Accumate 2300 Volumetric Infusion Pump
|
|
- Shannon Grant
- 5 years ago
- Views:
Transcription
1 Accumate 2300 Volumetric Infusion Pump Operating Manual Log o
2 Important This Operating Manual is for clinician use only. Do not permit patients to have access to this Operating Manual. Please read the entire Operating Manual before operating the Accumate 2300 volumetric infusion pump. Failure to properly follow warnings, cautions and instructions could result in death or serious injury to the patient. 2
3 Warnings If the pump is used to deliver life-sustaining medication, an additional pump must be available. Prior to use, always check the audible and visual alarms. The pump is not to be used for delivery of blood or cellular blood products. Do not use this pump on mains power if the integrity of the protective earth system is suspect. To prevent a possible explosion hazard do not use this pump in presence of inflammable anesthetic gases. Do not use this pump in close proximity to equipment such as surgical diathermy with generates high levels of RFI. Do not permit the use of mobile telephones in wards where this pump is in use. If the pump is dropped or hit, inspect the pump for damage. Do not use a pump that is damaged or not functioning properly. Contact Customer Service to return a pump for service. Do not permit this pump to used if standing in a pool of liquid. System delivery inaccuracies may occur as a result of back pressure or fluid resistance, which depends upon drug viscosity, catheter size, and extension set tubing. 3
4 Warnings Always read and follow the manual that accompany the source container and IV administration sets you are using. Carefully follow the manual for loading, removing, and reloading the set, as well as recommended set change intervals. Change pumping section on tubing every 24 hours. Do not use tubing for more than 72 hours. Always check pump data prior to starting infusion. After starting an infusion, make sure drops are falling in the drip chamber. If no drops are falling, make sure the roller clamp is open. If the roller clamp is open and still no drops are falling, replace and dispose of the set. The air detector cannot recognize the introduction of air at 3-way stopcocks, infusion ports, and other lines/tubes below the pump. To avoid mechanical or electronic damage, do not steam autoclave or immerse the pump in any fluids or cleaning solutions. Always disconnect electrical power cord from outlet before cleaning to prevent electrical shock. 4
5 Cautions Do not operate the pump at temperatures below +10 C(50 F) or above 40 C(104 F) and humidity levels below 30% or above 85%. Do not store the pump at temperatures below +5 C(41 F) or above 40 C(104 F) and humidity levels below 10% or above 90%. Do not store the pump with the Administration Set attached. Do not store the pump for prolonged periods with the battery installed (above 3months). Do not immerse the pump in cleaning fluid or water or allow solution to soak into the pump, accumulate on the key pad, or enter the battery compartment. Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners. Do not use the pump to therapeutic levels of ionizing radiation. Do not use the pump directly to ultrasound. Do not use the pump in the vicinity of magnetic resonance imaging (MRI) equipment. Do not use the pump near ECG equipment. Do not sterilize the pump. All service and repair of the Accumate 2300 pumps must be performed by Woo Young Medical or its authorized agents. 5
6 Table of Contents Introduction... 7 Features of the Accumate 2300 Front View & Key Pad...8 Controls and Indicators...9 R e a r V i e w Directions for Operation Preparation of the IV-set...12 Turning ON the Power - Rear...13 Turning ON the Power - Front...14 Insertion of the IV-set...15 Setting the Flow Rate and Delivery Limit...17 Clearing the Volume Delivered...18 Operating the Pump...18 Purge Infusion (stop mode)...19 Bolus Infusion (start mode)...19 T itration...20 Data Lock...21 Adjust Sound Occlusion Level Near End of Infusion...24 End of Infusion Alarms and Troubleshooting...26 Specifications Maintenance Routine Maintenance Procedures...30 Cleaning and Storage...31 Replacing the AC fuses...32 Battery operation Occlusion Pressure Limits...34 Start-up & Trumpet Curves...35 Warranty...38 Service and Packaging Contact
7 Introduction The Accumate 2300 volumetric infusion pump is a fully featured infusion pump suitable for critical care and generation applications. The Advantages High precision. Purge, automatic & manual bolus capability. Easy to operate in only three steps. Power On > Select (rate, total, VTBI) > Start Easy to read, large display. History function. Historical data of infusion records. (warnings, settings, starts and stops of infusion, etc.) Micro-infusion for neonatal and high risk application. None dedicated sets, available from various manufacturers. Programmable, ultrasonic air bubble detector. Multi-position clamp for vertical or horizontal fixation to poles. SAFETY STANDARD The Accumate 2300 meets all safety standards for medical electrical devices, corresponding to IEC and IEC
8 Features of the Accumate Front View & Key Pad b c d e a n m j l k i f g h o p q r s t u 8 y x w v
9 Controls and Indicators NO. Name Function a Handle To handle for carrying b Name Plate To display the pump model and name c Display All essential information is displayed (flow rate, delivery limit, volume delivered, alarms information) d Locker To lock pump door e Key Pad To set infusion data f Finger To squeeze the tube with the spring housing by peristaltic motion g Occlusion Sensor To detect the downstream occlusion of the tube h Clamp Lever To prevent free flow when the pump door is opened i Spring Housing To squeeze the tube with the peristaltic fingers j Door Handle To lock pump door with locker k Door To protect the tubing set when running l Door Magnet To detect whether the door is open or not m Bubble Sensor To detect air in the tube n LED To indicates pump is infusing or lights up during an alarm (normal : green, malfunction : red) 9
10 Controls and Indicators NO. Name Function o Volume Delivered Displays Volume Delivered (VTBI:Volume To Be Infused) p Delivery Limit Displays delivery limit or alarms information q Flow Rate Displays flow rate (infusion rate) r s t SELECT /SILENCE BATTERY LEVEL EXTERNAL POWER To select setting mode (flow rate, delivery limit, volume delivered) / To silence alarm for 2minutes To display the charging status of internal rechargeable battery To inform whether the external power is connected or not u PURGE To access PURGE or BOLUS key v POWER ON/OFF Press once to switch the pump ON. Press and hold down for 2 seconds to switch the pump OFF w START To start the infusion x STOP /CLEAR To stop the infusion /To clear the value when setting data y UP/DOWN To increase or decrease of value shown on display 10
11 Features of the Accumate Rear View a b l c d k e j f g h i NO. Name Function a Multi-position Clamp To fix the pump to vertical pole stand or horizontal bar of bed b Rear Support To support to pole stand c Drop Sensor Port To connect the drop sensor with the pump d Nurse Call/ Comm. Port To connect the nurse call jack or communication cable e Front Panel Lock Switch To lock the front key pad f 12V DC Jack To supply the power of 12V in ambulance car g Protective Ground To connect the protective earth line h Spec Plate To inform the product information i AC Outlet/Fuse Socket To connect the AC 90~245V main power cord where the 1.25A fuse is j Rechargeable Battery To hold the NIMH rechargeable battery k Power Switch To switch the main power l Clamp Knob To fix the clamp 11
12 Direction for Operation Preparation of the IV-set Check for cleanliness, completeness, and damage. If unit appears damaged, do not use it. Check the IV administration sets. If other IV administration sets are used, the operating safety of the pump can no longer be guaranteed. Patient safety may, as a result, be compromised. Recommended IV-set BD(BECTON DICKINSON) :A122 KV(KOREA VACCINE) :143T/243T Carefully purge the infusion set, without allowing any air bubbles to enter, until the drip chamber is to 1/3 full. If air has entered, repeat purgeprocedure. Close the tubing roller clamp. chamber 1/3 close 12
13 Direction for Operation Turning ON the Power - Rear Attach the pump to an IV pole by turning the knob on pole clamp. Plug the pump's power cord into an electrical outlet. Turn ON the MAIN POWER switch. MAIN POWER switch must be ON to charge the battery. 13
14 Direction for Operation Turning ON the Power - Front Check whether the AC power LED is turned on. External Power LED display will light. Press the POWER ON key on the front panel. When turning the pump OFF, the pump must be at STOP mode in order for the POWER key to respond. Wait the self test procedure. Check the previously used data. 14
15 Direction for Operation Insertion of the IV-set Insert the IV-set chamber into the IV bottle/bag. Fill 1/3 of the chamber. Take cautions to avoid any air bubbles. Open the door by pulling the DOOR HANDLE. Completely insert the IV-set into the BUBBLE SENSOR. Push the CLAMP LEVER to right to open. Place the IV-set into the clamp. This is a safety feature that prevents free flow when the door is opened during infusion. 15
16 Direction for Operation Insert the IV-set in the occlusion guide. Check if the IV-set is straightforward. CAUTION : Make sure that IV tube is NOT loosely fastened into the internal chamber of the pump. It should be straightforward. Close the door by pushing the tightly DOOR HANDLE. The roller clamp should be below the pump. Open the roller clamp to be ready to infuse. Check that no drops are falling in the drip chamber. 16
17 Direction for Operation Setting the Flow Rate and Delivery Limit Press the SELECT key. The digits in the LED digital display will blink in 7seconds. Use the UP( ) DOWN( ) keys to set the desired rate of infusion. The rate of infusion can be set from 0.1 to 999.9ml/hr. Press the SELECT key. The digits in the LED digital display will blink in 7seconds. Use the UP( ) DOWN( ) keys to set the desired volume to be infused. The volume to be infused can be set from 0.1 to 9,999ml. 17
18 Direction for Operation Clearing the Volume Delivered Press the SELECT key. The digits in the LED digital display will blink in 7seconds. Press the STOP/CLR key. Operating the Pump Press the START key. Check the three LED turned on sequentially. 18
19 Direction for Operation Purge Infusion (stop mode) Press and hold the PURGE key when the pump is not infusion. Press and hold the PURGE key until fluid flows and the purging of the IV-set is complete. CAUTION : PURGE will infuse at the rate of 999.9ml/hr. The volume used during purging will be displayed, but it is not added to the volume delivered. The pump will not purge if the lock has been enabled. Bolus Infusion (start mode) Press the PURGE key when the pump is infusion. PURGE will infuse at the rate of 999.9ml/hr. The feature allows a bolus to be given during an infusion. Enter the amount of bolus to be given in ml and Press the START key. The amount of bolus can be set from 1.0 to 99.9ml. 19
20 Direction for Operation Titration (To change the flow rate while infusing) Press the SELECT key. The digits in the LED digital display will blink in 7seconds. Use the UP( ) DOWN( ) keys to set the desired rate of infusion. The rate of infusion can be set from 0.1 to 999.9ml/hr. Press the START key. CAUTION : The pump will not titration if the lock has been enabled. 20
21 Direction for Operation Data Lock (This features prevents unauthorized use by locking out all key except SILENCE key.) Turn ON [KEY LOCK] button to lock the key pad on the front panel. Press the any key and check the displayed in '----'. Check the set up and pump status again. 21
22 Direction for Operation Adjust Sound (This feature allows the volume of the alarm tone to be changed) Turn ON [KEY LOCK] button to lock the key pad on the front panel. Press and hold the SELECT key and UP( ) /DOWN( ) keys to increase or decrease the sound level. Select the volume of the alarm tone to be changed. caution : The volume of the alarm to change can be set from 1(45dB/1m) to 5(65dB/1m) 22
23 Direction for Operation Occlusion Level (The pressure at which a downstream occlusion alarm will be triggered varied. The higher the pressure limit, the less sensitive the pump is changes in fluid resistance.) Turn ON [KEY LOCK] button to lock the key pad on the front panel. Press and hold the SELECT key and UP( )/DOWN( ) keys to increase or decrease the occlusion level. Select the occlusion level to be changed. caution : The occlusion level to change can be set low(7± 2psi), center(11± 3psi), or high(15± 3psi). 23
24 Direction for Operation Near End of Infusion (Sets the near end of infusion warning time, as time left to End of Infusion.) Turn ON [KEY LOCK] button to lock the key pad on the front panel. Press and hold the SELECT key and UP( )/DOWN( ) keys to increase or decrease. Select the near end of infusion to be changed. caution : The end of infusion to change can be set from 0(no set) to 15minutes. 24
25 Direction for Operation End of Infusion (The pump has reached the end of the infusion) In certain conditions the Accumate 2300 automatically switches to the KVO (Keep-Vein- Open) rate. The KVO-rate is set to 1.0ml/hr. However, if the flow-rate set by the user is lower than 1.0ml/hr, the KVO-rate is equal to flow-rate. This alarm occurs 5 minutes before infusion is completed. Press the SILENCE key. This alarm occurs when infusion is completed. (KVO : Keep Vein Open) Press the STOP key to reset. 25
26 Alarms and Troubleshooting Alarms are indicated by a combination of an audible alarm, flashing alarm indicator and abbreviated message in the display. First press the SILENCE key to silence the alarm for a maximum of 2 minutes, then check the display for an alarm message. Press the STOP key to cancel the alarm message. If the infusion has stopped, rectify the cause of the alarm then press the START key resume the infusion. 26
27 Alarms and Troubleshooting Alarm Indication & Sound Air-in-line Air bb-bb-bb Door Open OPEn bb-bb-bb Occlusion OccL bb-bb-bb Drive Error Err bb-bb-bb Low battery BAtt b-b-b End of infusion End, conp b-b-b No Good ng-? silence Possible Cause Air bubble air-in-line sensor IV tube not properly inserted into the Bubble Sensor Door opened during infusion Downstream tubing kinked Filter blocked Closed roller clamp IV positional or infiltrated Catheter or vein too small Occlusion limit too low Motor malfunction Mechanical error 30 minutes until battery is fully discharged Programmed infusion is complete Self-Test NG Internal Malfunction 27 Action Remove the air from the tube Completely insert the IV tube into the Bubble Sensor Close the door Correct kinks Correct blocked filter Open roller clamp Check IV site Increase occlusion limit Request A/S Request A/S Connect pump to AC power Stop infusion and reset Request A/S
28 Specifications Name Volumetric Infusion Pump Model Accumate 2300 Pumping System Flow Rate Range Delivery Limit Range Total Volume Purge Rate Range Accuracy KVO( Keep-Vein- Open) Linear peristaltic ml/hr (0.1ml/hr increments) ml (0.1 ml increments), 1,000-9,999ml(1ml increments) Equals to the upper limit of the delivery limit range Equals to the upper limit of the flow-rate range +/- 5% nominal ml/hr(0.1ml/hr increments) Adjustable occlusion[ psi(3 level : low, center, high) Adjustable alarm volume (level up or down) Infusion completed Alarms Low battery(30 min. before infusion stop) Battery depleted(3 min. before power off) Air-in-line Door open Drive error Near End of Infusion Titration while running Others Self testing (System check) One key bolus Front panel lock 28
29 Specifications EMS Electrical Safety Dimensions Weight Power Requirements Battery Battery Operation Battery Charging Time DC 12V Class I, Type BF 140(W)*225(H)*140(D)mm Approximately 2.5Kg AC V, 50/60Hz Built-in rechargeable dc 12V(NI-MH 2.3Ah) Ca. 100ml/hr 7hrs Operation Conditions (Relative humidity 30-85%) Storage Conditions 5-40 (Relative humidity 10-90%) Accessories Warranty Options AC power cord for 220V, Operating manual 2 years from the date of delivery RS-485, nurse call, drop sensor CAUTION : Specifications are subject to change without notice. 29
30 Maintenance Routine Maintenance Procedures To ensure that pump remains in good operating condition, it is important to keep it clean and carry out routine maintenance procedures described below. All servicing should only be performed by a qualified service engineer with reference to the Technical Service Manual for this product. As required Thoroughly clean external surfaces of the pump before and after prolonged period of storage. At least once per year Inspect AC power supply plug and cable for damage. Perform functional tests as outlined in the Technical Service Manual. Operate the pump on battery power until the battery low alarm then charge the battery to confirm battery operation and charging. 30
31 Maintenance Cleaning and Storage Before the transfer of the pump to a new patient and periodically during the use, clean the pump by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant/detergent solution. Disinfectants which are known to be corrosive to metals must not be used. Concentrated Isopropyl alcohol based cleaners will degrade plastic parts. The IV-set and extension lines are disposable single use items and should be discarded after use according to their manufacturers' instructions. If the pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packing for protection. Once every 3 months during storage, carry out functional tests as described in the Technical Service Manual and ensure that the internal battery is fully charged. Caution : Before cleaning always switch OFF disconnect from the AC power supply. Never allow liquid to enter the casing and excess fluid build up on the pump. Do not steam autoclave, ethylene oxide sterilize or immerse this pump in any fluid. Keep the pump plugged into electrical power during storage. 31
32 Maintenance Replacing the AC fuses If the pump continually illuminates the battery indicator and the AC power indicator light does not illuminate when the pump is connected to the AC power supply and switched ON, suspect that either the power supply fuse in the AC plug or the internal fuses have blown. First check the power supply fuse in the AC mains plug. If the AC power indicator light does not illuminate remove the pump from service. It is recommended that only a qualified service engineer replaces the AC fuses. For further information regarding the replacement of internal AC fuses refer to the Technical Service Manual. 32
33 Maintenance Battery Operation The internal rechargeable battery allows continued operation when the AC power is unavailable, for example during patient transfer or AC power failure. A fully charged battery will provide over 4 hours of operation at typical infusion rates. A low battery alert is activated approximately 30 minutes before the battery depleted alarm. The battery is maintenance free, sealed Nickel Metal Hydride and requires no routine servicing. However, to achieve optimum operation, ensure that the battery is fully recharged after full discharge, before storage, and at regular 3 month intervals during storage. Charge retention will eventually degrade. Where retention is critical the internal battery should be replaced every 3 years. It is recommended that only a qualified service engineer replaces the Battery. For further information regarding the replacement of batteries refer to the Technical Service Manual. 33
34 ml Occlusion Level(psi) Occlusion Pressure Limits Time to alarm following occlusion is achieved in less than 5minutes at rates of 25ml/hr and higher by the appropriate selection of occlusion levels. The following graphs show the typical values for time to alarm and bolus volume that can be expected in the event of an occlusion when the IV-set (KV 243T) is selected Time to Alarm - 25ml/hr 0 1:00 2:00 3:00 4:00 min : sec Bolus Volume Occlusion Level(psi) 34
35 Start-up Curves & Trumpet Curves In this pump, as with all infusion systems, the action of the pumping mechanism and variations in individual IV-sets cause short-term fluctuations in rate accuracy. The following curves show typical performance of the system in two ways : 1) the delay in onset of fluid flow when infusion commences (start-up curves)... and 2) the accuracy of fluid delivery over various time periods is measured (trumpet curves). The start-up curves represent continuous flow versus operating time from the start of the infusion. They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual representation of uniformity. Trumpet curves are derived from the second hour of this data. Tests performed per IEC standard. 35
36 Start-up Curves & Trumpet Curves Trumpet curves are named for their characteristic shape. They display discrete data averaged over particular time periods or 'observation windows', not continuous data versus operating time. Over long observation windows, short term fluctuations have ㅣ little effect on accuracy as represented by the flat part of the curve. As the observation window is reduced, short term fluctuations have greater effects as represented by the "mouth" of the trumpet. Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered. Short term fluctuations in rate accuracy may have impact depending on the half-life of the drug being infused, therefore the clinical effect cannot be determined from the trumpet curves alone. Caution : Start-up and trumpet curves may not be indicative of operation under negative pressure. 36
37 Error(%) Rate(ml/hr) Start-up Curves & Trumpet Curves Start-up (25ml/hr) 17s 2m 5m 9m Time 13m 17m 31m Trumpet Curve (25ml/hr) 15.00% 10.00% 5.00% 0.00% -5.00% 17s 1m 5m 11m 19m 31m % % Observation Window(mins) 37
38 Warranty Woo Young Medical Co., Ltd. offers a 24-month warranty on every Accumate 2300 unit (except for the battery), effective from the date of delivery. The battery is free from defects in material and workmanship under normal use and service for a period one hundred eighty(180) days from the date of delivery by Woo Young Medical to the original purchaser. This includes the repair or replacement of damaged parts as a result of design/ manufacturing errors or material defects. This warranty do not apply if Accumate 2300 has been altered, misused, or repaired by persons not authorized by Woo Young Medical. The warranty do not cover the elimination of problems caused by incorrect operation, inappropriate handling normal wear and tear. 38
39 Service & Packaging Service If the pump fails to respond to the operating or troubleshooting procedures listed in this manual and the cause cannot be determined, discontinue use and forward it to an authorized Woo Young Medical's agency. With each complaint, please include pump serial number and full description of difficulty, including all setting, types of fluids, time, and alarm messages. Return IV-set used if possible. Woo Young Medical cannot assume responsibility for loss or damage to returned pumps while they are in transit. Packaging The pump packaging is reusable. Package Contents : Accumate 2300 Infusion Pump, power cable, operation manual 39
Nimbus Flex Ambulatory Infusion Pump
Nimbus Flex Ambulatory Infusion Pump Patient Manual Continuous Mode Infusions (Software Version v5.1.1) Read this entire manual prior to operating the Nimbus Ambulatory Infusion Pump. Failure to follow
More informationSentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual
Sentinel Enteral Feeding Pump Model: S-1000-SI Operating Manual CONTENTS 1. Introduction.. 2 2. Control Panel 5 3. Operating Instructions.. 6 4. Indicators and Alarms.. 9 5. Maintenance.. 10 6. Troubleshooting
More informationAmbulatory. Infusion. Pumps
PATIENT INFORMATION PCA Patient Controlled Analgesia Delivery Mode Model 6100 and 6101 Ambulatory Infusion Pumps Table of Contents Introduction... 1 Warnings... 2 Cautions... 4 CADD-Prizm Pump (Diagram)...
More informationBodyGuardTM 323. BodyGuard 323 Multi-Therapy Ambulatory Infusion Pump. Patient Guide
BodyGuardTM 323 BodyGuard 323 Multi-Therapy Ambulatory Infusion Pump Patient Guide How to Use This Booklet This booklet gives you information about how to use your BodyGuard 323 infusion pump. It supplements
More informationSR7 Micro-infusion Syringe Pump. Contents
Contents Introduction 2 A. About this manual.2 B. Manufacturer s Responsibility.2 C. Warranty 3 D. Unpacking and Accessories.. 3 E. General Safety...3 1.Overview.. 7 2.Operating Principle....7 3.Technical
More informationCole-Parmer Pro-Spense
Cole-Parmer Pro-Spense Portable Syringe Pump User s Manual Model #74905-44 Publication 5607-006-REV-A Contents Caution...2 Application Fields...2 Specifications...3 Operation Procedures: Keypad...4 Keypad
More informationQuick Reference Guide
Quick Reference Guide 3 Spike Set Cassette Inflow Connector Cassette Outflow Connector Return Line Latch Handle Patient Line Air Trap Keypad Cassette Customer Service Toll Free: 1-866-944-9992 Office:
More informationBodyGuard 121 Twins Infusion System Operator Manual
BodyGuard 121 Twins Infusion System Operator Manual 0344 Manufacturer: Caesarea Medical Electronics GmbH. 23 Staufenburg st. 72805 Lichtenstein, Germany +800-323-575-00 E-mail: Sales@cme-infusion.com www.cme-infusion.com
More informationCADD -Solis VIP Ambulatory Infusion Pump Model 2120
CADD -Solis VIP Ambulatory Infusion Pump Model 2120 Clinician Guide Contents I Description of the CADD -Solis VIP Pump... 4 Indications... 4 Pump Diagram... 5 Pump Descriptions... 5 Keypad... 6 I Pump
More informationUnit 2, Rivington View Business Park, Station Road, Blackrod, Bolton BL6 5BN Tel: +44(0)
The P900 enteral syringe pump has been designed to deliver small s of enteral feeds, accurately and safely, using enteral syringes, but cannot be used with IV syringes Unit 2, Rivington View Business Park,
More informationIntermittent Delivery Mode
Intermittent Delivery Mode CADD -Solis VIP Ambulatory Infusion Pump Model 2120 Patient Information Guide Your Clinician s: Name: Phone Number: Instructions: Pump placement during bathing/showering: Pump
More informationBodyGuard 323 Infusion Pump System Operator Manual
. BodyGuard 323 Infusion ump System Operator Manual 0344 Manufacturer: Caesarea Medical Electronics Ltd. European address: Staufenburgstr. 23 Lichtenstein.O.B. 1248 Lichtenstein 72805 Germany International
More informationBodyGuard 323 Infusion Pump System Operator Manual
. BodyGuard 323 Infusion ump System Operator Manual 0473 Manufacturer: USA Headquarters: Caesarea Medical Electronics Ltd. CME America, LLC European headquarters: 14998 W. 6 th Avenue, Unit 830 Staufenburgstr.
More informationCADD-Legacy 1 Ambulatory Infusion Pump Model 6400
OPERATOR S MANUAL CADD-Legacy 1 Ambulatory Infusion Pump Model 6400 CONTINUOUS DELIVERY This online version differs from the printed version. Certain information that is not intended for patients has been
More informationCADD-TPN. Model 5700 Operator s Manual. Deltec. Ambulatory Infusion System. SIMS Deltec, Inc., St. Paul, MN U.S.A.
CADD-TPN Ambulatory Infusion System Model 5700 Operator s Manual Deltec SIMS Deltec, Inc., St. Paul, MN 55112 U.S.A. This manual pertains only to the Deltec CADD-TPN, Model 5700, ambulatory infusion system.
More informationOPERATOR S MANUAL. Ambulatory Infusion Pump Model 6300 PATIENT CONTROLLED ANALGESIA
OPERATOR S MANUAL CADD-Legacy PCA Ambulatory Infusion Pump Model 6300 PATIENT CONTROLLED ANALGESIA This Operator s Manual is for clinician use only. Read the entire Operator s Manual before operating the
More informationCompanion. Enteral Pump. Operating Manual. For Enteral Use Only Not for IV Use
Companion Enteral Pump Operating Manual For Enteral Use Only Not for IV Use CONTENTS Pump Characteristics................................................. 3 Indications for Use..................................................
More informationOperator Manual of Guangzhou Huaxi Medical Technology Co., Ltd. JORVET INFUSION PUMP J1060 OPERATOR MANUAL
Operator Manual of Guangzhou Huaxi Medical Technology Co., Ltd. JORVET INFUSION PUMP J1060 OPERATOR MANUAL Please read manual before operating your JorVet Infusion Pump Jorgensen Laboratories 1450 N. Van
More informationBODYGUARD 595 REGIONAL ANALGESIA INFUSION PUMP
BODYGUARD 595 REGIONAL ANALGESIA INFUSION PUMP The BodyGuard 595 Regional Analgesia Infusion Pump is a dedicated, colourcoded pain management system for enhanced recovery wound infiltration and nerve blocks,
More informationCompanion. Enteral Nutrition Pump. Operating Manual. For Enteral Use Only Not for IV Use NOTES. No Purchase No.
NOTES No. 54748-Purchase No. 00084-Lease/Rent Companion Enteral Nutrition Pump Operating Manual #54748/00084 35691 LITHO IN USA For Enteral Use Only Not for IV Use Distributed by: Abbott Nutrition Abbott
More informationOperator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:
Storage System Ultrasound Probe Cabinet Operator's Manual Manufactured by: CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA 319.248.6757 / 800.445.6741 WWW.CIVCO.COM Copyright 2018 All
More informationMOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE
MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE TABLE OF CONTENTS Introduction...2 Important keys...2 Installing/Changing the Batteries...3 Preparing Your Therapy...4
More informationCADD-Prizm PCS II. Operator's Manual. Ambulatory Infusion Pump. Model 6101
x CADD-Prizm PCS II Ambulatory Infusion Pump Model 6101 Operator's Manual s This manual concerns only the CADD-Prizm PCS II (Pain Control System) Model 6101 Ambulatory Infusion Pump. This pump can be
More informationDeltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps Modular Infusion System
Deltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps Modular Infusion System Operator s Manual Published by Smiths Medical MD, Inc. All possible care has been taken in the preparation of this
More informationDigital Apex Locator. ipex OPERATION MANUAL OM-E0285E
Digital Apex Locator ipex OPERATION MANUAL 0197 OM-E0285E The EU directive 93/42/EEC was applied in the design and production of this medical device. Thank you for purchasing the ipex. This is apex locator.
More informationBODYGUARD 323 AMBULATORY INFUSION PUMP
BODYGUARD 323 AMBULATORY INFUSION PUMP The BodyGuard 323 Ambulatory Infusion Pump is a multi-therapy pump suitable for the hospital and homecare environments with three programmable modes (TPN, continuous
More informationCADD-1. Model 5100 HFX Operator s Manual. Note: Colors are Black and Cyan (as second color) Deltec. Ambulatory Infusion Pump
Note: Colors are Black and Cyan (as second color) CADD-1 Ambulatory Infusion Pump Model 5100 HFX Operator s Manual Deltec SIMS Deltec, Inc., St. Paul, MN 55112 U.S.A. i This manual pertains only to the
More informationQuickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority
Quickdraw Model 2400 Series Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com
More informationNIKI T34 pump. Knowing your way around the NIKI T34 pump
NIKI T34 pump Knowing your way around the NIKI T34 pump 2. Display Screen 3. Barrel clamp 4. Collar & Plunger Sensors 5. LED Indicator 1. Info Key 9. Up & Down arrow keys 8. YES (START)/ NO(STOP) keys
More informationOPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems
OPERATION AND MAINTENANCE INSTRUCTION MANUAL AEU-17B & AEU-17BV2 Implant / Surgery Systems TABLE OF CONTENTS: Specifications........................i Introduction.........................1 Package Contents....................1
More informationModel 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority
VX-2 Model 2310 Series Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com techsupport@sscor.com
More informationSURGICAL POWER & ACCESSORIES. Rx Only PM-X00-710/PM-X IR-12.13
SURGICAL POWER & ACCESSORIES Rx Only PM-X00-710/PM-X00-715-IR-12.13 Table of Contents Introduction Page 4 Intended Use Page 4 Warnings Page 5 Cautions Page 5 Notes Page 5 Explanation of Symbols Page 6
More informationDCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority
DCell Suction Model DM10-001 Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com
More informationBio Lion Table Top Centrifuge XC-L5
Bio Lion Table Top Centrifuge XC-L5 Operation Manual Table of contents Section Specification Page 1 Pictures Page 2-3 Starting Safety information Page 4 Set up Page 5-7 Operation Page 7-8 Troubleshooting
More informationTIF8800X. Combustible Gas Detector. an spx brand. Owner s Manual. 99 Washington Street Melrose, MA Phone Toll Free
an spx brand TIF8800X Combustible Gas Detector 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Owner s Manual Safety Precautions
More informationFLEXIFLO Companion. Enteral Nutrition Pump OPERATING MANUAL
FLEXIFLO Companion Enteral Nutrition Pump OPERATING MANUAL TABLE OF CONTENTS PAGE PUMP CHARACTERISTICS Pump Characteristics 3 Pediatric Use 3 Precautions 3 Illustrations 4-5 Control Dial Settings 6 Visual
More informationSonic Meter. Soni. Operating Manual. Rev. A
Sonic Meter Soni Operating Manual Rev. A Important: Do not attempt to operate this instrument without completely reading this manual and understanding the correct operating procedures. For online technical
More informationOPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems
OPERATION AND MAINTENANCE INSTRUCTION MANUAL AEU-707A & AEU-707AV2 Implant / Surgery Systems TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the
More informationOperator s Manual. spectrum 35700BAX & 35700ABB. Generation 2 Operating System Pump Operating Software v6.05 For Use with MDL Editor v6.2.
spectrum Operator s Manual 35700BAX & 35700ABB Generation 2 Operating System Pump Operating Software v6.05 For Use with MDL Editor v6.2.4 41018-6.05/6.2.4 Rev H November 29, 2012 SIGMA 711 Park Avenue
More informationMedi-Temp III FW601 Blood/Fluid Warmer Operator s Manual/Service Manual P/N /04
Medi-Temp III FW601 Blood/Fluid Warmer Operator s Manual/Service Manual P/N 12195-000 3/04 Table of Contents Sec Topic Page 1.0 Indications for Use... 1 2.0 Product Description... 2 3.0 Instructions for
More informationLIST OF FIGURES... ii. LIST OF WARNINGS and CAUTIONS... iii CHAPTER 1 - GENERAL INFORMATION... 1 CHAPTER 2 - DIRECTIONS FOR USE...
TABLE OF CONTENTS Page LIST OF FIGURES... ii LIST OF WARNINGS and CAUTIONS... iii CHAPTER 1 - GENERAL INFORMATION... 1 CHAPTER 2 - DIRECTIONS FOR USE... 2 Pump Features... 2 The Keypad... 4 The Display...
More informationSyringe Driver Introduction. Christine Hull & Anita Webb Staff Nurses, Hospice in the Home 2015
Syringe Driver Introduction Christine Hull & Anita Webb Staff Nurses, Hospice in the Home 2015 Aims and Objectives To gain an understanding of using the McKinleyT34 syringe driver. To review knowledge
More informationBio Lion Table Top Centrifuge XC-H165
Bio Lion Table Top Centrifuge XC-H165 Operation Manual Table of contents Section Specification Page 1 Pictures Page 2-3 Starting Safety information Page 4 Set up Page 5-7 Operation Page 7-8 Troubleshooting
More informationSmartVent. Instructions for Use Version 4 (Dec 2009)
SmartVent Instructions for Use Version 4 (Dec 2009) SmartVent is a radioaerosol delivery system that has been developed to give unrivalled performance for lung ventilation scintigraphy. Consistent, excellent
More information230VAC Power Inverter 400W Owner s Manual
400W 230VAC Power Inverter 400W Owner s Manual For safe and optimum performance, the Enerdrive epower Inverter must be used properly. Carefully read and follow all instructions and guidelines in this manual
More informationCordless Rechargeable Saw Instructions for Use
Technical data Voltage: DC 10.8V Weight: 1.25Kg Stroke rate: 0-2100/min Stroke: 15mm Cutting capacity: max diameter in wood 80mm / in soft metal 7mm Charging time: Between 5.0-5.5 Hours Battery: 1.3Ah
More informationEickemeyer Operating Manual Syringe Pump "Advanced"
Eickemeyer Contents Operating Manual Syringe Pump "Advanced" Chapter Page No. Rev.No.0.01 This manual has been compiled to users of the Syringe Pump in Installation, operation & routine maintenance. 1.
More informationFLEXIFLO QUANTUM Enteral Pump
FLEXIFLO QUANTUM Enteral Pump Operating Manual For Enteral Use Only Not for IV Use CONTENTS Pump Characteristics................................................ 3 Pediatric Use......................................................
More informationCGD8800X CGD8800X220 Combustible Gas Detector
CGD8800X CGD8800X220 Combustible Gas Detector Background and Purpose The CGD8800X Combustible Gas Detector offers significantly improved hydrocarbon detection while featuring visual concentration indicators
More informationCADD-PCA. Model 5800R Operator s Manual. Deltec. Ambulatory Infusion Pump. SIMS Deltec, Inc., St. Paul, MN U.S.A.
CADD-PCA Ambulatory Infusion Pump Model 5800R Operator s Manual Deltec SIMS Deltec, Inc., St. Paul, MN 55112 U.S.A. i This manual pertains only to the Deltec CADD-PCA, Model 5800R, infusion pump. The issue
More informationXenon Fiberoptic Lightsource
Xenon Fiberoptic Lightsource Operating Instructions for ASB-XE-175W Lightsource 2659A Pan American Frw N.E Albuquerque, New Mexico 87107 USA (English) Page 1 of 12 TABLE OF CONTENTS PAGE 1. INTRODUCTION
More informationUser Manual. Charger for LTO Batteries. User Manual. Language specific user manuals are available on
User Manual Charger for LTO Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d emploi Manual
More informationDedicated, ambulatory nerve block and wound infiltration treatment
Dedicated, ambulatory nerve block and wound infiltration treatment The BodyGuard 595 TM Regional Analgesia Infusion Pump is specially designed and colour coded for safe and effective nerve block and wound
More informationEvolution 3E INSTRUCTION MANUAL. Part 1/2 Ref.: 65051/US. MORIA S.A. 15, rue Georges Besse - F Antony France #65051/US-C-02.
INSTRUCTION MANUAL Part 1/2 Ref.: 65051/US MORIA S.A. 15, rue Georges Besse - F-92160 Antony France #65051/US-C-02.2017 1/22 CONTENTS I. EQUIPMENT LIST 3 II. DESCRIPTION OF THE CONTROL UNIT 4 A. Front
More informationUser Manual Stealth s Power Tilt
User Manual Stealth s Power Tilt Copyright 2014 Stealth Products, Inc. Copyright 2014 Stealth Products, Inc. All rights reserved. Published by Stealth Products, Inc. November 7, 2014 P45D61R1 Customer
More informationSolar Hybrid Power Generating System CPS1200EOH12SC CPS2200EOH24SC CPS3000EOH24SC. User s Manual K01-C
Solar Hybrid Power Generating System CPS1200EOH12SC CPS2200EOH24SC CPS3000EOH24SC User s Manual K01-C000304-02 2 TABLE OF CONTENTS 1 IMPORTANT SAFETY INSTRUCTIONS..4 2 INSTALLATION....5 2-1 Unpacking...5
More informationOperating Instructions. International English. CCS Dust Extractor
Operating Instructions International English CCS Dust Extractor Electric Cast Saw Dust Removal System Warnings & Safety Rules WARNING!: THIS EQUIPMENT MUST BE EARTHED! Danger!: Possible explosion hazard
More informationFreeGo. pump. Training Manual. Freedom for life
FreeGo pump Training Manual Freedom for life Contents Section 1 Set-up What you need Set up the FreeGo pump table top stand 2 Preparation Prepare the FreeGo pump Plug in and turn on Page 1 2 3 4 3 The
More informationOPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems
OPERATION AND MAINTENANCE INSTRUCTION MANUAL AEU-707A & AEU-707AV2 Implant / Surgery Systems TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the
More informationDIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE
DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE W.CHRISTIE (INDUSTRIAL) LTD CHRISTIE HOUSE, MEADOWBANK ROAD, ROTHERHAM, SOUTH YORKSHIRE, S61 2NF, UK T: +44(0)1709 550088 F: +44(0)1709 550030 E:
More informationSurePower Battery Pack Guide
? SurePower Battery Pack Guide 9650-0536-01 Rev. B The issue date for the SurePower Battery Pack Guide (9650-0536-01 Rev. B) is September, 2010. If more than three years have elapsed since the issue date,
More informationOPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH
OPERATING MANUAL POWER DRIVE SYSTEM Cat. No. 400005 525-1300-00 RevH System components Symbol Definitions Qty Description Catalog Number 1 Power Drive System 300106 1 Controller 300108 1 Cutter Drive
More informationB-RAD Select USER MANUAL TABLE OF CONTENTS
TABLE OF CONTENTS TABLE OF CONTENTS... 1 MANUAL REVISION HISTORY... 2 IMPORTANT SAFETY NOTICE... 3 1.0 General Information... 5 1.1 System Components... 5 1.2 Specifications... 5 1.2.1 Torque Ranges...
More informationUser s Manual. Versaflo 200 Syringe Pump. For Veterinary Use Only
User s Manual For Veterinary Use Only Versaflo 200 Syringe Pump Midmark Corporation 60 Vista Drive, Box 286 Versailles, OH 45380-0286 Customer Service and Technical Service TEL: 1-800-MIDMARK (1-800-643-6275)
More informationOperator's Manual Tracer III Nerve Locator, Model NL-3
Operator's Manual Tracer III Nerve Locator, Model NL-3 2003-2004, Life-Tech, Inc. All rights reserved. M1043 Rev C IMPORTANT! - IN THIS MANUAL WARNING STATEMENTS: Indicate information or procedures that,
More informationPower Inverter 400 MW Owner s Manual
Power Inverter 400 MW 1204 Owner s Manual For safe and optimum performance, the Power Inverter must be used properly. Carefully read and follow all instructions and guidelines in this manual and give special
More informationPower Inverter. User s Manual. Diamond Series CPD1200EILCD
Power Inverter Diamond Series CPD1200EILCD User s Manual 2 TABLE OF CONTENTS 1 IMPORTANT SAFETY INSTRUCTIONS..4 2 INSTALLATION....5 2-1 Unpacking...5 2-2 Product Overview & Outlook..5 2-3 Power Requirements
More informationSyringe Infusion Solutions
Syringe Infusion Solutions Caesarea Medical Electronics T34 Deliver Syringe Infusion Anywhere rovide complete small-dose infusion therapy safely and economically wherever your patients receive care. The
More information6 Bay Battery Charger Operation Manual
6 Bay Battery Charger Operation Manual Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216 USA Customer Service: 904 737 7611 Toll Free 877 677 2832 EC REP RMS UK, Ltd. 28 Trinity Road
More informationFOR YOUR PERMANENT FILE
FOR YOUR PERMANENT FILE WRENCH MODEL NUMBER: operation manual torque and angle electronic WRENCH WRENCH SERIAL NUMBER: For Warranty Claims, Contact CDI Torque Products at (626) 965-0668. LIMITED WARRANTY
More informationSurePower TM II Battery Pack Guide
TM SurePower TM II Battery Pack Guide? 1:00+ 2:00+ 3:00+ TM 9650-000840-01 Rev. B The issue date for the SurePower II Battery Pack Guide (REF 9650-000840-01 Rev. B) is December, 2012. If more than three
More informationCADD-Solis Infusion Pump
Introduction CADD-Solis Infusion Pump Information for Parents Your child has been started on an infusion using the CADD-Solis pump. Your nurse specialist will have explained the reasons for starting the
More informationPatient s Manual for Using Nimbus Ambulatory Pump for Continuous Peripheral Nerve Block Application
Patient s Manual for Using Nimbus Ambulatory Pump for Continuous Peripheral Nerve Block Application PCA Mode Infusions (Software Version v2.1.3) Read this entire Manual prior to operating the Nimbus Ambulatory
More informationSyringe Infusion Solutions
Syringe Infusion Solutions Caesarea Medical Electronics T34 Deliver Syringe Infusion Anywhere rovide complete small-dose infusion therapy safely and economically wherever your patients receive care. The
More information1. D-Lux Introduction ENGLISH Parts & Components Product Overview. 4. Precautions. 5. Hand-piece Components Installation
1. D-Lux Introduction ENGLISH D-Lux is a new advanced cordless LED curing light. Made from the highest quality aluminum, D-Lux dissipates heat quickly to prevent overheating problem so that it can produce
More informationUser s Manual. Automatic Switch-Mode Battery Charger
User s Manual Automatic Switch-Mode Battery Charger IMPORTANT Read, understand, and follow these safety rules and operating instructions before using this battery charger. Only authorized and trained service
More informationLanguage specific user manuals are available on
User Manual WEB Charger for LiFePO 4 Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d
More informationNature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual
Version 1.1 Version 2 Nature Power Inverters True Sinewave Inverter Modified Sinewave Inverter Owner s Manual!!!!!!!!!!! 38304 38204 For safe and optimum performance, the Power Inverter must be used properly.
More informationCADD -Solis VIP Ambulatory Infusion Pump. Technical Manual Model 2120
X CADD -Solis VIP Ambulatory Infusion Pump Technical Manual Model 2120 s The issue date of this Technical Manual is included on the back cover. In the event one year has elapsed between the issue date
More informationModel 2053 Conductivity Meter Instruction Manual
Model 2053 Conductivity Meter Instruction Manual Printed in U.S.A. 09/2006 Ship Date: Serial Number: Calibrated By: Table of Contents Page 1. Introduction.. 1 2. Shipping Checklist.. 1 3. Specifications...
More informationOPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-70A Peristaltic Infusion Pump
OPERATION and MAINTENANCE INSTRUCTION MANUAL AEU-70A Peristaltic Infusion Pump Your new AEU-70A Aseptico Surgery Infusion Pump is one of the finest units available to the medical profession. The system
More informationLaerdal Suction Unit BEMIS Disposable Canister
Laerdal Suction Unit BEMIS Disposable Canister EN User Guide Cat. No. 78002001 www.laerdal.com 1 Cautions and Warnings Read this User Guide carefully, and become thoroughly familiar with the operation
More informationINSTRUCTION MANUAL VISCOTESTER VT-05 / VT-06 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
INSTRUCTION MANUAL VISCOTESTER VT-05 / VT-06 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.4.17 Organization
More informationFor detailed instructions, specifications, warnings, warranties, and additional information on operating the CADD -Solis ambulatory infusion pump,
X s For detailed instructions, specifications, warnings, warranties, and additional information on operating the CADD -Solis ambulatory infusion pump, please refer to the manuals supplied with the product.
More informationInstructions For Use VATEA. Endodontic Irrigation Device
Instructions For Use VATEA Endodontic Irrigation Device VATEA Irrigation System Instructions for use 1. Indications for use The VATEA system is intended to be attached to dental handpieces to deliver irrigation
More informationChair-Check II User Manual
Chair-Check II User Manual 0163-437-H Reducing the Risk of Falls Through Innovative Technology Since 1978 1-800-523-7956 www.bedcheck.com 2008 The Stanley Works All Rights Reserved Bed-Check, Chair-Check,
More informationSwitching DC Power Supply
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Model 1693, 1694 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary
More informationBattery Power Inverters
Battery Power Inverters Renogy 500W 1000W 2000W Pure Sine Wave Inverter Manual 2775 E. Philadelphia St., Ontario, CA 91761 1-800-330-8678 1 Version 1.4 Important Safety Instructions Please save these instructions.
More informationPodiatry Procedures Chair. For Models: Barrier-Free. User s Guide
Podiatry Procedures Chair For Models: 646 647 Barrier-Free User s Guide Product Information (The information below is required when calling for service.) Dealer : Date of Purchase: Model / Serial Number:
More informationPRODUCT CODE: 2M8153, 2M8153K
PRODUCT CODE: 2M8153, 2M8153K. Prior to using this pump, read this manual carefully to fully understand the pump s functionality and to ensure safe and proper operation. Warning There are risks associated
More information1.Safe testing IMPORTANT:
Electricity is dangerous and can cause injury and death. Always treat it with the greatest of respect and care. If you are not quite sure how to proceed, then stop, take advice from a qualified person.
More informationMulti Therapy Solutions
Multi Therapy Solutions Caesarea Medical Electronics BodyGuard 323 Color Vision Advanced Multi Therapy Infusion ump Join the frontline of infusion therapy with BodyGuard 323 Color Vision, the latest multi-purpose
More informationRemRest. Service Manual. CAUTION: USA Federal law restricts this device to sale by or on the order of a physician. P/N Rev A
RemRest Tiers 1-3 900 Series Service Manual CAUTION: USA Federal law restricts this device to sale by or on the order of a physician. P/N 22054 Rev A Table of Contents IEC Symbols 3 SIZE 4.7" H X 8.6"
More informationOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS Digital Clamp Meter DSA500A IMPORTANT: RECEIVING INSTRUCTIONS Visually inspect all components for shipping damage. If you find damage, notify the carrier at once. Shipping damage
More informationMEDICAL IRRIGATION K PUMP OPERATING MANUAL: Manufactured for: K.M.I 3185 Palisades Dr Corona CA USA Model No.
OPERATING MANUAL: MEDICAL IRRIGATION K PUMP Manufactured for: K.M.I 3185 Palisades Dr Corona CA 92880-9432 USA 1-866-412-7867 Model No. 1000-0031 A-1299-5091 Edition 01 TABLE OF CONTENTS Title SAFETY PRECAUTIONS
More informationOperating Manual. For Enteral Use Only Not for IV Use
Operating Manual For Enteral Use Only Not for IV Use CONTENTS Pump Characteristics................................................. 3 Indications for Use..................................................
More informationOperators Guide: RoboSign Stop/Go Traffic Control System
Operators Guide: RoboSign Stop/Go Traffic Control System RoboSign Remote controlled Stop/Go temporary traffic control system Operators Guide Table of Contents Operators Guide: RoboSign Stop/Go Traffic
More informationNT-500 Technical Support Guide. For use by qualified company approved Technicians
NT-500 Technical Support Guide For use by qualified company approved Technicians INTRODUCTION The NT-500 is designed to offer the full range of features required to perform radiofrequency denervation The
More informationSLA Battery Capacity Analyzer
Model: 600B SLA Battery Capacity Analyzer USER MANUAL Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation,
More informationMOdel No: MM19.V3 WARNING CAUTION 1. SAFETY INSTRUCTIONS
Instructions for: Digital Multimeter - 6 Function MOdel No: MM19.V3 Thank you for purchasing a Sealey product. Manufactured to a high standard this product will, if used according to these instructions
More information