientific Instruments. Table 2 lists the wire and cable types used in the instrument. Table 2. Wire and cable used in the NO x instrument Wire/cable application Wire cable description MS number External AC power distribution Teflon-insulated 3 conductor, 16 M27500/16SP3S23 gauge, shielded and Teflonjacketed Internal AC power distribution Teflon-insulated, 16 gauge M22759/32-16 Internal DC power distribution and Teflon-insulated, 22 gauge M22759/32-22 signals External signal and data Teflon-insulated 4 conductor, 24 M22759/24SRS23 communication gauge, shielded and Teflonjacketed Detector signals RG178 coaxial cable, Teflon insulation, braided shield, Teflon jacket n/a Several different connector types are used for AC power and signal distribution these are listed in Table 3. Table 3. Connectors used for external cabling in the NO x instrument. Connector purpose Connector description Connector part number

7 AC power from the AC distribution box AC power connection to the individual instrument chassis Circular plastic connector manufactured by Bulgin, 3 conductor, 12A contact rating, IP68 Threaded bayonet metal shell connector manufactured by AMP, 13 A contact rating Inter-chassis signal transmission MS crimp contact DB connectors (9 and 25 pin). Manufactured by Cannon, 3 A contact rating. BNC 50 ohm solder-type BNC manufactured by Pasternak PX0731/P PX0730/S MS3106A16-7P MS3106A16-7S M24308 series Meets MS3106A16-7S Circuit breakers and line filters Each of the instruments chassis is circuit protected using a thermal circuit breaker rated at 5 A. Inside each instrument chassis the AC power is filtered to minimize interference from noise either into or out of the particular instrument chassis. A block diagram of the external cabling interconnections is shown in Figure 2. Finally, all external cable and the pneumatic PFA Teflon tubing are sleeved in flame-resistant/self extinguishing sleeving. 5

8 Figure 2. Block diagram of external cabling connections for the NO x instrument. 6

9 Hazard assessment of the installed equipment. The NO x instrument presents two hazards that the operator must be aware of. First the instrument uses pure oxygen as a reagent gas for the ozonizer. While the cylinder regulator has pressure relief valves and the oxygen within the instrument is electronically pressure controlled and mechanically flow controlled, care should be taken to ensure that the oxygen connection to the instrument from the cylinder remain free of oil and other debris. Secondly, the instrument produces a low flow of 3% ozone, which, if it escaped the confines of the instrument would be an irritant gas. The instrument installation includes two measures to mitigate this hazard. First, there is a ozone destruction trap on the exhaust side of the pump. Second, the effluent from the ozone destruction trap is vented outside the cabin. If the ozone destruction trap failed, and if the ozone was vented into the cabin the pungent odor of ozone would be readily recognized by the scientific observer and the instrument could be shut down. Finally, every effort has been made to isolate the instrument operator from AC line voltages, but normal caution should be observed. Stress report for the installed equipment. The NO x instrument is installed in a NERC-provided, type-certified instrument rack and on a separate pump pallet, so the two cases will be considered separately. Instrument Rack: For the instrument rack the AQD instrument chassis are secured to the rack in two ways. First, the instruments are bolted through the ⅛ 6061-T6 front panel in four places to the steel DIN rails of the rack using 6mm screws. The front panel attachment acts to resist both forward and rearward movement of the instruments. Second, the instruments are secured to cold-rolled steel telescoping slides to both the front and rear Din rails on each side. The telescoping slides act to resist both sideways and up/down movement of the instruments. The fasteners used to attach the instruments to these points are all at least 55 ksi tensile strength. Considering the heaviest of the instruments as an example, the NO detector at 59 pounds, the loads and safety factors for attachment are shown in Table 4. Table 4. Strength of fasteners attaching the NO detector to the instrument rack. Attachment point Screw size screw area (in^2) Number of screws Holding Strength Box weight (pounds) Safety factor at 9 G Front panel 6mm Slide to instrument Size Slide to DIN rails Size The other instruments are identically attached, but weigh much less, so the safety factors noted for the NO detector are much greater for them. The computer attachment is considered further in the section Additional Considerations. 7

10 Given that the instruments are securely attached to the rack we can consider the overall load of the rack in light of the aforementioned NERC guideline document. The instrument was designed in the software program SolidWorks, so we used that software, in conjunction with measured and assigned weights, to evaluate the load on the instrument rack. Figure 3 shows the rack model, mass, and center of gravity for the loaded rack. The result of this calculation is that the center of mass for the populated rack is at 19.9 inches with a total mass, including the rack itself, of 208 pounds. The intersection of this mass and center of Figure 3. Mass and center of gravity analysis of the loaded NO x rack. 8

11 gravity fit within the Acceptable Rack Load envelope specified in the NERC engineering guidance document. Pump Pallet: The components that are mounted to the pump pallet, including two vacuum pumps and two gas cylinders, are each secured to the ¼ thick plate with four grade-8 ¼ bolts. The material of the bolts has a tensile strength of 120 Ksi. Following the example above, we considered the heaviest of the components, the Edwards vacuum pump at 59 pounds to estimate the stress. Since these components are bolted directly to the plate we must first consider the center of gravity. Figure 4. Mass and center of gravity analysis of the loaded pump pallet. Again, similar to the Solidworks calculation for the rack, we used measured and assigned weights to arrive at the result seen in Figure 4. We calculate a total mass of 115 pounds with a center of gravity of 4.3 inches. We estimate the mechanical advantage of the center of gravity to be about 10, given the center of gravity and the approximate bottom side of the grade 8 bolts where they attach to the various components (nominally 0.5 inch). In either a forward or sideways load we conservatively assign the stress to only two of the four bolts. Given these conditions the stress applied by the Edwards vacuum pump at 9 g would be 5300 pounds, versus the 7600 pound tensile holding strength of two bolts, resulting in a safety factor of 1.4. Additional Considerations. The two additional considerations are the attachment of the computer to the computer tray and the attachment of the gas cylinders to the gas bottle mounts. In both cases we referred to the online calculator website to identify and use appropriate estimates of the strength of materials using the attachment methods employed. For the computer attachment with used a calculation based on plate theory with one unclamped edge to approximate the 0.02 thick type 304 stainless steel z-clamps used to secure the computer to the computer tray. There are three similar clamps employed, two on the front edge of the 5 pound computer and one on the rear edge of the computer. We assumed a 9 g forward load on the computer which translates physically to an upward load on the three clamps. Following the data and figures provide at : 9

12 Diagram of the calculated stress, Maximum stress along the free edge of the clamp, We find that the maximum stress is 154 MPa, well below the 203 MPa yield strength of the stainless material and a safety factor of 1.3. For the gas cylinders, which are secured with two ¾ wide, 0.03 thick type 304 stainless steel band clamps we assumed a fully clamped plate with the stress applied to the upper ½ of the clamp. At 9 g, and a cylinder weight of 40 pounds we calculate a stress on one of the band clamps of 32 psi (360 pounds distributed over 11 in 2 ). Following the calculator for a fully clamped plate we find a maximum stress on 0.02 MPa, a very small number compared to the 205 MPa yield and 515 Mpa tensile strength of the material. Flammability report/analysis. Oxygen is classified as flammable because it is a combustion accelerant. However, the oxygen used in this system is contained with the confines of the instrument. Environmental testing summary. Since this is a bespoke instrument we have not submitted it en suite for environmental testing. Mechanically speaking we have followed established guidelines for materials, fasteners, and fastener attachment methods. Electrically, the system uses commercial, CE-marked power supplies and other electronic components so we assume that radiated emissions are within established guidelines. Instructions for continued checks, maintenance and periodic service. While the NO x instrument may require period cleaning and internal component maintenance for optimal performance, the routine checks from and operational point of view are limited to regular examination of the system attach points, detailed above. We expect that the shortest term replaceable components are the ozone destruction traps (estimated lifetime is 1000 hours of operation), and the Edwards vacuum pump (5000 hours of operation). Calibration of the instrument on an every-flight basis will alert the operator to the need for my detailed cleaning and maintenance. 10

13 System block diagram 11

Technical Offer. Antek Model 6200 Analyzer Determine Total Sulfur and Total Nitrogen in a single analyzer

Technical Offer. Antek Model 6200 Analyzer Determine Total Sulfur and Total Nitrogen in a single analyzer Technical Offer Antek Model 6200 Analyzer 2015.4 Determine Total Sulfur and Total Nitrogen in a single analyzer Meet sulfur regulations and control your process with continuous sulfur and/or nitrogen measurements

More information

AMI. American Magnetics, Inc. Excellence in Magnetics and Cryogenics

AMI. American Magnetics, Inc. Excellence in Magnetics and Cryogenics Cryogenic Liquid Level Instruments Versatile Reliable Affordable AMI American Magnetics, Inc. Excellence in Magnetics and Cryogenics A Versatile Level Measurement System... A Variety of Liquids Nitrogen

More information

MIATECH SOLUTIONS, A PIECE OF THE PUZZLE. Bio Turbo 300. Bio Turbo 1000

MIATECH SOLUTIONS, A PIECE OF THE PUZZLE. Bio Turbo 300. Bio Turbo 1000 MIATECH SOLUTIONS, A PIECE OF THE PUZZLE Four Stage Filtration Step 1: Air Filter. Bio Turbo 100 Bio Turbo 300 Bio Turbo 1000 Step 2: Cell Destruct Filter. Step 3: Generation destroys Ethylene and Airborne

More information

REPORT NUMBER: MID-001 ORIGINAL ISSUE DATE: Sept 27, 2013 REVISED DATE: NA

REPORT NUMBER: MID-001 ORIGINAL ISSUE DATE: Sept 27, 2013 REVISED DATE: NA REPORT NUMBER: 101351744MID-001 ORIGINAL ISSUE DATE: Sept 27, 2013 REVISED DATE: NA TEST REPORT EVALUATION CENTER Intertek Verification Center 8431 Murphy Dr Middleton, WI 53562 x RENDERED TO Marc Doheny

More information

INDIAN INSTITUTE OF TECHNOLOGY KANPUR Kanpur , Uttar Pradesh, India Centre for Lasers and Photonics

INDIAN INSTITUTE OF TECHNOLOGY KANPUR Kanpur , Uttar Pradesh, India Centre for Lasers and Photonics INDIAN INSTITUTE OF TECHNOLOGY KANPUR Kanpur 208016, Uttar Pradesh, India Centre for Lasers and Photonics Enquiry no.: CELP/RV/EQP/MHR/2017/1 Enquiry date: 23/04/2018 Closing date: 17/05/2018 Sealed quotations

More information

DETAIL SPECIFICATION SHEET

DETAIL SPECIFICATION SHEET INCH-POUND 30 June 2014 SUPERSEDING MIL-DTL-16377/53C 27 September 2005 DETAIL SPECIFICATION SHEET FIXTURES, LIGHTING; INCANDESCENT AND LIGHT EMITTING DIODE (LED), DETAIL LIGHTING, LANTERN, HAND, PORTABLE

More information

Biogas safety control equipment

Biogas safety control equipment Biogas safety control equipment As environmental concerns and energy efficiency are increasing, efficient economical and safe gas control systems become an increasingly important part of operating the

More information

Emergency lighting units EM powerled

Emergency lighting units EM powerled BASIC FX SR 50 W Combined emergency lighting LED Driver Product description Independent fixed output LED Driver for mains operation with integrated Simple CORRIDOR FUNCTION (CF) Emergency lighting function

More information

VSO - MI thermally compensated proportional valve

VSO - MI thermally compensated proportional valve The VSO - MI is a miniature proportional valve specifically designed for medical device manufacturers and incorporates thermal compensation to provide precise flow control and stability over a wide operating

More information

The pumps shall be Oil Free and Variable speed controlled by Pressure Transducer, PLC and inverter.

The pumps shall be Oil Free and Variable speed controlled by Pressure Transducer, PLC and inverter. 1 General Model MS100-3-60 expandable/modular duplex central vacuum system consisting of two Becker or approved equivalent KVX 3.140/1-400 vacuum pumps mounted on a common skid with expandable automatic

More information

Lifeline Medical Systems

Lifeline Medical Systems Installation, Operation and Service Manual Lifeline Medical Systems Fully-Automatic Manifold For Use With Liquid Cylinders LQ x HP Liquid x High-Pressure Models Part No. 6-847703-00 Rev. C00 Pg. 1 Introduction

More information

Each pump shall be equipped with carbon/graphite vanes. A vacuum relief valve shall be incorporated into the design of the pump.

Each pump shall be equipped with carbon/graphite vanes. A vacuum relief valve shall be incorporated into the design of the pump. 1 General Model MS300-4-530 expandable/modular central vacuum system consisting of 4 Becker or approved equivalent VTLF2.250 vacuum pumps mounted on a common skid with expandable automatic alternating

More information

Installation Guide Smart-UPS On-Line SRT1000/SRT1500 XLA Tower/Rack-Mount

Installation Guide Smart-UPS On-Line SRT1000/SRT1500 XLA Tower/Rack-Mount Installation Guide Smart-UPS On-Line SRT1000/SRT1500 XLA Tower/Rack-Mount Important Safety Messages Read the instructions carefully to become familiar with the equipment before attempting to install, operate,

More information

USER MANUAL RHF SERIES

USER MANUAL RHF SERIES Hukseflux Thermal Sensors USER MANUAL RHF SERIES Ring heat flux sensors Copyright by Hukseflux manual v1824 www.hukseflux.com info@hukseflux.com Warning statements Follow the installation instructions

More information

VSO Thermally Compensated Proportional Valve

VSO Thermally Compensated Proportional Valve 4 VSO Thermally Compensated Proportional Valve Typical Applications Gas Chromatography Mass Spectrometry Ventilators O 2 Concentrators/Conservers Anaesthesia Delivery & Monitors Pressure & Control Mass

More information

Miniature Proportional Valve Thermally Compensated Proportional Valve. Physical Properties

Miniature Proportional Valve Thermally Compensated Proportional Valve. Physical Properties VSO Miniature Proportional Valve Thermally Compensated Proportional Valve Typical Applications Gas Chromatography Mass Spectrometry Ventilators O 2 Concentrators/Conservers Anaesthesia Delivery & Monitors

More information

Miniature Proportional Valves Precision Fluidics

Miniature Proportional Valves Precision Fluidics Precision Fluidics 2 Electronic Pressure Control When you partner with the global leader in motion and control technologies, expect to move your business and the world forward. From miniature solenoid

More information

Matrix APAX. 380V-415V 50Hz TECHNICAL REFERENCE MANUAL

Matrix APAX. 380V-415V 50Hz TECHNICAL REFERENCE MANUAL Matrix APAX 380V-415V 50Hz TECHNICAL REFERENCE MANUAL WARNING High Voltage! Only a qualified electrician can carry out the electrical installation of this filter. Quick Reference ❶ Performance Data Pages

More information

JUMO MIDAS S05 Pressure transmitter Type

JUMO MIDAS S05 Pressure transmitter Type Data Sheet 40.00 Page /8 JUMO MIDAS S05 Pressure transmitter Type 4000 Brief description The JUMO MIDAS S05 pressure transmitter is available with both relative pressure and absolute pressure measurement

More information

Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA RG-131

Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA RG-131 ... the heart of your system. Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA 27106 RG-131 24 VOLT Various Ah, VALVE REGULATED, LEAD-ACID, AIRCRAFT BATTERY DECLARATION

More information

Micro Diaphragm Pumps (air/gas)

Micro Diaphragm Pumps (air/gas) 10 Up to 800 mlpm Free Flow Typical Markets Safety Patient Therapy Patient Monitoring Typical Applications Portable Gas Detection Gas Sampling Medical Instruments Trace Detection Sidestream CO 2 Negative

More information

800 Pin Receiver and ITA Modules and Contacts 2

800 Pin Receiver and ITA Modules and Contacts 2 TABLE OF CONTENTS 800 Pin Signal Module 800 Pin Receiver and ITA Modules and Contacts 2 Interface Devices (IDs) CASS Receiver 4 Light Weight CASS Receiver 5 80 Series Receiver 6 CASS Connector Module Configuration

More information

EGR MASS FLOW SYSTEM CONVERSION KIT (FROM TWO-PIECE TUBE WITH BELLOWS-TYPE CONNECTIONS TO ONE-PIECE TUBE WITH HOSE CONNECTIONS) ASET AC ENGINES

EGR MASS FLOW SYSTEM CONVERSION KIT (FROM TWO-PIECE TUBE WITH BELLOWS-TYPE CONNECTIONS TO ONE-PIECE TUBE WITH HOSE CONNECTIONS) ASET AC ENGINES SERVICE BULLETIN REV (Not applicable to Mack Trucks Australia) (Supersedes SB-214-026 dated 2/2/04) NUMBER: SB-214-026 DATE: 7/30/04 MODEL: ASET AC EGR MASS FLOW SYSTEM CONVERSION KIT (FROM TWO-PIECE TUBE

More information

20 amps 220 amps, 60 Hz VDC VAC G3PA-420B-VD-2 DC amps 440 amps, 60 Hz G3PA-450B-VD-2 DC12-24

20 amps 220 amps, 60 Hz VDC VAC G3PA-420B-VD-2 DC amps 440 amps, 60 Hz G3PA-450B-VD-2 DC12-24 Solid State Relays (SSRs) G3PA/G3PB Long Service Life for Circuits that Cycle Frequently D D D D D Built-in heat sink increases life and reliability Voltage turn-on at zero crossing reduces initial inrush

More information

ESCORT LC Pump. Instruction Manual WARNING

ESCORT LC Pump. Instruction Manual WARNING ESCORT LC Pump Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece of complex equipment,

More information

WATERFLUX 3070 Quick Start

WATERFLUX 3070 Quick Start WATERFLUX 3070 Quick Start Battery powered electromagnetic water meter Electronic Revision ER 4.3.0_ up to ER 4.3.4_ (SW.REV 4.2.2_ up to 4.2.5_) KROHNE CONTENTS WATERFLUX 3070 1 Safety instructions 4

More information

Description of Bolt Torque Tables

Description of Bolt Torque Tables Maximum (preferred) torque required per bolt (foot pounds) to arrive at the maximum available / recommended gasket stress shown in the previous column. (4) Maximum gasket stress recommended or available

More information

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation M T E C o r p o r a t i o n MATRIX FILTER SERIES B 380-415 Volts, 50HZ USER MANUAL PART NO. INSTR - 015 REL. 060628 2006 MTE Corporation IMPORTANT USER INFORMATION NOTICE The MTE Corporation Matrix Filter

More information

PTS SERIES With AE402 Indicator

PTS SERIES With AE402 Indicator This document is hosted by Old Will Knott Scales www.oldwillknottscales.com Adam Equipment PTS SERIES With AE402 Indicator (P.N. 3106610715, Revision E, April 2011) Adam Equipment 2011 Adam Equipment 2011

More information

Preparing the Site. Information About the Site Requirements CHAPTER

Preparing the Site. Information About the Site Requirements CHAPTER 2 CHAPTER This chapter describes the basic site requirements that you should be aware of as you prepare to install your Cisco Nexus 7000 Series switches. This chapter includes the following sections: Information

More information

Replacing the Internal Oil Filter Elements for HS85 and OS 85 Screw Compressors

Replacing the Internal Oil Filter Elements for HS85 and OS 85 Screw Compressors Maintenance Bulletin (MB-0038) Version 1, Sept 2015 Replacing the Internal Oil Filter Elements for HS85 and OS 85 Screw Compressors The internal oil filter elements are recommended to be changed on new

More information

First Responder System Identification Quiz. Presented by the NY-SUN PV Trainers Network

First Responder System Identification Quiz. Presented by the NY-SUN PV Trainers Network First Responder System Identification Quiz Presented by the NY-SUN PV Trainers Network SYSTEM 1 Questions: 1) Is there a solar electric system at the site? 2) What is your next step? SYSTEM 1 3) Given

More information

PETRODIST PILODIST. Crude oil distillation systems according to ASTM - standards for fractionation and boiling analysis

PETRODIST PILODIST. Crude oil distillation systems according to ASTM - standards for fractionation and boiling analysis With kind permission of BP Gelsenkirchen GmbH PETRODIST Crude oil distillation systems according to ASTM - standards for fractionation and boiling analysis Version 9/ January 2012 1 PETRODIST is a brand

More information

Each pump shall be equipped with carbon/graphite vanes. A vacuum relief valve shall be incorporated into the design of the pump.

Each pump shall be equipped with carbon/graphite vanes. A vacuum relief valve shall be incorporated into the design of the pump. General 1 Model MS 9-2-60 expandable/modular duplex central vacuum system consisting of two Becker or approved equivalent VT4.25 vacuum pumps mounted on a common skid with expandable automatic alternating

More information

VRLA Batteries. Battery Installation And Start up Guide

VRLA Batteries. Battery Installation And Start up Guide TECHNICAL MANUAL 41-7525 VRLA Batteries 26-206 Ampere-Hour Capacity Battery Installation And Start up Guide (For Rack Mounted Systems) 41-7525/0514/CD www.cdtechno.com Table of Contents 12V VRLA Battery

More information

HT STRAIGHT ADAPTER, WRENCH FLATS ( )

HT STRAIGHT ADAPTER, WRENCH FLATS ( ) HT7006 - STRAIGHT ADAPTER, WRENCH FLATS (12387118) MIL-DTL-38999 SERIES III; MIL-DTL-38999 SERIES IV HT7 0 06-11 - 6-8 SERIES ADAPTER ANGLE STRAIGHT - 0 CONNECTOR SERIES (INTERFACE) FINISH 8 - ELECTROLESS

More information

The pumps shall be Oil Free and Variable speed controlled by Pressure Transducer, PLC and inverter.

The pumps shall be Oil Free and Variable speed controlled by Pressure Transducer, PLC and inverter. General Model MS150-4-150 expandable/modular central vacuum system consisting of 4 Becker or approved equivalent KVX3.140-400 vacuum pumps mounted on a common skid with expandable automatic alternating

More information

High Frequency SineWave Guardian TM

High Frequency SineWave Guardian TM High Frequency SineWave Guardian TM 380V 480V INSTALLATION GUIDE FORM: SHF-IG-E REL. January 2018 REV. 002 2018 MTE Corporation High Voltage! Only a qualified electrician can carry out the electrical installation

More information

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 *

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * PRODUCT: Heavy Duty Stock Car Steel Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards

More information

Focus Area Level Report Including Knowledge and Skills, and Performance Indicators

Focus Area Level Report Including Knowledge and Skills, and Performance Indicators Including Knowledge and Skills, and PST 01. Use physical science principles and engineering applications with power, structural and technical systems to solve problems and improve performance. PST 01.01.

More information

Emergency lighting units EM converterled. EM converterled BASIC 200 V BASIC series

Emergency lighting units EM converterled. EM converterled BASIC 200 V BASIC series EM converter EM converter BASIC 200 V BASIC series Product description lighting Driver for manual testing For modules with a forward voltage of 50 200 V ow profile casing (21 x 30 mm cross-section) For

More information

1291BL Series Technical Specification Single Axis Position and Rate Table System

1291BL Series Technical Specification Single Axis Position and Rate Table System Datasheet 1291BL Series Technical Specification Single Axis Position and Rate Table System DESCRIPTION The Model 1291BL Single Axis Position and Rate Table System is designed to provide precise position,

More information

Product Data. Features/Benefits. 35K Bypass Terminal. 110 to 4400 cfm

Product Data. Features/Benefits. 35K Bypass Terminal. 110 to 4400 cfm Product Data 35K Bypass Terminal 110 to 4400 cfm Carrier s 35K Series bypass terminals offer: 20-gage, galvanized steel casing construction 1/2-in. thick, dual density fiberglass insulation meeting NFPA

More information

REFERENCE MANUAL FORM: MX-TRM-E REL REV MTE

REFERENCE MANUAL FORM: MX-TRM-E REL REV MTE Matrix APAX 380V-415V 50Hz TECHNICAL REFERENCE MANUAL FORM: MX-TRM-E REL. September 2014 REV. 002 2014 MTE Corporation WARNING High Voltage! Only a qualified electrician can carry out the electrical installation

More information

Emergency lighting units EM converterled. EM converterled BASIC 90 V BASIC series

Emergency lighting units EM converterled. EM converterled BASIC 90 V BASIC series BASIC 90 V BASIC series Product description lighting Driver for manual testing For self-contained emergency lighting For modules with a forward voltage of 40 97 V SEV for output voltage < 120 V DC ow profile

More information

06xx024S - STRAIGHT SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING 06xx024Y - ONE PIECE SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING

06xx024S - STRAIGHT SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING 06xx024Y - ONE PIECE SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING 06xx024S - STRAIGHT SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING 06xx024Y - ONE PIECE SHRINK BOOT ADAPTER, NON-DETENT SELF LOCKING LENGTH STYLE Y SPECIFICATION: WHEN THE DESIRED ADAPTER S C DIAMETER IS

More information

1291BL Series Technical Specification Single-Axis Rate and Positioning Table System

1291BL Series Technical Specification Single-Axis Rate and Positioning Table System Datasheet 1291BL Series Technical Specification Single-Axis Rate and Positioning Table System DESCRIPTION The Model 1291BL Single Axis Positioning and Rate Table System is designed to provide precise position,

More information

Electric Heat Tracing INSTALLATION PROCEDURES

Electric Heat Tracing INSTALLATION PROCEDURES Electric Heat Tracing INSTALLATION PROCEDURES Electric Heat Tracing Complete Electric Heat Tracing System... Illustration A: Typical Heat Tracing Installation A complete electric heat tracing system will

More information

Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA 91790

Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA 91790 Concorde Battery Corporation 2009 San Bernardino Road West Covina, California, USA 91790 RG-150 Series 24 VOLT 3.5 Ah, VALVE REGULATED, LEAD-ACID, AIRCRAFT BATTERY DECLARATION OF DESIGN PERFORMANCE TO

More information

Installation and Operation Manual

Installation and Operation Manual Industrial Process Installation and Operation Manual Advantage Actuator 2.0 Table of Contents Table of Contents Introduction and Safety...2 Safety message levels...2 User health and safety...2 Transportation

More information

MSC Manifolds - DIN for Steam Distribution and Condensate Collection Installation and Maintenance Instructions

MSC Manifolds - DIN for Steam Distribution and Condensate Collection Installation and Maintenance Instructions 1170650/3 IM-P117-17 ST Issue 3 MSC Manifolds - DIN for Steam Distribution and Condensate Collection Installation and Maintenance Instructions 1 General safety information 2 General product information

More information

Lifeline Medical Systems

Lifeline Medical Systems Installation, Operation and Service Manual Lifeline Medical Systems High Pressure Fully-Automatic Manifold Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783

More information

M-6642-MB Boss Oil Cooler INSTALLATION INSTRUCTIONS

M-6642-MB Boss Oil Cooler INSTALLATION INSTRUCTIONS !!! PLEASE READ THE FOLLOWING INSTRUCTIONS CAREFULLY PRIOR TO INSTALLATION!!! Caution: If you are not confident that you can complete the installation safely, have it completed by a certified technician!

More information

Emergency lighting units EM converterled. EM converterled BASIC 50 V BASIC series

Emergency lighting units EM converterled. EM converterled BASIC 50 V BASIC series BASIC 50 V BASIC series Product description lighting Driver for manual testing For self-contained emergency lighting For modules with a forward voltage of 10 52 V SEV for output voltage < 60 V DC ow profile

More information

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer Application Note: 42165 A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer Kristian J. Hoffman, Angela Seipel, Application Specialists,

More information

Levalarm Gold. Installation, Operation, & Maintenance Instructions R500.E253A 03/16/2016

Levalarm Gold. Installation, Operation, & Maintenance Instructions R500.E253A 03/16/2016 Installation, Operation, & Maintenance Instructions R500.E253A 03/16/2016 Levalarm Gold STORAGE AND HANDLING The Reliance Levalarm Gold meets or exceeds all applicable specifications when shipped from

More information

GUIDE FOR MICROGENERATION INTERCONNECTION TO CITY OF MEDICINE HAT ELECTRIC DISTRIBUTION SYSTEM

GUIDE FOR MICROGENERATION INTERCONNECTION TO CITY OF MEDICINE HAT ELECTRIC DISTRIBUTION SYSTEM GUIDE FOR MICROGENERATION INTERCONNECTION TO CITY OF MEDICINE HAT ELECTRIC DISTRIBUTION SYSTEM Page 1 of 19 Table of Contents 1.0 SCOPE...4 2.0 PURPOSE...4 3.0 LIMITATIONS...5 4.0 GENERAL INTERCONNECTION

More information

Harris IRT Enterprises Digital Resistance Tester Model XP

Harris IRT Enterprises Digital Resistance Tester Model XP Harris IRT Enterprises Digital Resistance Tester Model 5012-06XP Specifications & Dimensions 2 Theory of Operation 3 Operator Controls & Connectors 4 Test Connections 5 Calibration Procedure 6-7 Options

More information

CONTENTS 1. INTRODUCTION SAFTY INSTRUCTION CABLE CONNECTION SYSTEM DESCRIPTION INVERTER OPERATION...

CONTENTS 1. INTRODUCTION SAFTY INSTRUCTION CABLE CONNECTION SYSTEM DESCRIPTION INVERTER OPERATION... CONTENTS 1. INTRODUCTION...... 1 2. SAFTY INSTRUCTION.......... 2 3. CABLE CONNECTION.......... 4 4. SYSTEM DESCRIPTION............ 5 5. INVERTER OPERATION... 11 6. TROUBLE SHOOTING GUIDE....... 16 7.

More information

Safety Contact Strips. Safety Contact Strips

Safety Contact Strips. Safety Contact Strips Contact Strips, Bumpers, Mounting and Controllers Profile materials EPDM or NBR rubber Provides housing for safety contact Available in sizes to fit many applications Safety Bumpers Foam rubber covered

More information

COLD PLATE SOFTWARE PROGRAM ANALYZES AIRCRAFT

COLD PLATE SOFTWARE PROGRAM ANALYZES AIRCRAFT COLD PLATE SOFTWARE PROGRAM ANALYZES AIRCRAFT DISPLAY T. Renaud Sanders, a Lockheed Martin Co. Nov, 2000 Introduction Finned heat exchangers, called cold plates, have been used for many years to cool military

More information

Miniature Pneumatic Solenoid Valve

Miniature Pneumatic Solenoid Valve 10 mm Solenoid-Actuated Poppet Valve Typical Markets Respiratory Patient Therapy Typical Applications Oxygen Concentrators - Sieve bed switching/equalization - Oxygen delivery Deep Vein Thrombosis - Cuff

More information

Cylinder Valve Assembly Series

Cylinder Valve Assembly Series Cylinder Valve Assembly Series CV, PCV, FCV and MCV series CV Series (Cylinder/Valve) Cylinder/Valve Assembly with: Cylinder & Valve PCV Series (Pre-Engineered Nozzle/Cylinder/Valve) Cylinder/Valve Assembly

More information

AIRCRAFT GENERAL KNOWLEDGE (1) AIRFRAME/SYSTEMS/POWERPLANT

AIRCRAFT GENERAL KNOWLEDGE (1) AIRFRAME/SYSTEMS/POWERPLANT 1 In flight, a cantilever wing of an airplane containing fuel undergoes vertical loads which produce a bending moment: A highest at the wing root B equal to the zero -fuel weight multiplied by the span

More information

SITRANS AS 100. Instruction Manual September 2001 ITRANS AS 100

SITRANS AS 100. Instruction Manual September 2001 ITRANS AS 100 SITRANS AS 100 Instruction Manual September 2001 ITRANS AS 100 Safety Guidelines Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the

More information

PIPING PACKAGES TABLE OF CONTENTS

PIPING PACKAGES TABLE OF CONTENTS PIPING PACKAGES TABLE OF CONTENTS ITEM PAGE General Notes and Legend........................................................ 3 Piping Package Code Descriptions & Diagrams VAV Products and HL, HP, VF, and

More information

Installation Guide Smart-UPS On-Line SRT1000/1500 UXI-NCLI, SRT1000/1500 UXI-LI, Tower/Rack-Mount

Installation Guide Smart-UPS On-Line SRT1000/1500 UXI-NCLI, SRT1000/1500 UXI-LI, Tower/Rack-Mount Installation Guide Smart-UPS On-Line SRT1000/1500 UXI-NCLI, SRT1000/1500 UXI-LI, Tower/Rack-Mount Important Safety Messages Read the instructions carefully to become familiar with the equipment before

More information

Hydro Plant Risk Assessment Guide

Hydro Plant Risk Assessment Guide September 2006 Hydro Plant Risk Assessment Guide Appendix E8: Battery Condition Assessment E8.1 GENERAL Plant or station batteries are key components in hydroelectric powerplants and are appropriate for

More information

CP-250E-60/72-208/240-MC4 Microinverter with Modular Trunk Cable

CP-250E-60/72-208/240-MC4 Microinverter with Modular Trunk Cable CP-250E-60/72-208/240-MC4 Microinverter with Modular Trunk Cable Chilicon Power Aug 2016 1 CONTENTS CP-250E Microinverter System... 3 The CP-100 Cortex Gateway... 3 Important Safety Information... 4 Inverter

More information

Linear Flexible Joint Cart Plus Single Inverted Pendulum (LFJC+SIP)

Linear Flexible Joint Cart Plus Single Inverted Pendulum (LFJC+SIP) Linear Motion Servo Plants: IP01 and IP02 Linear Flexible Joint Cart Plus Single Inverted Pendulum (LFJC+SIP) User Manual Table of Contents 1. Linear Flexible Joint Cart Plus Single Inverted Pendulum System

More information

Crispin Valves Operating Guide. Crispin

Crispin Valves Operating Guide. Crispin Crispin Valves Operating Guide Crispin Since 1905 Crispin Multiplex Manufacturing Co. 600 Fowler Avenue Berwick, PA 18603 1-800-AIR-VALV T: (570) 752-4524 F: (570) 752-4962 www.crispinvalve.com sales@crispinvalve.com

More information

Emergency lighting units EM powerled

Emergency lighting units EM powerled M BASIC 1 W Emergency lighting Driver 1 W Product description Emergency lighting Driver for manual testing For self-contained emergency lighting on maintained operation SEV for output voltage < 60 V DC

More information

DHE Chassis and Cooling

DHE Chassis and Cooling DHE Chassis and Cooling DES Mechanical Workshop at FNAL November 2, 2005 Vaidas Simaitis High Energy Physics Group University of Illinois at Urbana-Champaign 1 6-SLOT + 4-SLOT CHASSIS 2 FAN MANIFOLDS ADDED

More information

PRYCO, INC. P. O. BOX 108 Mechanicsburg, IL OPERATIONS AND MAINTENANCE MANUAL For PUMP SETS

PRYCO, INC. P. O. BOX 108 Mechanicsburg, IL OPERATIONS AND MAINTENANCE MANUAL For PUMP SETS PRYCO, INC. P. O. BOX 108 Mechanicsburg, IL 62545 Telephone: 217 / 364-4467 Fax: 217 / 364-4494 OPERATIONS AND MAINTENANCE MANUAL For PUMP SETS WHAT TO DO FIRST Upon receiving your new pump set system

More information

Table of Contents. 1. Model Designation And Unit Models Guide Specifications Technical Data Performance Data 10

Table of Contents. 1. Model Designation And Unit Models Guide Specifications Technical Data Performance Data 10 Table of Contents Page No. 1. Model Designation And Unit Models 1 2. Features 2 3. Guide Specifications 4 4. Standard Unit Components 5 5. Technical Data 6 6. Performance Data 10 7. Electrical Data 19

More information

W91/W94 Series TEMPERATURE REGULATORS. Self-Operated Temperature Regulators. Design & Operation W91 Non-Indicating W94 Dial Thermometer

W91/W94 Series TEMPERATURE REGULATORS. Self-Operated Temperature Regulators. Design & Operation W91 Non-Indicating W94 Dial Thermometer Design & Operation W91 Non-Indicating W94 Dial Thermometer Watson McDaniel reserves the right to change the designs and/or materials of its products without notice. 2010 Watson McDaniel Company CAPILLARY

More information

WARNING. Diode Package The Murphy diode package ( ) is designed to allow the fuel shutoff valve to be used with dual Magneto Ignition systems.

WARNING. Diode Package The Murphy diode package ( ) is designed to allow the fuel shutoff valve to be used with dual Magneto Ignition systems. M2582, M508, and M580 Series Electromechanical and Pneumatic, Fuel Gas Shutoff Valves Installation and Operation Instructions Please read the following instructions before installing. A visual inspection

More information

Product overview. Technical data

Product overview. Technical data VALVES & CONTROLS ANDERSON GREENWOOD BLANKETING SYSTEMS The Complete Blanketing and Safety Valve System! Product overview Pad Valves Anderson Greenwood is a world leader in tank protection and blanketing.

More information

Emergency lighting units EM powerled

Emergency lighting units EM powerled EM power EM power SEFTEST 1 2 W Combined emergency lighting Driver 1 4 W Product description Emergency lighting Driver with self-test function For self-contained emergency lighting SEV for output voltage

More information

Operators Manual ECO-AIR-IVS Internal Valve Actuation System

Operators Manual ECO-AIR-IVS Internal Valve Actuation System 2534-A Shell Road Georgetown, TX 78628 www.parafour.com 512-626-4099 Operators Manual ECO-AIR-IVS Internal Valve Actuation System Ver 1.0 Warnings:!! This device is electrically operated and should only

More information

Emergency lighting units EM powerled

Emergency lighting units EM powerled M BASIC 1 W Emergency lighting Driver 1 W Product description Emergency lighting Driver for manual testing For self-contained emergency lighting on maintained operation SEV for output voltage < 60 V DC

More information

EVAP system, servicing

EVAP system, servicing Page 1 of 65 20-130 EVAP system, servicing EVAP system components 1 - Cap nut 10 Nm 2 - Cover 3 - Stud For EVAP canister 15 Nm 4 - Sealing piece 5 - Bleed line To EVAP canister purge regulator valve -

More information

Amphenol 348 Series High Density Connectors

Amphenol 348 Series High Density Connectors 348 Series High Density Connectors 12-093-6 348 Series I, II Table of Contents 348 Series Page No. High Density Connectors........................................................................ 1 Specifications.................................................................................

More information

ATD WATT INVERTER

ATD WATT INVERTER ATD-5954 1500 WATT INVERTER Welcome. This ATD product has been carefully engineered and manufactured to give you dependable operation. Please read this manual thoroughly before operating your new ATD product,

More information

PowerOhm Installation Manual for LG ATV Series Braking Modules

PowerOhm Installation Manual for LG ATV Series Braking Modules PowerOhm Installation Manual for LG ATV Series Braking Modules IMPORTANT: These instructions should be read thoroughly before installation. All warnings and precautions should be observed for both personal

More information

ARGONITE Fire Protection Systems

ARGONITE Fire Protection Systems GENERAL INFORMATION The Chemetron ARGONITE Fire Protection Systems are clean agent, automatic extinguishing systems using Argonite (IG55) and consisting of four basic components and their associated accessories.

More information

SmartOnline 120V 2.2kVA 1.8kW Double-Conversion UPS, 2U, Extended Run, Network Card Slot, LCD, USB, DB9

SmartOnline 120V 2.2kVA 1.8kW Double-Conversion UPS, 2U, Extended Run, Network Card Slot, LCD, USB, DB9 SmartOnline 120V 2.2kVA 1.8kW Double-Conversion UPS, 2U, Extended Run, Network Card Slot, LCD, USB, DB9 MODEL NUMBER: AL-84111 Specifications OUTPUT Output Volt Amp Capacity (VA) 2200 Output kva Capacity

More information

BAe 146 NOXY OPERATION MANUAL

BAe 146 NOXY OPERATION MANUAL BAe 146 NOXY OPERATION MANUAL AQD-307200 By: Martin Buhr Air Quality Design, Inc. Golden, Colorado Prepared for: AAM Cranfield, UK 2008 TABLE O CONTENTS Section Page 1. GENERAL SYSTEM DESCRIPTION...3 1.1

More information

Application Note Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines

Application Note Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines Application Note 83404 Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines Woodward reserves the right to update any portion of this publication

More information

S v S e r i e s F M f i r e e x t i n g u i s h i n g s y s t e m

S v S e r i e s F M f i r e e x t i n g u i s h i n g s y s t e m Clean S v S e r i e s F M - 2 0 0 f i r e e x t i n g u i s h i n g s y s t e m Advantages The Sv Series Clean Agent Fire Extinguishing System utilizes FM-200 as the extinguishing medium. FM-200 is a colorless,

More information

Ignition Reliability in SGT-750 for Gas Blends at Arctic Conditions. Magnus Persson Combustion Expert / Distributed Generation / Sweden

Ignition Reliability in SGT-750 for Gas Blends at Arctic Conditions. Magnus Persson Combustion Expert / Distributed Generation / Sweden Ignition Reliability in SGT-750 for Gas Blends at Arctic Conditions Magnus Persson Combustion Expert / Distributed Generation / Sweden siemens.com/power-gas Table of content Objectives of the Project SGT-750

More information

Emergency lighting units EM converterled

Emergency lighting units EM converterled EM converter EM converter BASIC icd 50 V BASIC series Product description lighting Driver for manual testing For self-contained emergency lighting For modules with a forward voltage of 10 54 V SEV for

More information

VENT SILENCER PRODUCT GUIDE

VENT SILENCER PRODUCT GUIDE VENT SILENCER Copyright 200 by PULSCO Incorporated. All rights reserved. Reproduction without permission is prohibited. PRODUCT GUIDE PULSCO VENT SILENCER TABLE OF CONTENTS DESCRIPTION PAGE PULSCO VENT

More information

Emergency lighting units EM converterled

Emergency lighting units EM converterled EM converter EM converter BASIC icd/imh 250 V BASIC series Product description lighting Driver for manual testing For self-contained emergency lighting For modules with a forward voltage of 50 250 V ow

More information

From. McCROMETER, INC WEST STETSON AVENUE HEMET, CA Phone (951) Toll Free (800) Fax (951)

From. McCROMETER, INC WEST STETSON AVENUE HEMET, CA Phone (951) Toll Free (800) Fax (951) SPI Mag MODEL 282L SINGLE POINT INSERTION ELECTROMAGNETIC FLOW METER 2 SENSOR SUBMITTAL PACKAGE From McCROMETER, INC. 3255 WEST STETSON AVENUE HEMET, CA 92545 Phone (951) 652-6811 Toll Free (800) 220-2279

More information

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation

M T E C o r p o r a t i o n MATRIX FILTER. SERIES B Volts, 50HZ USER MANUAL PART NO. INSTR REL MTE Corporation M T E C o r p o r a t i o n MATRIX FILTER SERIES B 380-415 Volts, 50HZ USER MANUAL PART NO. INSTR - 015 REL. 040709 2003 MTE Corporation IMPORTANT USER INFORMATION NOTICE The MTE Corporation Matrix Filter

More information

Installation Guide Smart-UPS 1200/1500 VA 100/120/230 Vac Rack-Mount 1U

Installation Guide Smart-UPS 1200/1500 VA 100/120/230 Vac Rack-Mount 1U Installation Guide Smart-UPS 1200/1500 VA 100/120/230 Vac Rack-Mount 1U Important Safety Messages SAVE THESE INSTRUCTIONS - This section contains important instructionsthat should be followed during installation

More information

Technical specifications

Technical specifications KINGDOM OF SAUDI ARABIA THE SAUDI FUND FOR DEVELOPMENT ( SFD) ITB -207-7 Technical specifications Project name: Reconstruction, Furniture and Equipping of Three Schools Photovoltaic Solar System for Holly

More information

RADIO FREQUENCY SYSTEMS

RADIO FREQUENCY SYSTEMS C14520.016-E.MU0 DOC.:10000008168-02 LAB4 DEHYDRATOR USER MANUAL 1 DRYING CHAMBER 2 HEATER 3 PUMP 4 AIR BACKWASHING HOLE 5 HUMIDITY PROBE 6 DIGITAL PRESSURE GAUGE 7 RELIEF VALVE 8 SHUT OFF VALVE 9 DRY

More information

An advisory circular may also include technical information that is relevant to the rule standards or requirements.

An advisory circular may also include technical information that is relevant to the rule standards or requirements. Revision 0 Electrical Load Analysis 2 August 2016 General Civil Aviation Authority advisory circulars contain guidance and information about standards, practices, and procedures that the Director has found

More information